Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis

PURPOSE: There is no specific literature on the best implantation position of the Femoral Neck System (FNS) for treating Pauwels type III femoral neck fracture in young adults. METHODS: Use finite-element analysis to compare the mechanical properties of implantation positions: FNS in the central pos...

Descripción completa

Detalles Bibliográficos
Autores principales: Kuang, Ximu, Jian, Guojian, Xie, Desheng, Chen, Xiaolin, Liu, Haoyuan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10449665/
https://www.ncbi.nlm.nih.gov/pubmed/36964769
http://dx.doi.org/10.1007/s00068-023-02253-5
_version_ 1785095009375092736
author Kuang, Ximu
Jian, Guojian
Xie, Desheng
Chen, Xiaolin
Liu, Haoyuan
author_facet Kuang, Ximu
Jian, Guojian
Xie, Desheng
Chen, Xiaolin
Liu, Haoyuan
author_sort Kuang, Ximu
collection PubMed
description PURPOSE: There is no specific literature on the best implantation position of the Femoral Neck System (FNS) for treating Pauwels type III femoral neck fracture in young adults. METHODS: Use finite-element analysis to compare the mechanical properties of implantation positions: FNS in the central position, FNS in the low position, and FNS in the low position combined with cannulated screw (CS). The CT data of the femur were imported into the mimics20.0 to obtain the three-dimensional model of the femur; imported into geomagic2017 and SolidWorks 2017 for optimizations; models of FNS and CS are built on the basis of the device manuals. Grouping is as follows: FNS group, FNS-LOW group, and FNS-CS group. Assemble and import them into abaques6.14 for load application. The displacement distribution and von Mises Stress value of them were compared. RESULTS: On femoral stability and stress distribution, the FNS-CS group performs best, followed by the FNS-LOW group, and finally FNS group. The FNS-LOW group has an improvement over the FNS group but not by much. CONCLUSION: In operations, when the implantation position of the central guide wire is not at the center of the femoral neck but slightly lower, it is recommended not to adjust the wire repeatedly in pursuit of the center position; for femoral neck fractures that are extremely unstable at the fracture end or require revision, the insertion strategy of FNS in the low position combined with CS can be adopted to obtain better fixation effects.
format Online
Article
Text
id pubmed-10449665
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-104496652023-08-26 Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis Kuang, Ximu Jian, Guojian Xie, Desheng Chen, Xiaolin Liu, Haoyuan Eur J Trauma Emerg Surg Original Article PURPOSE: There is no specific literature on the best implantation position of the Femoral Neck System (FNS) for treating Pauwels type III femoral neck fracture in young adults. METHODS: Use finite-element analysis to compare the mechanical properties of implantation positions: FNS in the central position, FNS in the low position, and FNS in the low position combined with cannulated screw (CS). The CT data of the femur were imported into the mimics20.0 to obtain the three-dimensional model of the femur; imported into geomagic2017 and SolidWorks 2017 for optimizations; models of FNS and CS are built on the basis of the device manuals. Grouping is as follows: FNS group, FNS-LOW group, and FNS-CS group. Assemble and import them into abaques6.14 for load application. The displacement distribution and von Mises Stress value of them were compared. RESULTS: On femoral stability and stress distribution, the FNS-CS group performs best, followed by the FNS-LOW group, and finally FNS group. The FNS-LOW group has an improvement over the FNS group but not by much. CONCLUSION: In operations, when the implantation position of the central guide wire is not at the center of the femoral neck but slightly lower, it is recommended not to adjust the wire repeatedly in pursuit of the center position; for femoral neck fractures that are extremely unstable at the fracture end or require revision, the insertion strategy of FNS in the low position combined with CS can be adopted to obtain better fixation effects. Springer Berlin Heidelberg 2023-03-25 2023 /pmc/articles/PMC10449665/ /pubmed/36964769 http://dx.doi.org/10.1007/s00068-023-02253-5 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Kuang, Ximu
Jian, Guojian
Xie, Desheng
Chen, Xiaolin
Liu, Haoyuan
Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis
title Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis
title_full Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis
title_fullStr Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis
title_full_unstemmed Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis
title_short Choose the appropriate implantation position of the Femoral Neck System in the femoral neck: a finite-element analysis
title_sort choose the appropriate implantation position of the femoral neck system in the femoral neck: a finite-element analysis
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10449665/
https://www.ncbi.nlm.nih.gov/pubmed/36964769
http://dx.doi.org/10.1007/s00068-023-02253-5
work_keys_str_mv AT kuangximu choosetheappropriateimplantationpositionofthefemoralnecksysteminthefemoralneckafiniteelementanalysis
AT jianguojian choosetheappropriateimplantationpositionofthefemoralnecksysteminthefemoralneckafiniteelementanalysis
AT xiedesheng choosetheappropriateimplantationpositionofthefemoralnecksysteminthefemoralneckafiniteelementanalysis
AT chenxiaolin choosetheappropriateimplantationpositionofthefemoralnecksysteminthefemoralneckafiniteelementanalysis
AT liuhaoyuan choosetheappropriateimplantationpositionofthefemoralnecksysteminthefemoralneckafiniteelementanalysis