Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation
BACKGROUND: Ophthalmic brachytherapy using radioactive plaques is an effective technique for the treatment of uveal melanoma. Ru-106 eye plaques are considered as interesting issue due to their steep gradient dose. The pre-planning evaluation of dosimetric parameters is essential for the treatment p...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Shiraz University of Medical Sciences
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10440413/ https://www.ncbi.nlm.nih.gov/pubmed/37609514 http://dx.doi.org/10.31661/jbpe.v0i0.2010-1206 |
_version_ | 1785093146424639488 |
---|---|
author | Arjmand, Mojtaba Ghassemi, Fariba Rafiepour, Payman Zeinali, Reyhaneh Riazi-Esfahani, Hamid Beiki-Ardakani, Akbar |
author_facet | Arjmand, Mojtaba Ghassemi, Fariba Rafiepour, Payman Zeinali, Reyhaneh Riazi-Esfahani, Hamid Beiki-Ardakani, Akbar |
author_sort | Arjmand, Mojtaba |
collection | PubMed |
description | BACKGROUND: Ophthalmic brachytherapy using radioactive plaques is an effective technique for the treatment of uveal melanoma. Ru-106 eye plaques are considered as interesting issue due to their steep gradient dose. The pre-planning evaluation of dosimetric parameters is essential for the treatment planning system. OBJECTIVE: The current study aims at providing dose distributions of six Ru-106 eye plaques (CCA, CCB, CGD, CIB, COB and COD) using radiochromic EBT3 film, Geant4 Monte Carlo toolkit and the treatment planning software (Plaque Simulator). MATERIAL AND METHODS: In this experimental study, an in-house phantom was employed for depth dose measurements with EBT3 films. Also, Geant4.10.5 scoring mesh was implemented to obtain the 2D dose distribution of the plaques. The results were compared with Plaque Simulator software and the manufacturer’s (BEBIG) data. The gamma index criterion (3%/3 mm) was used to evaluate dose distributions obtained by the film measurements and Geant4 simulation. RESULTS: A good agreement was achieved between simulation and experimental results. Gamma index passing rate was 94.2%, 89.3%, 88.2%, 82.2%, 92.2% and 90.1% for CCA, CCB, CGD, CIB, COB and COD plaques, respectively. Absolute dose rate (mGy/min) obtained by EBT3 film at the depth of 2 mm was 79.4 mGy/min, 81.0 mGy/min, 78.6 mGy/min, 62.2 mGy/min, 75.2 mGy/min and 81.2 mGy/min for CCA, CCB, CGD, CIB, COB and COD plaques, respectively. CONCLUSION: The measured dose distributions and lateral dose profiles may be utilized in the treatment planning system to cover clinical volumes such as the clinical target volume and the gross tumor volume. |
format | Online Article Text |
id | pubmed-10440413 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Shiraz University of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-104404132023-08-22 Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation Arjmand, Mojtaba Ghassemi, Fariba Rafiepour, Payman Zeinali, Reyhaneh Riazi-Esfahani, Hamid Beiki-Ardakani, Akbar J Biomed Phys Eng Original Article BACKGROUND: Ophthalmic brachytherapy using radioactive plaques is an effective technique for the treatment of uveal melanoma. Ru-106 eye plaques are considered as interesting issue due to their steep gradient dose. The pre-planning evaluation of dosimetric parameters is essential for the treatment planning system. OBJECTIVE: The current study aims at providing dose distributions of six Ru-106 eye plaques (CCA, CCB, CGD, CIB, COB and COD) using radiochromic EBT3 film, Geant4 Monte Carlo toolkit and the treatment planning software (Plaque Simulator). MATERIAL AND METHODS: In this experimental study, an in-house phantom was employed for depth dose measurements with EBT3 films. Also, Geant4.10.5 scoring mesh was implemented to obtain the 2D dose distribution of the plaques. The results were compared with Plaque Simulator software and the manufacturer’s (BEBIG) data. The gamma index criterion (3%/3 mm) was used to evaluate dose distributions obtained by the film measurements and Geant4 simulation. RESULTS: A good agreement was achieved between simulation and experimental results. Gamma index passing rate was 94.2%, 89.3%, 88.2%, 82.2%, 92.2% and 90.1% for CCA, CCB, CGD, CIB, COB and COD plaques, respectively. Absolute dose rate (mGy/min) obtained by EBT3 film at the depth of 2 mm was 79.4 mGy/min, 81.0 mGy/min, 78.6 mGy/min, 62.2 mGy/min, 75.2 mGy/min and 81.2 mGy/min for CCA, CCB, CGD, CIB, COB and COD plaques, respectively. CONCLUSION: The measured dose distributions and lateral dose profiles may be utilized in the treatment planning system to cover clinical volumes such as the clinical target volume and the gross tumor volume. Shiraz University of Medical Sciences 2023-08-01 /pmc/articles/PMC10440413/ /pubmed/37609514 http://dx.doi.org/10.31661/jbpe.v0i0.2010-1206 Text en Copyright: © Journal of Biomedical Physics and Engineering https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License, ( http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Arjmand, Mojtaba Ghassemi, Fariba Rafiepour, Payman Zeinali, Reyhaneh Riazi-Esfahani, Hamid Beiki-Ardakani, Akbar Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation |
title | Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation |
title_full | Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation |
title_fullStr | Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation |
title_full_unstemmed | Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation |
title_short | Dosimetric Investigation of Six Ru-106 Eye Plaques by EBT3 Radiochromic Films and Monte Carlo Simulation |
title_sort | dosimetric investigation of six ru-106 eye plaques by ebt3 radiochromic films and monte carlo simulation |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10440413/ https://www.ncbi.nlm.nih.gov/pubmed/37609514 http://dx.doi.org/10.31661/jbpe.v0i0.2010-1206 |
work_keys_str_mv | AT arjmandmojtaba dosimetricinvestigationofsixru106eyeplaquesbyebt3radiochromicfilmsandmontecarlosimulation AT ghassemifariba dosimetricinvestigationofsixru106eyeplaquesbyebt3radiochromicfilmsandmontecarlosimulation AT rafiepourpayman dosimetricinvestigationofsixru106eyeplaquesbyebt3radiochromicfilmsandmontecarlosimulation AT zeinalireyhaneh dosimetricinvestigationofsixru106eyeplaquesbyebt3radiochromicfilmsandmontecarlosimulation AT riaziesfahanihamid dosimetricinvestigationofsixru106eyeplaquesbyebt3radiochromicfilmsandmontecarlosimulation AT beikiardakaniakbar dosimetricinvestigationofsixru106eyeplaquesbyebt3radiochromicfilmsandmontecarlosimulation |