A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement

BACKGROUND: Acrylic-based bone cement (polymethyl methacrylate [PMMA]) is a material commonly used in orthopaedic surgeries; however, during PMMA polymerization, a highly exothermic reaction occurs. The heat released in polymerization can damage nearby materials including poorly heat-resistant cross...

Descripción completa

Detalles Bibliográficos
Autores principales: Thompson, Zoe, Hothi, Harry, Brillantes, Jacqueline, Khoshbin, Amir, Atrey, Amit
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10517260/
https://www.ncbi.nlm.nih.gov/pubmed/37745962
http://dx.doi.org/10.1016/j.artd.2023.101184
_version_ 1785109285009620992
author Thompson, Zoe
Hothi, Harry
Brillantes, Jacqueline
Khoshbin, Amir
Atrey, Amit
author_facet Thompson, Zoe
Hothi, Harry
Brillantes, Jacqueline
Khoshbin, Amir
Atrey, Amit
author_sort Thompson, Zoe
collection PubMed
description BACKGROUND: Acrylic-based bone cement (polymethyl methacrylate [PMMA]) is a material commonly used in orthopaedic surgeries; however, during PMMA polymerization, a highly exothermic reaction occurs. The heat released in polymerization can damage nearby materials including poorly heat-resistant cross-linked polyethylene (XLPE). Both PMMA and XLPE are used in total hip arthroplasty and could interact during femoral stem fixation. We sought to determine if the exothermic polymerization of PMMA could alter the surface characteristics of XLPE acetabular liners. METHODS: Six XLPE liners were assigned to one of 4 experimental categories with varying volumes of PMMA applied in a manner that mimicked how the 2 materials would come into contact intraoperatively. Measurements were taken both pre- and post-intervention using a coordinate measuring machine for geometric and gravimetric analysis. Light microscopy was conducted postintervention to examine the surface for damage. RESULTS: Coordinate measuring machine measurements showed minimal gross deformation in all 6 liners, but there were isolated surface deposits in 4 of 6 liners. The average maximal surface deviations, when compared to the control, for liners exposed to 1 cc of cement, 2 cc of cement, or 1 cc of cement with a femoral head implant attached were 26.6 μm, 77.2 μm, and 26.4 μm, respectively. All but one liner showed an increase in volume following intervention when compared to the control. Subtle scratches were identified using light microscopy on all 6 liners. CONCLUSIONS: XLPE shows areas of isolated surface deformation in a dose-dependent manner but with minimal gross deformation after interacting with highly exothermic PMMA.
format Online
Article
Text
id pubmed-10517260
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-105172602023-09-24 A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement Thompson, Zoe Hothi, Harry Brillantes, Jacqueline Khoshbin, Amir Atrey, Amit Arthroplast Today Original Research BACKGROUND: Acrylic-based bone cement (polymethyl methacrylate [PMMA]) is a material commonly used in orthopaedic surgeries; however, during PMMA polymerization, a highly exothermic reaction occurs. The heat released in polymerization can damage nearby materials including poorly heat-resistant cross-linked polyethylene (XLPE). Both PMMA and XLPE are used in total hip arthroplasty and could interact during femoral stem fixation. We sought to determine if the exothermic polymerization of PMMA could alter the surface characteristics of XLPE acetabular liners. METHODS: Six XLPE liners were assigned to one of 4 experimental categories with varying volumes of PMMA applied in a manner that mimicked how the 2 materials would come into contact intraoperatively. Measurements were taken both pre- and post-intervention using a coordinate measuring machine for geometric and gravimetric analysis. Light microscopy was conducted postintervention to examine the surface for damage. RESULTS: Coordinate measuring machine measurements showed minimal gross deformation in all 6 liners, but there were isolated surface deposits in 4 of 6 liners. The average maximal surface deviations, when compared to the control, for liners exposed to 1 cc of cement, 2 cc of cement, or 1 cc of cement with a femoral head implant attached were 26.6 μm, 77.2 μm, and 26.4 μm, respectively. All but one liner showed an increase in volume following intervention when compared to the control. Subtle scratches were identified using light microscopy on all 6 liners. CONCLUSIONS: XLPE shows areas of isolated surface deformation in a dose-dependent manner but with minimal gross deformation after interacting with highly exothermic PMMA. Elsevier 2023-09-19 /pmc/articles/PMC10517260/ /pubmed/37745962 http://dx.doi.org/10.1016/j.artd.2023.101184 Text en © 2023 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Research
Thompson, Zoe
Hothi, Harry
Brillantes, Jacqueline
Khoshbin, Amir
Atrey, Amit
A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement
title A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement
title_full A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement
title_fullStr A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement
title_full_unstemmed A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement
title_short A Geometric Analysis of Polyethylene Liners Exposed to Acrylic-based Bone Cement
title_sort geometric analysis of polyethylene liners exposed to acrylic-based bone cement
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10517260/
https://www.ncbi.nlm.nih.gov/pubmed/37745962
http://dx.doi.org/10.1016/j.artd.2023.101184
work_keys_str_mv AT thompsonzoe ageometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT hothiharry ageometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT brillantesjacqueline ageometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT khoshbinamir ageometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT atreyamit ageometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT thompsonzoe geometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT hothiharry geometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT brillantesjacqueline geometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT khoshbinamir geometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement
AT atreyamit geometricanalysisofpolyethylenelinersexposedtoacrylicbasedbonecement