Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease

OBJECTIVE: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensi...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Chul, Song, Jun Hyeong, Kim, Seung Hyoun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Academy of Rehabilitation Medicine 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10475817/
https://www.ncbi.nlm.nih.gov/pubmed/37536665
http://dx.doi.org/10.5535/arm.23019
_version_ 1785100798739349504
author Kim, Chul
Song, Jun Hyeong
Kim, Seung Hyoun
author_facet Kim, Chul
Song, Jun Hyeong
Kim, Seung Hyoun
author_sort Kim, Chul
collection PubMed
description OBJECTIVE: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensities. METHODS: CPX was performed for patients with coronary artery disease (CAD). Twelve lead electrocardiograph (ECG) was the gold standard and Apple watch 7 (AW7), Galaxy watch 4 (GW4) and Bio Patch Mobicare 200 (MC200) were applied for comparison. Paired absolute difference (PAD), mean absolute percentage error (MAPE) and intraclass correlation coefficient (ICC) were evaluated for each device. RESULTS: Forty-four participants with CAD were included. All the devices showed MAPE under 2% and ICC above 0.9 in rest, exercise and recovery phases (MC200=0.999, GW4=0.997, AW7=0.998). When comparing exercise and recovery phase, PAD of MC200 and AW7 in recovery phase were significantly bigger than PAD of exercise phase (p<0.05). Although not significant, PAD of GW4 tended to be bigger in recovery phase, too. Also, when stratified by HR 20, ICC of all the devices were highest under HR of 100, and ICC decreased as HR increased. However, except for ICC of GW4 at HR above 160 (=0.867), all ICCs exceeded 0.9 indicating excellent accuracy. CONCLUSION: The HR measurement of the devices validated in this study shows a high concordance with the ECG device, so CAD patients may benefit from the devices during high-intensity exercise under conditions where HR is measured reliably.
format Online
Article
Text
id pubmed-10475817
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Korean Academy of Rehabilitation Medicine
record_format MEDLINE/PubMed
spelling pubmed-104758172023-09-05 Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease Kim, Chul Song, Jun Hyeong Kim, Seung Hyoun Ann Rehabil Med Original Article OBJECTIVE: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensities. METHODS: CPX was performed for patients with coronary artery disease (CAD). Twelve lead electrocardiograph (ECG) was the gold standard and Apple watch 7 (AW7), Galaxy watch 4 (GW4) and Bio Patch Mobicare 200 (MC200) were applied for comparison. Paired absolute difference (PAD), mean absolute percentage error (MAPE) and intraclass correlation coefficient (ICC) were evaluated for each device. RESULTS: Forty-four participants with CAD were included. All the devices showed MAPE under 2% and ICC above 0.9 in rest, exercise and recovery phases (MC200=0.999, GW4=0.997, AW7=0.998). When comparing exercise and recovery phase, PAD of MC200 and AW7 in recovery phase were significantly bigger than PAD of exercise phase (p<0.05). Although not significant, PAD of GW4 tended to be bigger in recovery phase, too. Also, when stratified by HR 20, ICC of all the devices were highest under HR of 100, and ICC decreased as HR increased. However, except for ICC of GW4 at HR above 160 (=0.867), all ICCs exceeded 0.9 indicating excellent accuracy. CONCLUSION: The HR measurement of the devices validated in this study shows a high concordance with the ECG device, so CAD patients may benefit from the devices during high-intensity exercise under conditions where HR is measured reliably. Korean Academy of Rehabilitation Medicine 2023-08 2023-08-04 /pmc/articles/PMC10475817/ /pubmed/37536665 http://dx.doi.org/10.5535/arm.23019 Text en Copyright © 2023 by Korean Academy of Rehabilitation Medicine https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Kim, Chul
Song, Jun Hyeong
Kim, Seung Hyoun
Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
title Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
title_full Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
title_fullStr Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
title_full_unstemmed Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
title_short Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
title_sort validation of wearable digital devices for heart rate measurement during exercise test in patients with coronary artery disease
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10475817/
https://www.ncbi.nlm.nih.gov/pubmed/37536665
http://dx.doi.org/10.5535/arm.23019
work_keys_str_mv AT kimchul validationofwearabledigitaldevicesforheartratemeasurementduringexercisetestinpatientswithcoronaryarterydisease
AT songjunhyeong validationofwearabledigitaldevicesforheartratemeasurementduringexercisetestinpatientswithcoronaryarterydisease
AT kimseunghyoun validationofwearabledigitaldevicesforheartratemeasurementduringexercisetestinpatientswithcoronaryarterydisease