Dispersion-corrected extracorporeal arterial input functions in PET studies of mice: a comparison to intracorporeal microprobe measurements

BACKGROUND: Kinetic modelling of dynamic PET typically requires knowledge of the arterial radiotracer concentration (arterial input function, AIF). Its accurate determination is very difficult in mice. AIF measurements in an extracorporeal shunt can be performed; however, this introduces catheter di...

Descripción completa

Detalles Bibliográficos
Autores principales: Cufe, Juela, Gierse, Florian, Schäfers, Klaus P., Hermann, Sven, Schäfers, Michael A., Backhaus, Philipp, Büther, Florian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10522560/
https://www.ncbi.nlm.nih.gov/pubmed/37752319
http://dx.doi.org/10.1186/s13550-023-01031-z
Descripción
Sumario:BACKGROUND: Kinetic modelling of dynamic PET typically requires knowledge of the arterial radiotracer concentration (arterial input function, AIF). Its accurate determination is very difficult in mice. AIF measurements in an extracorporeal shunt can be performed; however, this introduces catheter dispersion. We propose a framework for extracorporeal dispersion correction and validated it by comparison to invasively determined intracorporeal AIFs using implanted microprobes. RESULTS: The response of an extracorporeal radiation detector to radioactivity boxcar functions, characterised by a convolution-based dispersion model, gave best fits using double-gamma variate and single-gamma variate kernels compared to mono-exponential kernels for the investigated range of flow rates. Parametric deconvolution with the optimal kernels was performed on 9 mice that were injected with a bolus of 39 ± 25 MBq [(18)F]F-PSMA-1007 after application of an extracorporeal circulation for three different flow rates in order to correct for dispersion. Comparison with synchronous implantation of microprobes for invasive aortic AIF recordings showed favourable correspondence, with no significant difference in terms of area-under-curve after 300 s and 5000 s. One-tissue and two-tissue compartment model simulations were performed to investigate differences in kinetic parameters between intra- and extracorporeally measured AIFs. Results of the modelling study revealed kinetic parameters close to the chosen simulated values in all compartment models. CONCLUSION: The high correspondence of simultaneously intra- and extracorporeally determined AIFs and resulting model parameters establishes a feasible framework for extracorporeal dispersion correction. This should allow more precise and accurate kinetic modelling in small animal experiments. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s13550-023-01031-z.