A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09‐A3), carryover (according to CLSI...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10431393/ https://www.ncbi.nlm.nih.gov/pubmed/37431092 http://dx.doi.org/10.1002/jcla.24929 |
_version_ | 1785091190132047872 |
---|---|
author | Pontis, Adeline Delanoe, Mélanie Schilliger, Nathalie Carlo, Audrey Guéret, Pierre Nédélec‐Gac, Fabienne Gouin‐Thibault, Isabelle |
author_facet | Pontis, Adeline Delanoe, Mélanie Schilliger, Nathalie Carlo, Audrey Guéret, Pierre Nédélec‐Gac, Fabienne Gouin‐Thibault, Isabelle |
author_sort | Pontis, Adeline |
collection | PubMed |
description | AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09‐A3), carryover (according to CLSI H57‐A), APTT sensitivity to heparin (according to CLSI H47‐A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15‐A3) and accuracy. The assays tested on sthemO 301 were: PT, APTT (silica and kaolin activators), fibrinogen (Fib), thrombin time (TT), chromogenic and clotting protein C (PC) activity, and von Willebrand factor antigen (VWF:Ag) levels. RESULTS: All intra‐assay and inter‐assay precision CVs were below the maximal precision limit proposed by the French Group for Hemostasis and Thrombosis (GFHT). Accuracy was verified with bias below GFHT criteria and most Z‐scores were between −2 and +2. No clinically relevant carryover was detected. Silica APTT reagent sensitivity to unfractionated heparin was moderate, as expected. Productivity results were consistent over the 10 repeats performed. The overall agreement between the two systems was excellent for all assays, with Spearman rank correlation coefficient all above 0.9 and slopes of Passing–Bablok correlation near 1 and intercepts close to 0. CONCLUSION: For the methods tested, sthemO 301 system met all the criteria to implement a novel coagulation analyzer in the laboratory and result comparability with STA R Max® 2 was good. |
format | Online Article Text |
id | pubmed-10431393 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-104313932023-08-17 A performance evaluation of sthemO 301 coagulation analyzer and associated reagents Pontis, Adeline Delanoe, Mélanie Schilliger, Nathalie Carlo, Audrey Guéret, Pierre Nédélec‐Gac, Fabienne Gouin‐Thibault, Isabelle J Clin Lab Anal Research Articles AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09‐A3), carryover (according to CLSI H57‐A), APTT sensitivity to heparin (according to CLSI H47‐A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15‐A3) and accuracy. The assays tested on sthemO 301 were: PT, APTT (silica and kaolin activators), fibrinogen (Fib), thrombin time (TT), chromogenic and clotting protein C (PC) activity, and von Willebrand factor antigen (VWF:Ag) levels. RESULTS: All intra‐assay and inter‐assay precision CVs were below the maximal precision limit proposed by the French Group for Hemostasis and Thrombosis (GFHT). Accuracy was verified with bias below GFHT criteria and most Z‐scores were between −2 and +2. No clinically relevant carryover was detected. Silica APTT reagent sensitivity to unfractionated heparin was moderate, as expected. Productivity results were consistent over the 10 repeats performed. The overall agreement between the two systems was excellent for all assays, with Spearman rank correlation coefficient all above 0.9 and slopes of Passing–Bablok correlation near 1 and intercepts close to 0. CONCLUSION: For the methods tested, sthemO 301 system met all the criteria to implement a novel coagulation analyzer in the laboratory and result comparability with STA R Max® 2 was good. John Wiley and Sons Inc. 2023-07-10 /pmc/articles/PMC10431393/ /pubmed/37431092 http://dx.doi.org/10.1002/jcla.24929 Text en © 2023 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Articles Pontis, Adeline Delanoe, Mélanie Schilliger, Nathalie Carlo, Audrey Guéret, Pierre Nédélec‐Gac, Fabienne Gouin‐Thibault, Isabelle A performance evaluation of sthemO 301 coagulation analyzer and associated reagents |
title | A performance evaluation of sthemO 301 coagulation analyzer and associated reagents |
title_full | A performance evaluation of sthemO 301 coagulation analyzer and associated reagents |
title_fullStr | A performance evaluation of sthemO 301 coagulation analyzer and associated reagents |
title_full_unstemmed | A performance evaluation of sthemO 301 coagulation analyzer and associated reagents |
title_short | A performance evaluation of sthemO 301 coagulation analyzer and associated reagents |
title_sort | performance evaluation of sthemo 301 coagulation analyzer and associated reagents |
topic | Research Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10431393/ https://www.ncbi.nlm.nih.gov/pubmed/37431092 http://dx.doi.org/10.1002/jcla.24929 |
work_keys_str_mv | AT pontisadeline aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT delanoemelanie aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT schilligernathalie aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT carloaudrey aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT gueretpierre aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT nedelecgacfabienne aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT gouinthibaultisabelle aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT pontisadeline performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT delanoemelanie performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT schilligernathalie performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT carloaudrey performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT gueretpierre performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT nedelecgacfabienne performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents AT gouinthibaultisabelle performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents |