The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri

BACKGROUND: Shoulder arthroplasty is becoming increasingly common. With evolving implant designs, multiple humeral stem options exist for the surgeon to choose from. New stemless and short-stem systems are modular, remove less native bone stock, and better adapt to patient anatomy. It has been sugge...

Descripción completa

Detalles Bibliográficos
Autores principales: Ryan, Weston K., Vander Voort, Wyatt D., Saad, Maarouf A., Wu, Edward, Garcia-Nolen, Tanya C., Bayne, Christopher O., Szabo, Robert M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10499865/
https://www.ncbi.nlm.nih.gov/pubmed/37719832
http://dx.doi.org/10.1016/j.jseint.2023.04.011
_version_ 1785105801868738560
author Ryan, Weston K.
Vander Voort, Wyatt D.
Saad, Maarouf A.
Wu, Edward
Garcia-Nolen, Tanya C.
Bayne, Christopher O.
Szabo, Robert M.
author_facet Ryan, Weston K.
Vander Voort, Wyatt D.
Saad, Maarouf A.
Wu, Edward
Garcia-Nolen, Tanya C.
Bayne, Christopher O.
Szabo, Robert M.
author_sort Ryan, Weston K.
collection PubMed
description BACKGROUND: Shoulder arthroplasty is becoming increasingly common. With evolving implant designs, multiple humeral stem options exist for the surgeon to choose from. New stemless and short-stem systems are modular, remove less native bone stock, and better adapt to patient anatomy. It has been suggested that shorter stem implants may be protective against periprosthetic fracture; however, this has not been mechanistically evaluated. Therefore, this study aimed to biomechanically test synthetic humeri with long-stem, short-stem, and stemless arthroplasty components in a torsional manner to evaluate their response to loading and characterize failure. METHODS: Twenty-four synthetic humeri were implanted with long stem, short stem, or stemless uncemented prosthesis, 8 in each group. Humeri were mounted in a custom testing jig with a morse taper interfacing with a mechanical testing system. After a 20N axial force, specimens were torsionally loaded to failure at 15 degrees/sec, with 50 Hz collection. Torque vs. rotation curves were generated for each specimen, and stiffness, yield, ultimate strength, and failure load were measured. ANOVA and post hoc pairwise comparisons were used to assess effect of stem type on mechanical test variable. The association of the stem type with fracture type was analyzed by a Fisher’s Exact test. Statistical significance was set at P < .05. RESULTS: During torsional loading, long-stem implants were significantly stiffer than short or stemless implants. The angle of implant yielding was similar across stem designs; however, stemless implants had a lower yield torque. This correlated with a decreased yield energy in stemless compared to short stems as well. Maximum torque and failure torque was also significantly higher in short-stem and long-stem implants compared to stemless. DISCUSSION: Periprosthetic fractures in shoulder arthroplasty are a concern in low-energy trauma, and stem design likely plays a significant role in early implant-bone failure. Our results suggest stemless implants under torsional load fail at lower stress and are less stiff than stemmed implants. The failure mechanism of stemless implants through metaphyseal cancellous bone emphasizes the effect bone quality has on implant fixation. There is likely a balance of torsional stability to survive physiologic loads while minimizing diaphyseal stress and risk of diaphyseal periprosthetic fracture. This combined with revision and fixation options represent decisions the surgeon is faced with when performing shoulder arthroplasty.
format Online
Article
Text
id pubmed-10499865
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-104998652023-09-15 The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri Ryan, Weston K. Vander Voort, Wyatt D. Saad, Maarouf A. Wu, Edward Garcia-Nolen, Tanya C. Bayne, Christopher O. Szabo, Robert M. JSES Int Shoulder BACKGROUND: Shoulder arthroplasty is becoming increasingly common. With evolving implant designs, multiple humeral stem options exist for the surgeon to choose from. New stemless and short-stem systems are modular, remove less native bone stock, and better adapt to patient anatomy. It has been suggested that shorter stem implants may be protective against periprosthetic fracture; however, this has not been mechanistically evaluated. Therefore, this study aimed to biomechanically test synthetic humeri with long-stem, short-stem, and stemless arthroplasty components in a torsional manner to evaluate their response to loading and characterize failure. METHODS: Twenty-four synthetic humeri were implanted with long stem, short stem, or stemless uncemented prosthesis, 8 in each group. Humeri were mounted in a custom testing jig with a morse taper interfacing with a mechanical testing system. After a 20N axial force, specimens were torsionally loaded to failure at 15 degrees/sec, with 50 Hz collection. Torque vs. rotation curves were generated for each specimen, and stiffness, yield, ultimate strength, and failure load were measured. ANOVA and post hoc pairwise comparisons were used to assess effect of stem type on mechanical test variable. The association of the stem type with fracture type was analyzed by a Fisher’s Exact test. Statistical significance was set at P < .05. RESULTS: During torsional loading, long-stem implants were significantly stiffer than short or stemless implants. The angle of implant yielding was similar across stem designs; however, stemless implants had a lower yield torque. This correlated with a decreased yield energy in stemless compared to short stems as well. Maximum torque and failure torque was also significantly higher in short-stem and long-stem implants compared to stemless. DISCUSSION: Periprosthetic fractures in shoulder arthroplasty are a concern in low-energy trauma, and stem design likely plays a significant role in early implant-bone failure. Our results suggest stemless implants under torsional load fail at lower stress and are less stiff than stemmed implants. The failure mechanism of stemless implants through metaphyseal cancellous bone emphasizes the effect bone quality has on implant fixation. There is likely a balance of torsional stability to survive physiologic loads while minimizing diaphyseal stress and risk of diaphyseal periprosthetic fracture. This combined with revision and fixation options represent decisions the surgeon is faced with when performing shoulder arthroplasty. Elsevier 2023-05-24 /pmc/articles/PMC10499865/ /pubmed/37719832 http://dx.doi.org/10.1016/j.jseint.2023.04.011 Text en https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Shoulder
Ryan, Weston K.
Vander Voort, Wyatt D.
Saad, Maarouf A.
Wu, Edward
Garcia-Nolen, Tanya C.
Bayne, Christopher O.
Szabo, Robert M.
The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri
title The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri
title_full The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri
title_fullStr The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri
title_full_unstemmed The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri
title_short The effect of shoulder prosthesis stem length on failure due to torsional loading. A biomechanical study in composite humeri
title_sort effect of shoulder prosthesis stem length on failure due to torsional loading. a biomechanical study in composite humeri
topic Shoulder
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10499865/
https://www.ncbi.nlm.nih.gov/pubmed/37719832
http://dx.doi.org/10.1016/j.jseint.2023.04.011
work_keys_str_mv AT ryanwestonk theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT vandervoortwyattd theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT saadmaaroufa theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT wuedward theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT garcianolentanyac theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT baynechristophero theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT szaborobertm theeffectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT ryanwestonk effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT vandervoortwyattd effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT saadmaaroufa effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT wuedward effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT garcianolentanyac effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT baynechristophero effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri
AT szaborobertm effectofshoulderprosthesisstemlengthonfailureduetotorsionalloadingabiomechanicalstudyincompositehumeri