A performance evaluation of sthemO 301 coagulation analyzer and associated reagents

AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09‐A3), carryover (according to CLSI...

Descripción completa

Detalles Bibliográficos
Autores principales: Pontis, Adeline, Delanoe, Mélanie, Schilliger, Nathalie, Carlo, Audrey, Guéret, Pierre, Nédélec‐Gac, Fabienne, Gouin‐Thibault, Isabelle
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10431393/
https://www.ncbi.nlm.nih.gov/pubmed/37431092
http://dx.doi.org/10.1002/jcla.24929
_version_ 1785091190132047872
author Pontis, Adeline
Delanoe, Mélanie
Schilliger, Nathalie
Carlo, Audrey
Guéret, Pierre
Nédélec‐Gac, Fabienne
Gouin‐Thibault, Isabelle
author_facet Pontis, Adeline
Delanoe, Mélanie
Schilliger, Nathalie
Carlo, Audrey
Guéret, Pierre
Nédélec‐Gac, Fabienne
Gouin‐Thibault, Isabelle
author_sort Pontis, Adeline
collection PubMed
description AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09‐A3), carryover (according to CLSI H57‐A), APTT sensitivity to heparin (according to CLSI H47‐A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15‐A3) and accuracy. The assays tested on sthemO 301 were: PT, APTT (silica and kaolin activators), fibrinogen (Fib), thrombin time (TT), chromogenic and clotting protein C (PC) activity, and von Willebrand factor antigen (VWF:Ag) levels. RESULTS: All intra‐assay and inter‐assay precision CVs were below the maximal precision limit proposed by the French Group for Hemostasis and Thrombosis (GFHT). Accuracy was verified with bias below GFHT criteria and most Z‐scores were between −2 and +2. No clinically relevant carryover was detected. Silica APTT reagent sensitivity to unfractionated heparin was moderate, as expected. Productivity results were consistent over the 10 repeats performed. The overall agreement between the two systems was excellent for all assays, with Spearman rank correlation coefficient all above 0.9 and slopes of Passing–Bablok correlation near 1 and intercepts close to 0. CONCLUSION: For the methods tested, sthemO 301 system met all the criteria to implement a novel coagulation analyzer in the laboratory and result comparability with STA R Max® 2 was good.
format Online
Article
Text
id pubmed-10431393
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-104313932023-08-17 A performance evaluation of sthemO 301 coagulation analyzer and associated reagents Pontis, Adeline Delanoe, Mélanie Schilliger, Nathalie Carlo, Audrey Guéret, Pierre Nédélec‐Gac, Fabienne Gouin‐Thibault, Isabelle J Clin Lab Anal Research Articles AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09‐A3), carryover (according to CLSI H57‐A), APTT sensitivity to heparin (according to CLSI H47‐A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15‐A3) and accuracy. The assays tested on sthemO 301 were: PT, APTT (silica and kaolin activators), fibrinogen (Fib), thrombin time (TT), chromogenic and clotting protein C (PC) activity, and von Willebrand factor antigen (VWF:Ag) levels. RESULTS: All intra‐assay and inter‐assay precision CVs were below the maximal precision limit proposed by the French Group for Hemostasis and Thrombosis (GFHT). Accuracy was verified with bias below GFHT criteria and most Z‐scores were between −2 and +2. No clinically relevant carryover was detected. Silica APTT reagent sensitivity to unfractionated heparin was moderate, as expected. Productivity results were consistent over the 10 repeats performed. The overall agreement between the two systems was excellent for all assays, with Spearman rank correlation coefficient all above 0.9 and slopes of Passing–Bablok correlation near 1 and intercepts close to 0. CONCLUSION: For the methods tested, sthemO 301 system met all the criteria to implement a novel coagulation analyzer in the laboratory and result comparability with STA R Max® 2 was good. John Wiley and Sons Inc. 2023-07-10 /pmc/articles/PMC10431393/ /pubmed/37431092 http://dx.doi.org/10.1002/jcla.24929 Text en © 2023 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Pontis, Adeline
Delanoe, Mélanie
Schilliger, Nathalie
Carlo, Audrey
Guéret, Pierre
Nédélec‐Gac, Fabienne
Gouin‐Thibault, Isabelle
A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
title A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
title_full A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
title_fullStr A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
title_full_unstemmed A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
title_short A performance evaluation of sthemO 301 coagulation analyzer and associated reagents
title_sort performance evaluation of sthemo 301 coagulation analyzer and associated reagents
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10431393/
https://www.ncbi.nlm.nih.gov/pubmed/37431092
http://dx.doi.org/10.1002/jcla.24929
work_keys_str_mv AT pontisadeline aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT delanoemelanie aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT schilligernathalie aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT carloaudrey aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT gueretpierre aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT nedelecgacfabienne aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT gouinthibaultisabelle aperformanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT pontisadeline performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT delanoemelanie performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT schilligernathalie performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT carloaudrey performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT gueretpierre performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT nedelecgacfabienne performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents
AT gouinthibaultisabelle performanceevaluationofsthemo301coagulationanalyzerandassociatedreagents