The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients

OBJECTIVE: To investigate the role of segmental bioelectrical impedance technique (SBIA) in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients. METHOD: We selected the patients at the Department of Nephrology of the First Affiliated Hospital of Zhengzhou Unive...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Lifen, Zhang, Bei, Luo, Jing, Li, Zhengyan, Liu, Dong, Dou, Yanna, Yang, Lin, Xiao, Jing, Zhao, Zhanzheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Taylor & Francis 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10512755/
https://www.ncbi.nlm.nih.gov/pubmed/37724525
http://dx.doi.org/10.1080/0886022X.2023.2255678
_version_ 1785108430589001728
author Liu, Lifen
Zhang, Bei
Luo, Jing
Li, Zhengyan
Liu, Dong
Dou, Yanna
Yang, Lin
Xiao, Jing
Zhao, Zhanzheng
author_facet Liu, Lifen
Zhang, Bei
Luo, Jing
Li, Zhengyan
Liu, Dong
Dou, Yanna
Yang, Lin
Xiao, Jing
Zhao, Zhanzheng
author_sort Liu, Lifen
collection PubMed
description OBJECTIVE: To investigate the role of segmental bioelectrical impedance technique (SBIA) in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients. METHOD: We selected the patients at the Department of Nephrology of the First Affiliated Hospital of Zhengzhou University and measured the segmental bioelectrical impedance by a German Fresenius body composition analyzer (the Fresenius whole body composition measurement (BCM) machine was used as a segmental machine in this study). An alternating current (5 kHz, 0.05–0.7 mA) was continuously released during the measurement. The released current penetrated the peritoneal cavity on both sides of the body, from which the segmental resistance at a frequency of 5 kHz was obtained from the multifrequency data (R5/Ω). Baseline BIA measurements were initiated after the patient entered the supine position for 5–10 min, then dialysate was instilled into the peritoneal cavity. BIA measurements were performed at 10-min intervals during the retention of dialysate in the abdomen and finally ended when dialysate drainage was complete. Real-time intraperitoneal volume estimated by SBIA (IPV(SBIA))and ultrafiltration volume estimated by SBIA(UFV(SBIA)) was calculated. At the same time, the actual ultrafiltration volume at the end of peritoneal dialysis was weighed and measured (UFV(MEA)). RESULTS: A total of 30 patients were included in the study, 9 patients withdrew from the study due to subjective factors during the measurement process, and 21 patients completed the study. The correlation coefficient R(2) of UFV(SBIA) and UFV(MEA) was 0.21 (p < 0.05). Bland-Altman analysis showed that the bias of UFV(SBIA) to the actual UFV(MEA) was 0.12 L, and the 95% agreement limit was between −0.5 L and 0.74 L, which confirmed that UFV(SBIA) measured by electrical impedance method and UFV(MEA) measured by weighing method were in good agreement. The time required to reach the maximum ultrafiltration volume (UFV(SBIA)) was 108 ± 68 min, and the mean value of the maximum ultrafiltration volume (Max UFV(SBIA)) was 1.16 ± 0.60 L. CONCLUSION: The segmental bioelectrical impedance technique can be used to assess the intraperitoneal ultrafiltration volume of peritoneal dialysis patients in real-time and effectively. This method may guide the dialysis fluid retention time and the maximum ultrafiltration volume in PD patients.
format Online
Article
Text
id pubmed-10512755
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Taylor & Francis
record_format MEDLINE/PubMed
spelling pubmed-105127552023-09-22 The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients Liu, Lifen Zhang, Bei Luo, Jing Li, Zhengyan Liu, Dong Dou, Yanna Yang, Lin Xiao, Jing Zhao, Zhanzheng Ren Fail Research Article OBJECTIVE: To investigate the role of segmental bioelectrical impedance technique (SBIA) in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients. METHOD: We selected the patients at the Department of Nephrology of the First Affiliated Hospital of Zhengzhou University and measured the segmental bioelectrical impedance by a German Fresenius body composition analyzer (the Fresenius whole body composition measurement (BCM) machine was used as a segmental machine in this study). An alternating current (5 kHz, 0.05–0.7 mA) was continuously released during the measurement. The released current penetrated the peritoneal cavity on both sides of the body, from which the segmental resistance at a frequency of 5 kHz was obtained from the multifrequency data (R5/Ω). Baseline BIA measurements were initiated after the patient entered the supine position for 5–10 min, then dialysate was instilled into the peritoneal cavity. BIA measurements were performed at 10-min intervals during the retention of dialysate in the abdomen and finally ended when dialysate drainage was complete. Real-time intraperitoneal volume estimated by SBIA (IPV(SBIA))and ultrafiltration volume estimated by SBIA(UFV(SBIA)) was calculated. At the same time, the actual ultrafiltration volume at the end of peritoneal dialysis was weighed and measured (UFV(MEA)). RESULTS: A total of 30 patients were included in the study, 9 patients withdrew from the study due to subjective factors during the measurement process, and 21 patients completed the study. The correlation coefficient R(2) of UFV(SBIA) and UFV(MEA) was 0.21 (p < 0.05). Bland-Altman analysis showed that the bias of UFV(SBIA) to the actual UFV(MEA) was 0.12 L, and the 95% agreement limit was between −0.5 L and 0.74 L, which confirmed that UFV(SBIA) measured by electrical impedance method and UFV(MEA) measured by weighing method were in good agreement. The time required to reach the maximum ultrafiltration volume (UFV(SBIA)) was 108 ± 68 min, and the mean value of the maximum ultrafiltration volume (Max UFV(SBIA)) was 1.16 ± 0.60 L. CONCLUSION: The segmental bioelectrical impedance technique can be used to assess the intraperitoneal ultrafiltration volume of peritoneal dialysis patients in real-time and effectively. This method may guide the dialysis fluid retention time and the maximum ultrafiltration volume in PD patients. Taylor & Francis 2023-09-19 /pmc/articles/PMC10512755/ /pubmed/37724525 http://dx.doi.org/10.1080/0886022X.2023.2255678 Text en © 2023 The Author(s). Published by Informa UK Limited, trading as Taylor & Francis Group https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. The terms on which this article has been published allow the posting of the Accepted Manuscript in a repository by the author(s) or with their consent.
spellingShingle Research Article
Liu, Lifen
Zhang, Bei
Luo, Jing
Li, Zhengyan
Liu, Dong
Dou, Yanna
Yang, Lin
Xiao, Jing
Zhao, Zhanzheng
The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
title The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
title_full The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
title_fullStr The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
title_full_unstemmed The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
title_short The role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
title_sort role of segmental bioelectrical impedance technique in the assessment of intraperitoneal ultrafiltration volume with peritoneal dialysis patients
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10512755/
https://www.ncbi.nlm.nih.gov/pubmed/37724525
http://dx.doi.org/10.1080/0886022X.2023.2255678
work_keys_str_mv AT liulifen theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT zhangbei theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT luojing theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT lizhengyan theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT liudong theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT douyanna theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT yanglin theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT xiaojing theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT zhaozhanzheng theroleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT liulifen roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT zhangbei roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT luojing roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT lizhengyan roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT liudong roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT douyanna roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT yanglin roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT xiaojing roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients
AT zhaozhanzheng roleofsegmentalbioelectricalimpedancetechniqueintheassessmentofintraperitonealultrafiltrationvolumewithperitonealdialysispatients