Real-world Accuracy of CGM in Inpatient Critical and Noncritical Care Settings at a Safety-Net Hospital

OBJECTIVE: We sought to determine real-world accuracy of inpatient continuous glucose monitoring (CGM) at multiple levels of acuity in a large safety-net hospital. RESEARCH DESIGN AND METHODS: We analyzed records from hospitalized patients on Dexcom G6 CGM, including clinical, point of care (POC), a...

Descripción completa

Detalles Bibliográficos
Autores principales: Finn, Erin, Schlichting, Lindsay, Grau, Laura, Douglas, Ivor S., Pereira, Rocio I.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Diabetes Association 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10516250/
https://www.ncbi.nlm.nih.gov/pubmed/37561954
http://dx.doi.org/10.2337/dc23-0089
Descripción
Sumario:OBJECTIVE: We sought to determine real-world accuracy of inpatient continuous glucose monitoring (CGM) at multiple levels of acuity in a large safety-net hospital. RESEARCH DESIGN AND METHODS: We analyzed records from hospitalized patients on Dexcom G6 CGM, including clinical, point of care (POC), and laboratory (Lab) glucose, and CGM data. POC/Lab values were matched to the closest timed CGM value. Encounters were divided into not critically ill (NCI) versus critically ill (CI). CGM accuracy was evaluated. RESULTS: Paired readings (2,744 POC-CGM; 3,705 Lab-CGM) were analyzed for 233 patients with 239 encounters (83 NCI, 156 CI). POC-CGM aggregated and average mean absolute relative differences (MARD) were 15.1% and 17.1%. Lab-CGM aggregated and average MARDs were 11.4% and 12.2%. Accuracy for POC-CGM and Lab-CGM was 96.5% and 99.1% in Clarke Error Grid zones A/B. CONCLUSIONS: Real-world accuracy of inpatient CGM is acceptable for NCI and CI patients. Further exploration of conditions associated with lower CGM accuracy in real-world settings is warranted.