Diagnostic validation of smart wearable device embedded with single-lead electrocardiogram for arrhythmia detection
OBJECTIVE: To validate a single-lead electrocardiogram algorithm for identifying atrial fibrillation, atrial premature beats, ventricular premature beats, and sinus rhythm. METHODS: A total of 656 subjects aged 19 to 94 years were enrolled. Participants were simultaneously tested with a wristwatch (...
Autores principales: | Niu, Yonghong, Wang, Hao, Wang, Hong, Zhang, Hui, Jin, Zhigeng, Guo, Yutao |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10475230/ https://www.ncbi.nlm.nih.gov/pubmed/37667685 http://dx.doi.org/10.1177/20552076231198682 |