3D-printed dual holder system for simultaneous rat PET scanning: design and influence on quantification
BACKGROUND: The low throughout of small animal positron emission tomography (PET) images acquisitions represents a substantial limitation. The aims of this study were (i) to design a low-cost support for simultaneous dynamic PET scanning of two lying rats and (ii) to study its impact on brain image...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10485197/ https://www.ncbi.nlm.nih.gov/pubmed/37676475 http://dx.doi.org/10.1186/s13550-023-01027-9 |
Sumario: | BACKGROUND: The low throughout of small animal positron emission tomography (PET) images acquisitions represents a substantial limitation. The aims of this study were (i) to design a low-cost support for simultaneous dynamic PET scanning of two lying rats and (ii) to study its impact on brain image quantification. RESULTS: Accuracy of concentration measurement was 5.5% for one phantom in the field of view, and 5.7% for two phantoms measured simultaneously. Ratio concentration between phantoms showed an error of 6.7% ± 5.1% for Solo upper position, 6.7% ± 3.7% for Solo lower position, 5.9% ± 4.3% for Duo upper position, and 7.4% ± 6% for Duo lower position 6.7% for separated measures, and 6.6% for simultaneous measures. In vivo distribution profiles showed no difference between solo and duo uptakes. Region of Interest quantification in the whole brain showed 4.4% variability solo and 3.5% duo. The quantified test–retest bias was 8% in solo and 5% in duo, and the Intraclass Correlation Coefficient was comparable in solo and duo (0.969 vs. 0.966). CONCLUSIONS: Our results showed that simultaneous scans of two rats in INVEON do not affect quantification. The dual support system will allow us to reduce protocol costs and duration. |
---|