Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol
Objective: To evaluate the effect of low tube voltage (100 kV) combined with adaptive statistical iterative reconstruction-V (ASIR-V) on the visualization and image quality of the Adamkiewicz artery (AKA). Methods: One hundred patients were prospectively enrolled and randomly assigned into two group...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10417362/ https://www.ncbi.nlm.nih.gov/pubmed/37568857 http://dx.doi.org/10.3390/diagnostics13152495 |
_version_ | 1785088014616100864 |
---|---|
author | Deng, Jiantao Ma, Ting Yan, Jing Wu, Siyi Yan, Gaowu Li, Hongwei Li, Yong Zhao, Linwei Fan, Xiaoping McClure, Morgan A. Bhetuwal, Anup |
author_facet | Deng, Jiantao Ma, Ting Yan, Jing Wu, Siyi Yan, Gaowu Li, Hongwei Li, Yong Zhao, Linwei Fan, Xiaoping McClure, Morgan A. Bhetuwal, Anup |
author_sort | Deng, Jiantao |
collection | PubMed |
description | Objective: To evaluate the effect of low tube voltage (100 kV) combined with adaptive statistical iterative reconstruction-V (ASIR-V) on the visualization and image quality of the Adamkiewicz artery (AKA). Methods: One hundred patients were prospectively enrolled and randomly assigned into two groups (both n = 50). Group A (100 kV) was reconstructed with filtered back projection (FBP) and ASIR-V from 10% to 100% with 10% intervals. Group B (120 kV) was only reconstructed with FBP. The objective image quality was evaluated by using CT values of the aorta (CT(Aorta)), background noise, signal-to-noise ratio of the descending aorta (SNR(Aorta)), and contrast-to-noise ratio of the spinal cord (CNR(Spinal cord)). The subjective image quality and visualization scores of the AKA were assessed on a 5-point scale. Results: CT(Aorta) was significantly higher in Group A than in Group B (p < 0.001). When ASIR-V weights were ≥60%, significant differences were found in the background noise, SNR(Aorta), and CNR(Spinal cord) between the two groups (all p < 0.05). In Group A, compared with FBP, the subjective score gradually increased as ASIR-V increased to 80%, which decreased when ASIR-V exceeded 80%. The visualization scores of the AKA (≥60%) and the ability to detect vessel continuity (≥80%) gradually increased as the ASIR-V weights increased (p < 0.05). The effective radiation dose was reduced by about 40.36% in Group A compared to Group B. Conclusions: compared with conventional scanning protocol, using a combination of low tube voltage (100 kV) and 80% ASIR-V protocol could not only increase the visualization of the AKA, but also improve image quality and reduce the radiation doses. |
format | Online Article Text |
id | pubmed-10417362 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-104173622023-08-12 Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol Deng, Jiantao Ma, Ting Yan, Jing Wu, Siyi Yan, Gaowu Li, Hongwei Li, Yong Zhao, Linwei Fan, Xiaoping McClure, Morgan A. Bhetuwal, Anup Diagnostics (Basel) Article Objective: To evaluate the effect of low tube voltage (100 kV) combined with adaptive statistical iterative reconstruction-V (ASIR-V) on the visualization and image quality of the Adamkiewicz artery (AKA). Methods: One hundred patients were prospectively enrolled and randomly assigned into two groups (both n = 50). Group A (100 kV) was reconstructed with filtered back projection (FBP) and ASIR-V from 10% to 100% with 10% intervals. Group B (120 kV) was only reconstructed with FBP. The objective image quality was evaluated by using CT values of the aorta (CT(Aorta)), background noise, signal-to-noise ratio of the descending aorta (SNR(Aorta)), and contrast-to-noise ratio of the spinal cord (CNR(Spinal cord)). The subjective image quality and visualization scores of the AKA were assessed on a 5-point scale. Results: CT(Aorta) was significantly higher in Group A than in Group B (p < 0.001). When ASIR-V weights were ≥60%, significant differences were found in the background noise, SNR(Aorta), and CNR(Spinal cord) between the two groups (all p < 0.05). In Group A, compared with FBP, the subjective score gradually increased as ASIR-V increased to 80%, which decreased when ASIR-V exceeded 80%. The visualization scores of the AKA (≥60%) and the ability to detect vessel continuity (≥80%) gradually increased as the ASIR-V weights increased (p < 0.05). The effective radiation dose was reduced by about 40.36% in Group A compared to Group B. Conclusions: compared with conventional scanning protocol, using a combination of low tube voltage (100 kV) and 80% ASIR-V protocol could not only increase the visualization of the AKA, but also improve image quality and reduce the radiation doses. MDPI 2023-07-27 /pmc/articles/PMC10417362/ /pubmed/37568857 http://dx.doi.org/10.3390/diagnostics13152495 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Deng, Jiantao Ma, Ting Yan, Jing Wu, Siyi Yan, Gaowu Li, Hongwei Li, Yong Zhao, Linwei Fan, Xiaoping McClure, Morgan A. Bhetuwal, Anup Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol |
title | Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol |
title_full | Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol |
title_fullStr | Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol |
title_full_unstemmed | Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol |
title_short | Effect of Low Tube Voltage (100 kV) Combined with ASIR-V on the Visualization and Image Quality of the Adamkiewicz Artery: A Comparison with 120 kV Protocol |
title_sort | effect of low tube voltage (100 kv) combined with asir-v on the visualization and image quality of the adamkiewicz artery: a comparison with 120 kv protocol |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10417362/ https://www.ncbi.nlm.nih.gov/pubmed/37568857 http://dx.doi.org/10.3390/diagnostics13152495 |
work_keys_str_mv | AT dengjiantao effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT mating effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT yanjing effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT wusiyi effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT yangaowu effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT lihongwei effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT liyong effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT zhaolinwei effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT fanxiaoping effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT mccluremorgana effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol AT bhetuwalanup effectoflowtubevoltage100kvcombinedwithasirvonthevisualizationandimagequalityoftheadamkiewiczarteryacomparisonwith120kvprotocol |