Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers

Background We aimed to compare bio-mechanical outcomes of short-length 75%-length uni-cortical screw (SL75UCS) and full-length 100%-length screws (FL100S) under axial compression (AXC) and torsional compression (TRC) in cadaveric distal radius volar plate model. Methodology A total of 20 wrists from...

Descripción completa

Detalles Bibliográficos
Autores principales: Pehlivan, Ali T, Kilinc, Bekir E, Oc, Yunus, Vezirhuyuk, Mustafa, Yamak, Fatih, Bozdag, Ergun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10501418/
https://www.ncbi.nlm.nih.gov/pubmed/37719510
http://dx.doi.org/10.7759/cureus.43522
_version_ 1785106112106725376
author Pehlivan, Ali T
Kilinc, Bekir E
Oc, Yunus
Vezirhuyuk, Mustafa
Yamak, Fatih
Bozdag, Ergun
author_facet Pehlivan, Ali T
Kilinc, Bekir E
Oc, Yunus
Vezirhuyuk, Mustafa
Yamak, Fatih
Bozdag, Ergun
author_sort Pehlivan, Ali T
collection PubMed
description Background We aimed to compare bio-mechanical outcomes of short-length 75%-length uni-cortical screw (SL75UCS) and full-length 100%-length screws (FL100S) under axial compression (AXC) and torsional compression (TRC) in cadaveric distal radius volar plate model. Methodology A total of 20 wrists from 10 fresh frozen cadavers were included. A 2.5 mm titanium alloy distal radius anatomical plate was placed to the distal radii in full anatomical position, just proximal to the watershed line. Three bi-cortical screws to the shaft of the radius, followed by uni-cortical drilling for distal screwing were placed. Measurement by pulling the drill once it reached the opposite cortex was applied. We selected the screw lengths such that they corresponded to the SL75UCS. In the same configuration for each of the cadavers, we delivered six screws from distal radius holes of the anatomical plate. An oscillating handsaw was used to create an extra-articular distal radius fracture model (AO 23-A3.2). We created a dorsal AP model by performing a 1-cm wedge osteotomy from the dorsal aspect. Complete separation of the volar cortex was achieved. Potting was performed by embedding the shaft of the prepared radius into the polyurethane medium. We placed aluminum apparatus into the distal end to ensure applying of AXC and TRC in bio-mechanistic tests. Results No statistically significant difference of stiffness between the SL75UCS and FL100S both under AXC (p=0.88) and TRC (p=0.82). SL75UCS and FL100S groups did not differ in elastic limit under AXC (p=0.71) and TRC (p=0.71). Maximal force on SL75UCS and FL100S groups were also similar under both AXC (p=0.71) and TRC (p=0.50). Conclusions Our study findings suggest that drilling the dorsal cortex may not be necessary in the management of distal radius fractures. Instead, utilizing SL75UCS could serve as a viable alternative. This approach offers potential advantages in reducing the risk of extensor tendon complications associated with drilling or screw protrusion. It is a safe method under torsional load to avoid drilling of the dorsal cortex and SL75UCS could be performed in order to prevent from extensor tendon complications secondary to drilling or screw protrusion.
format Online
Article
Text
id pubmed-10501418
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-105014182023-09-15 Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers Pehlivan, Ali T Kilinc, Bekir E Oc, Yunus Vezirhuyuk, Mustafa Yamak, Fatih Bozdag, Ergun Cureus Orthopedics Background We aimed to compare bio-mechanical outcomes of short-length 75%-length uni-cortical screw (SL75UCS) and full-length 100%-length screws (FL100S) under axial compression (AXC) and torsional compression (TRC) in cadaveric distal radius volar plate model. Methodology A total of 20 wrists from 10 fresh frozen cadavers were included. A 2.5 mm titanium alloy distal radius anatomical plate was placed to the distal radii in full anatomical position, just proximal to the watershed line. Three bi-cortical screws to the shaft of the radius, followed by uni-cortical drilling for distal screwing were placed. Measurement by pulling the drill once it reached the opposite cortex was applied. We selected the screw lengths such that they corresponded to the SL75UCS. In the same configuration for each of the cadavers, we delivered six screws from distal radius holes of the anatomical plate. An oscillating handsaw was used to create an extra-articular distal radius fracture model (AO 23-A3.2). We created a dorsal AP model by performing a 1-cm wedge osteotomy from the dorsal aspect. Complete separation of the volar cortex was achieved. Potting was performed by embedding the shaft of the prepared radius into the polyurethane medium. We placed aluminum apparatus into the distal end to ensure applying of AXC and TRC in bio-mechanistic tests. Results No statistically significant difference of stiffness between the SL75UCS and FL100S both under AXC (p=0.88) and TRC (p=0.82). SL75UCS and FL100S groups did not differ in elastic limit under AXC (p=0.71) and TRC (p=0.71). Maximal force on SL75UCS and FL100S groups were also similar under both AXC (p=0.71) and TRC (p=0.50). Conclusions Our study findings suggest that drilling the dorsal cortex may not be necessary in the management of distal radius fractures. Instead, utilizing SL75UCS could serve as a viable alternative. This approach offers potential advantages in reducing the risk of extensor tendon complications associated with drilling or screw protrusion. It is a safe method under torsional load to avoid drilling of the dorsal cortex and SL75UCS could be performed in order to prevent from extensor tendon complications secondary to drilling or screw protrusion. Cureus 2023-08-15 /pmc/articles/PMC10501418/ /pubmed/37719510 http://dx.doi.org/10.7759/cureus.43522 Text en Copyright © 2023, Pehlivan et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Orthopedics
Pehlivan, Ali T
Kilinc, Bekir E
Oc, Yunus
Vezirhuyuk, Mustafa
Yamak, Fatih
Bozdag, Ergun
Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers
title Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers
title_full Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers
title_fullStr Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers
title_full_unstemmed Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers
title_short Evaluation of Unicortical Locking Screw Placement for Torsional Loads in Distal Radius Fractures: A Biomechanical Study in Cadavers
title_sort evaluation of unicortical locking screw placement for torsional loads in distal radius fractures: a biomechanical study in cadavers
topic Orthopedics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10501418/
https://www.ncbi.nlm.nih.gov/pubmed/37719510
http://dx.doi.org/10.7759/cureus.43522
work_keys_str_mv AT pehlivanalit evaluationofunicorticallockingscrewplacementfortorsionalloadsindistalradiusfracturesabiomechanicalstudyincadavers
AT kilincbekire evaluationofunicorticallockingscrewplacementfortorsionalloadsindistalradiusfracturesabiomechanicalstudyincadavers
AT ocyunus evaluationofunicorticallockingscrewplacementfortorsionalloadsindistalradiusfracturesabiomechanicalstudyincadavers
AT vezirhuyukmustafa evaluationofunicorticallockingscrewplacementfortorsionalloadsindistalradiusfracturesabiomechanicalstudyincadavers
AT yamakfatih evaluationofunicorticallockingscrewplacementfortorsionalloadsindistalradiusfracturesabiomechanicalstudyincadavers
AT bozdagergun evaluationofunicorticallockingscrewplacementfortorsionalloadsindistalradiusfracturesabiomechanicalstudyincadavers