Cargando…

Postcolumn Infusion as a Quality Control Tool for LC-MS-Based Analysis

[Image: see text] Postcolumn infusion has been widely used to study the matrix effect of analytical methods based on liquid chromatography coupled to mass spectrometry (LC-MS). Nevertheless, this methodology is usually only applied during a method development or validation. With this application not...

Descripción completa

Detalles Bibliográficos
Autores principales: González, Oskar, Dubbelman, Anne-Charlotte, Hankemeier, Thomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2022
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10443037/
https://www.ncbi.nlm.nih.gov/pubmed/35483670
http://dx.doi.org/10.1021/jasms.2c00022
Descripción
Sumario:[Image: see text] Postcolumn infusion has been widely used to study the matrix effect of analytical methods based on liquid chromatography coupled to mass spectrometry (LC-MS). Nevertheless, this methodology is usually only applied during a method development or validation. With this application note, we aim to demonstrate that the continuous use of postcolumn infusion can be also a very useful tool to monitor the quality of LC-MS analyses and easily detect flaws in the analytical method performance. Here we propose a protocol that can be transferred to other LC-MS platforms, and we show some real situations in bioanalysis in which postcolumn infusion proved to be extremely helpful in, for example, the evaluation of a sample treatment or the detection of unexpected sources of the matrix effect.