The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance

OBJECTIVES: The clinical importance of optimal post-repair mitral valve diastolic performance is increasingly being recognized. The haemodynamic effect of a partial annuloplasty band implantation, in comparison to a full ring, remains insufficiently explored. METHODS: Patients undergoing mitral valv...

Descripción completa

Detalles Bibliográficos
Autores principales: Tomšič, Anton, Sandoval, Elena, Meucci, Maria C, Nabeta, Takeru, Castella, Manuel, Muro, Anna, Klautz, Robert J M, Ajmone Marsan, Nina, Pereda, Daniel, Palmen, Meindert
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10517645/
https://www.ncbi.nlm.nih.gov/pubmed/37688566
http://dx.doi.org/10.1093/ejcts/ezad307
_version_ 1785109367835590656
author Tomšič, Anton
Sandoval, Elena
Meucci, Maria C
Nabeta, Takeru
Castella, Manuel
Muro, Anna
Klautz, Robert J M
Ajmone Marsan, Nina
Pereda, Daniel
Palmen, Meindert
author_facet Tomšič, Anton
Sandoval, Elena
Meucci, Maria C
Nabeta, Takeru
Castella, Manuel
Muro, Anna
Klautz, Robert J M
Ajmone Marsan, Nina
Pereda, Daniel
Palmen, Meindert
author_sort Tomšič, Anton
collection PubMed
description OBJECTIVES: The clinical importance of optimal post-repair mitral valve diastolic performance is increasingly being recognized. The haemodynamic effect of a partial annuloplasty band implantation, in comparison to a full ring, remains insufficiently explored. METHODS: Patients undergoing mitral valve repair for pure degenerative disease between 2011 and 2019 at 2 experienced heart valve centres were eligible for inclusion. Exclusion criteria were concomitant procedures other than tricuspid valve repair and ablation procedures for atrial fibrillation. Pre-discharge and follow-up echocardiograms (1–4 years after surgery) were analysed to assess haemodynamic mitral valve performance. RESULTS: Of 535 patients meeting the inclusion criteria, 364 (68.0%) patients underwent full annuloplasty ring and 171 (31.0%) partial band implantation. On predischarge echocardiogram, post-repair mitral valve gradient and area did not differ between groups [2.89 mmHg (IQR 2.26–3.72) vs 2.60 mmHg (IQR 1.91–3.55), P = 0.19 and 1.98 cm(2) (IQR 1.66–2.46) vs 2.03 cm(2) (IQR 1.55–3.06), P = 0.15]. However, multivariable linear regression analysis demonstrated band annuloplasty as a determinant of larger valve area (coefficient 0.467 cm(2), standard error 0.105, P < 0.001). On multivariable analysis, no significant impact on post-repair gradient was observed (–0.370 mmHg, standard error 0.167, P = 0.36). At follow-up, the differences between groups disappeared and multivariable regression analysis failed to demonstrate a significant impact of annuloplasty device type on mitral valve gradient (coefficient –0.095 mmHg, standard error 0.171, P = 1.00) or area (coefficient –0.085 cm(2), standard error 0.120, P = 1.00). These results were confirmed with a linear mixed model analysis. CONCLUSIONS: Partial band annuloplasty was related to an improved haemodynamic profile directly after valve repair for degenerative disease but the effect was short-lived. Our results suggest that the type of annuloplasty device has no durable impact on diastolic valve performance.
format Online
Article
Text
id pubmed-10517645
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-105176452023-09-24 The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance Tomšič, Anton Sandoval, Elena Meucci, Maria C Nabeta, Takeru Castella, Manuel Muro, Anna Klautz, Robert J M Ajmone Marsan, Nina Pereda, Daniel Palmen, Meindert Eur J Cardiothorac Surg Conventional Valve Operations OBJECTIVES: The clinical importance of optimal post-repair mitral valve diastolic performance is increasingly being recognized. The haemodynamic effect of a partial annuloplasty band implantation, in comparison to a full ring, remains insufficiently explored. METHODS: Patients undergoing mitral valve repair for pure degenerative disease between 2011 and 2019 at 2 experienced heart valve centres were eligible for inclusion. Exclusion criteria were concomitant procedures other than tricuspid valve repair and ablation procedures for atrial fibrillation. Pre-discharge and follow-up echocardiograms (1–4 years after surgery) were analysed to assess haemodynamic mitral valve performance. RESULTS: Of 535 patients meeting the inclusion criteria, 364 (68.0%) patients underwent full annuloplasty ring and 171 (31.0%) partial band implantation. On predischarge echocardiogram, post-repair mitral valve gradient and area did not differ between groups [2.89 mmHg (IQR 2.26–3.72) vs 2.60 mmHg (IQR 1.91–3.55), P = 0.19 and 1.98 cm(2) (IQR 1.66–2.46) vs 2.03 cm(2) (IQR 1.55–3.06), P = 0.15]. However, multivariable linear regression analysis demonstrated band annuloplasty as a determinant of larger valve area (coefficient 0.467 cm(2), standard error 0.105, P < 0.001). On multivariable analysis, no significant impact on post-repair gradient was observed (–0.370 mmHg, standard error 0.167, P = 0.36). At follow-up, the differences between groups disappeared and multivariable regression analysis failed to demonstrate a significant impact of annuloplasty device type on mitral valve gradient (coefficient –0.095 mmHg, standard error 0.171, P = 1.00) or area (coefficient –0.085 cm(2), standard error 0.120, P = 1.00). These results were confirmed with a linear mixed model analysis. CONCLUSIONS: Partial band annuloplasty was related to an improved haemodynamic profile directly after valve repair for degenerative disease but the effect was short-lived. Our results suggest that the type of annuloplasty device has no durable impact on diastolic valve performance. Oxford University Press 2023-09-09 /pmc/articles/PMC10517645/ /pubmed/37688566 http://dx.doi.org/10.1093/ejcts/ezad307 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (https://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Conventional Valve Operations
Tomšič, Anton
Sandoval, Elena
Meucci, Maria C
Nabeta, Takeru
Castella, Manuel
Muro, Anna
Klautz, Robert J M
Ajmone Marsan, Nina
Pereda, Daniel
Palmen, Meindert
The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
title The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
title_full The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
title_fullStr The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
title_full_unstemmed The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
title_short The impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
title_sort impact of annuloplasty ring or band implantation on post-repair mitral valve haemodynamic performance
topic Conventional Valve Operations
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10517645/
https://www.ncbi.nlm.nih.gov/pubmed/37688566
http://dx.doi.org/10.1093/ejcts/ezad307
work_keys_str_mv AT tomsicanton theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT sandovalelena theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT meuccimariac theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT nabetatakeru theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT castellamanuel theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT muroanna theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT klautzrobertjm theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT ajmonemarsannina theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT peredadaniel theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT palmenmeindert theimpactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT tomsicanton impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT sandovalelena impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT meuccimariac impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT nabetatakeru impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT castellamanuel impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT muroanna impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT klautzrobertjm impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT ajmonemarsannina impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT peredadaniel impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance
AT palmenmeindert impactofannuloplastyringorbandimplantationonpostrepairmitralvalvehaemodynamicperformance