Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020)
This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10484188/ https://www.ncbi.nlm.nih.gov/pubmed/37434533 http://dx.doi.org/10.1097/MBP.0000000000000662 |
_version_ | 1785102535669841920 |
---|---|
author | Duan, Baojing He, Le Zhang, Jianling Fang, Lixiu Li, Gang |
author_facet | Duan, Baojing He, Le Zhang, Jianling Fang, Lixiu Li, Gang |
author_sort | Duan, Baojing |
collection | PubMed |
description | This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the same arm in 88 adults (female : male = 47 : 41) with a mean age of 56.85 years using a mercury sphygmomanometer (two observers) and a DBP-6279B device (one supervisor). The AAMI/ESH/ISO 81060-2 : 2018 and Amd.1 : 2020 universal standards for the validation of BP-measuring devices in adults and adolescents were followed. A total of 259 valid pairs of data were used in the analysis. According to Criterion 1, the mean difference of SBP between the test device (DBP-6279B) and the reference device (the mercury sphygmomanometer) was 0.75 mmHg, with a SD of 7.66 mmHg. The mean difference in DBP was 1.13 mmHg, with a SD of 6.14 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 0.85 mmHg, and the SD was 6.56 mmHg, which was less than 6.88 mmHg and met the requirements. The mean difference in DBP was 1.27 mmHg, and the SD was 5.42 mmHg, which was less than 6.82 mmHg and met the requirements. DBP-6279B fulfilled the requirements of the AAMI/ESH/ISO universal standard (ISO 81060-2 : 2018 + Amd.1 : 2020); hence, it can be recommended for both clinical and self/home BP measurement in adults and adolescents. |
format | Online Article Text |
id | pubmed-10484188 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-104841882023-09-08 Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) Duan, Baojing He, Le Zhang, Jianling Fang, Lixiu Li, Gang Blood Press Monit Device and Technology This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the same arm in 88 adults (female : male = 47 : 41) with a mean age of 56.85 years using a mercury sphygmomanometer (two observers) and a DBP-6279B device (one supervisor). The AAMI/ESH/ISO 81060-2 : 2018 and Amd.1 : 2020 universal standards for the validation of BP-measuring devices in adults and adolescents were followed. A total of 259 valid pairs of data were used in the analysis. According to Criterion 1, the mean difference of SBP between the test device (DBP-6279B) and the reference device (the mercury sphygmomanometer) was 0.75 mmHg, with a SD of 7.66 mmHg. The mean difference in DBP was 1.13 mmHg, with a SD of 6.14 mmHg. The mean difference of both SBP and DBP was less than 5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to Criterion 2, the mean difference of SBP between the test device and the reference device was 0.85 mmHg, and the SD was 6.56 mmHg, which was less than 6.88 mmHg and met the requirements. The mean difference in DBP was 1.27 mmHg, and the SD was 5.42 mmHg, which was less than 6.82 mmHg and met the requirements. DBP-6279B fulfilled the requirements of the AAMI/ESH/ISO universal standard (ISO 81060-2 : 2018 + Amd.1 : 2020); hence, it can be recommended for both clinical and self/home BP measurement in adults and adolescents. Lippincott Williams & Wilkins 2023-10 2023-07-12 /pmc/articles/PMC10484188/ /pubmed/37434533 http://dx.doi.org/10.1097/MBP.0000000000000662 Text en Copyright © 2023 The Author(s). Published by Wolters Kluwer Health, Inc. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND) (https://creativecommons.org/licenses/by-nc-nd/4.0/) , where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. |
spellingShingle | Device and Technology Duan, Baojing He, Le Zhang, Jianling Fang, Lixiu Li, Gang Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) |
title | Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) |
title_full | Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) |
title_fullStr | Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) |
title_full_unstemmed | Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) |
title_short | Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020) |
title_sort | validation of the joytech dbp-6279b blood pressure monitor in adults and adolescents according to the aami/esh/iso universal standard(iso 81060-2:2018 + amd.1:2020) |
topic | Device and Technology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10484188/ https://www.ncbi.nlm.nih.gov/pubmed/37434533 http://dx.doi.org/10.1097/MBP.0000000000000662 |
work_keys_str_mv | AT duanbaojing validationofthejoytechdbp6279bbloodpressuremonitorinadultsandadolescentsaccordingtotheaamieshisouniversalstandardiso8106022018amd12020 AT hele validationofthejoytechdbp6279bbloodpressuremonitorinadultsandadolescentsaccordingtotheaamieshisouniversalstandardiso8106022018amd12020 AT zhangjianling validationofthejoytechdbp6279bbloodpressuremonitorinadultsandadolescentsaccordingtotheaamieshisouniversalstandardiso8106022018amd12020 AT fanglixiu validationofthejoytechdbp6279bbloodpressuremonitorinadultsandadolescentsaccordingtotheaamieshisouniversalstandardiso8106022018amd12020 AT ligang validationofthejoytechdbp6279bbloodpressuremonitorinadultsandadolescentsaccordingtotheaamieshisouniversalstandardiso8106022018amd12020 |