Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease
OBJECTIVE: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensi...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Academy of Rehabilitation Medicine
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10475817/ https://www.ncbi.nlm.nih.gov/pubmed/37536665 http://dx.doi.org/10.5535/arm.23019 |
_version_ | 1785100798739349504 |
---|---|
author | Kim, Chul Song, Jun Hyeong Kim, Seung Hyoun |
author_facet | Kim, Chul Song, Jun Hyeong Kim, Seung Hyoun |
author_sort | Kim, Chul |
collection | PubMed |
description | OBJECTIVE: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensities. METHODS: CPX was performed for patients with coronary artery disease (CAD). Twelve lead electrocardiograph (ECG) was the gold standard and Apple watch 7 (AW7), Galaxy watch 4 (GW4) and Bio Patch Mobicare 200 (MC200) were applied for comparison. Paired absolute difference (PAD), mean absolute percentage error (MAPE) and intraclass correlation coefficient (ICC) were evaluated for each device. RESULTS: Forty-four participants with CAD were included. All the devices showed MAPE under 2% and ICC above 0.9 in rest, exercise and recovery phases (MC200=0.999, GW4=0.997, AW7=0.998). When comparing exercise and recovery phase, PAD of MC200 and AW7 in recovery phase were significantly bigger than PAD of exercise phase (p<0.05). Although not significant, PAD of GW4 tended to be bigger in recovery phase, too. Also, when stratified by HR 20, ICC of all the devices were highest under HR of 100, and ICC decreased as HR increased. However, except for ICC of GW4 at HR above 160 (=0.867), all ICCs exceeded 0.9 indicating excellent accuracy. CONCLUSION: The HR measurement of the devices validated in this study shows a high concordance with the ECG device, so CAD patients may benefit from the devices during high-intensity exercise under conditions where HR is measured reliably. |
format | Online Article Text |
id | pubmed-10475817 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Korean Academy of Rehabilitation Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-104758172023-09-05 Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease Kim, Chul Song, Jun Hyeong Kim, Seung Hyoun Ann Rehabil Med Original Article OBJECTIVE: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensities. METHODS: CPX was performed for patients with coronary artery disease (CAD). Twelve lead electrocardiograph (ECG) was the gold standard and Apple watch 7 (AW7), Galaxy watch 4 (GW4) and Bio Patch Mobicare 200 (MC200) were applied for comparison. Paired absolute difference (PAD), mean absolute percentage error (MAPE) and intraclass correlation coefficient (ICC) were evaluated for each device. RESULTS: Forty-four participants with CAD were included. All the devices showed MAPE under 2% and ICC above 0.9 in rest, exercise and recovery phases (MC200=0.999, GW4=0.997, AW7=0.998). When comparing exercise and recovery phase, PAD of MC200 and AW7 in recovery phase were significantly bigger than PAD of exercise phase (p<0.05). Although not significant, PAD of GW4 tended to be bigger in recovery phase, too. Also, when stratified by HR 20, ICC of all the devices were highest under HR of 100, and ICC decreased as HR increased. However, except for ICC of GW4 at HR above 160 (=0.867), all ICCs exceeded 0.9 indicating excellent accuracy. CONCLUSION: The HR measurement of the devices validated in this study shows a high concordance with the ECG device, so CAD patients may benefit from the devices during high-intensity exercise under conditions where HR is measured reliably. Korean Academy of Rehabilitation Medicine 2023-08 2023-08-04 /pmc/articles/PMC10475817/ /pubmed/37536665 http://dx.doi.org/10.5535/arm.23019 Text en Copyright © 2023 by Korean Academy of Rehabilitation Medicine https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Kim, Chul Song, Jun Hyeong Kim, Seung Hyoun Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease |
title | Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease |
title_full | Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease |
title_fullStr | Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease |
title_full_unstemmed | Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease |
title_short | Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease |
title_sort | validation of wearable digital devices for heart rate measurement during exercise test in patients with coronary artery disease |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10475817/ https://www.ncbi.nlm.nih.gov/pubmed/37536665 http://dx.doi.org/10.5535/arm.23019 |
work_keys_str_mv | AT kimchul validationofwearabledigitaldevicesforheartratemeasurementduringexercisetestinpatientswithcoronaryarterydisease AT songjunhyeong validationofwearabledigitaldevicesforheartratemeasurementduringexercisetestinpatientswithcoronaryarterydisease AT kimseunghyoun validationofwearabledigitaldevicesforheartratemeasurementduringexercisetestinpatientswithcoronaryarterydisease |