Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study
Advances in stent design and technology have made stent loss during percutaneous coronary interventions rare. When an undeployed stent dislodges in the left main (LM) artery during percutaneous coronary angioplasty, the risk of life-threatening procedural complications is high. We report a 50-year-o...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10424754/ https://www.ncbi.nlm.nih.gov/pubmed/37584019 http://dx.doi.org/10.4103/heartviews.heartviews_120_22 |
_version_ | 1785089736783691776 |
---|---|
author | Abdelaziz, Abdelaziz Ahmed Taleb, Ahmed Maher Abo |
author_facet | Abdelaziz, Abdelaziz Ahmed Taleb, Ahmed Maher Abo |
author_sort | Abdelaziz, Abdelaziz Ahmed |
collection | PubMed |
description | Advances in stent design and technology have made stent loss during percutaneous coronary interventions rare. When an undeployed stent dislodges in the left main (LM) artery during percutaneous coronary angioplasty, the risk of life-threatening procedural complications is high. We report a 50-year-old male patient, a smoker, with a history of diabetes mellitus and hypertension with typical chest pain on minimal exertion. Electrocardiogram and echo revealed ischemic changes and regional wall motion abnormality. Culotte technique was used. A new 3 mm × 48 mm stent was inserted in the LM-left circumflex (LM-LCX) followed by stenting of the LM-left anterior descending (LM-LAD) ostia with a 3.5 mm × 18 mm stent. The two balloons were rewired and kissed. Stent slippage and dislodging in the LM artery can be corrected using the culotte technique to crush the undeployed stent behind the LM-LCX and LM-LAD stents. |
format | Online Article Text |
id | pubmed-10424754 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-104247542023-08-15 Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study Abdelaziz, Abdelaziz Ahmed Taleb, Ahmed Maher Abo Heart Views Case Report Advances in stent design and technology have made stent loss during percutaneous coronary interventions rare. When an undeployed stent dislodges in the left main (LM) artery during percutaneous coronary angioplasty, the risk of life-threatening procedural complications is high. We report a 50-year-old male patient, a smoker, with a history of diabetes mellitus and hypertension with typical chest pain on minimal exertion. Electrocardiogram and echo revealed ischemic changes and regional wall motion abnormality. Culotte technique was used. A new 3 mm × 48 mm stent was inserted in the LM-left circumflex (LM-LCX) followed by stenting of the LM-left anterior descending (LM-LAD) ostia with a 3.5 mm × 18 mm stent. The two balloons were rewired and kissed. Stent slippage and dislodging in the LM artery can be corrected using the culotte technique to crush the undeployed stent behind the LM-LCX and LM-LAD stents. Wolters Kluwer - Medknow 2023 2023-07-05 /pmc/articles/PMC10424754/ /pubmed/37584019 http://dx.doi.org/10.4103/heartviews.heartviews_120_22 Text en Copyright: © 2023 Heart Views https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Case Report Abdelaziz, Abdelaziz Ahmed Taleb, Ahmed Maher Abo Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study |
title | Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study |
title_full | Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study |
title_fullStr | Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study |
title_full_unstemmed | Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study |
title_short | Slippage of an Undeployed Stent in the Left Main Artery: A Case Report Study |
title_sort | slippage of an undeployed stent in the left main artery: a case report study |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10424754/ https://www.ncbi.nlm.nih.gov/pubmed/37584019 http://dx.doi.org/10.4103/heartviews.heartviews_120_22 |
work_keys_str_mv | AT abdelazizabdelazizahmed slippageofanundeployedstentintheleftmainarteryacasereportstudy AT talebahmedmaherabo slippageofanundeployedstentintheleftmainarteryacasereportstudy |