Cargando…

Design Optimization of a Phototherapy Extracorporeal Membrane Oxygenator for Treating Carbon Monoxide Poisoning

We designed a photo-ECMO device to speed up the rate of carbon monoxide (CO) removal by using visible light to dissociate CO from hemoglobin (Hb). Using computational fluid dynamics, fillets of different radii (5 cm and 10 cm) were applied to the square shape of a photo-ECMO device to reduce stagnan...

Descripción completa

Detalles Bibliográficos
Autores principales: Etim, Edidiong, Goulopoulos, Anastasia, Fischbach, Anna, Franco, Walfre
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10451272/
https://www.ncbi.nlm.nih.gov/pubmed/37627854
http://dx.doi.org/10.3390/bioengineering10080969