Inflow from a Cardiopulmonary Assist System to the Pulmonary Artery and Its Implications for Local Hemodynamics—a Computational Fluid Dynamics Study
When returning blood to the pulmonary artery (PA), the inflow jet interferes with local hemodynamics. We investigated the consequences for several connection scenarios using transient computational fluid dynamics simulations. The PA was derived from CT data. Three aspects were varied: graft flow rat...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10480287/ https://www.ncbi.nlm.nih.gov/pubmed/36662482 http://dx.doi.org/10.1007/s12265-022-10349-3 |
Sumario: | When returning blood to the pulmonary artery (PA), the inflow jet interferes with local hemodynamics. We investigated the consequences for several connection scenarios using transient computational fluid dynamics simulations. The PA was derived from CT data. Three aspects were varied: graft flow rate, anastomosis location, and inflow jet path length from anastomosis site to impingement on the PA wall. Lateral anastomosis locations caused abnormal flow distribution between the left and right PA. The central location provided near-physiological distribution but induced higher wall shear stress (WSS). All effects were most pronounced at high graft flows. A central location is beneficial regarding flow distribution, but the resulting high WSS might promote detachment of local thromboembolisms or influence the autonomic nervous innervation. Lateral locations, depending on jet path length, result in lower WSS at the cost of an unfavorable flow distribution that could promote pulmonary vasculature changes. Case-specific decisions and further research are necessary. GRAPHICAL ABSTRACT: [Image: see text] SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s12265-022-10349-3. |
---|