Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry

[Image: see text] Accurate quantitative analysis in liquid chromatography–mass spectrometry (LC–MS) benefits from calibration curves generated in the same matrix as the study sample. In the case of endogenous compound quantification, as no blank matrix exists, the multitargeted internal calibration...

Descripción completa

Detalles Bibliográficos
Autores principales: Visconti, Gioele, de Figueiredo, Miguel, Strassel, Oriane, Boccard, Julien, Vuilleumier, Nicolas, Jaques, David, Ponte, Belén, Rudaz, Serge
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2023
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10500547/
https://www.ncbi.nlm.nih.gov/pubmed/37655548
http://dx.doi.org/10.1021/acs.analchem.3c02069
_version_ 1785105932375556096
author Visconti, Gioele
de Figueiredo, Miguel
Strassel, Oriane
Boccard, Julien
Vuilleumier, Nicolas
Jaques, David
Ponte, Belén
Rudaz, Serge
author_facet Visconti, Gioele
de Figueiredo, Miguel
Strassel, Oriane
Boccard, Julien
Vuilleumier, Nicolas
Jaques, David
Ponte, Belén
Rudaz, Serge
author_sort Visconti, Gioele
collection PubMed
description [Image: see text] Accurate quantitative analysis in liquid chromatography–mass spectrometry (LC–MS) benefits from calibration curves generated in the same matrix as the study sample. In the case of endogenous compound quantification, as no blank matrix exists, the multitargeted internal calibration (MTIC) is an attractive and straightforward approach to avoid the need for extensive matrix similarity evaluation. Its principle is to take advantage of stable isotope labeled (SIL) standards as internal calibrants to simultaneously quantify authentic analytes using a within sample calibration. An MTIC workflow was developed for the simultaneous quantification of metabolites related to chronic kidney disease (CKD) using a volumetric microsampling device to collect 20 μL of serum or plasma, followed by a single-step extraction with acetonitrile/water and liquid chromatography–tandem mass spectrometry (LC–MS/MS) analysis. Since a single concentration of internal calibrant is necessary to calculate the study sample concentration, the instrument response function was investigated to determine the best SIL concentration. After validation, the trueness of 16 endogenous analytes in authentic human serum ranged from 72.2 to 116.0%, the repeatability from 1.9 to 11.3%, and the intermediate precision ranged overall from 2.1 to 15.4%. The proposed approach was applied to plasma samples collected from healthy control participants and two patient groups diagnosed with CKD. Results confirmed substantial concentration differences between groups for several analytes, including indoxyl sulfate and cortisone, as well as metabolite enrichment in the kynurenine and indole pathways. Multitargeted methodologies represent a major step toward rapid and straightforward LC–MS/MS absolute quantification of endogenous biomarkers, which could change the paradigm of MS use in clinical laboratories.
format Online
Article
Text
id pubmed-10500547
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher American Chemical Society
record_format MEDLINE/PubMed
spelling pubmed-105005472023-09-15 Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry Visconti, Gioele de Figueiredo, Miguel Strassel, Oriane Boccard, Julien Vuilleumier, Nicolas Jaques, David Ponte, Belén Rudaz, Serge Anal Chem [Image: see text] Accurate quantitative analysis in liquid chromatography–mass spectrometry (LC–MS) benefits from calibration curves generated in the same matrix as the study sample. In the case of endogenous compound quantification, as no blank matrix exists, the multitargeted internal calibration (MTIC) is an attractive and straightforward approach to avoid the need for extensive matrix similarity evaluation. Its principle is to take advantage of stable isotope labeled (SIL) standards as internal calibrants to simultaneously quantify authentic analytes using a within sample calibration. An MTIC workflow was developed for the simultaneous quantification of metabolites related to chronic kidney disease (CKD) using a volumetric microsampling device to collect 20 μL of serum or plasma, followed by a single-step extraction with acetonitrile/water and liquid chromatography–tandem mass spectrometry (LC–MS/MS) analysis. Since a single concentration of internal calibrant is necessary to calculate the study sample concentration, the instrument response function was investigated to determine the best SIL concentration. After validation, the trueness of 16 endogenous analytes in authentic human serum ranged from 72.2 to 116.0%, the repeatability from 1.9 to 11.3%, and the intermediate precision ranged overall from 2.1 to 15.4%. The proposed approach was applied to plasma samples collected from healthy control participants and two patient groups diagnosed with CKD. Results confirmed substantial concentration differences between groups for several analytes, including indoxyl sulfate and cortisone, as well as metabolite enrichment in the kynurenine and indole pathways. Multitargeted methodologies represent a major step toward rapid and straightforward LC–MS/MS absolute quantification of endogenous biomarkers, which could change the paradigm of MS use in clinical laboratories. American Chemical Society 2023-09-01 /pmc/articles/PMC10500547/ /pubmed/37655548 http://dx.doi.org/10.1021/acs.analchem.3c02069 Text en © 2023 The Authors. Published by American Chemical Society https://creativecommons.org/licenses/by/4.0/Permits the broadest form of re-use including for commercial purposes, provided that author attribution and integrity are maintained (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Visconti, Gioele
de Figueiredo, Miguel
Strassel, Oriane
Boccard, Julien
Vuilleumier, Nicolas
Jaques, David
Ponte, Belén
Rudaz, Serge
Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry
title Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry
title_full Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry
title_fullStr Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry
title_full_unstemmed Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry
title_short Multitargeted Internal Calibration for the Quantification of Chronic Kidney Disease-Related Endogenous Metabolites Using Liquid Chromatography–Mass Spectrometry
title_sort multitargeted internal calibration for the quantification of chronic kidney disease-related endogenous metabolites using liquid chromatography–mass spectrometry
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10500547/
https://www.ncbi.nlm.nih.gov/pubmed/37655548
http://dx.doi.org/10.1021/acs.analchem.3c02069
work_keys_str_mv AT viscontigioele multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT defigueiredomiguel multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT strasseloriane multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT boccardjulien multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT vuilleumiernicolas multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT jaquesdavid multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT pontebelen multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry
AT rudazserge multitargetedinternalcalibrationforthequantificationofchronickidneydiseaserelatedendogenousmetabolitesusingliquidchromatographymassspectrometry