Augmented Reality Surgical Navigation in Minimally Invasive Spine Surgery: A Preclinical Study

Background: In minimally invasive spine surgery (MISS), where the surgeon cannot directly see the patient’s internal anatomical structure, the implementation of augmented reality (AR) technology may solve this problem. Methods: We combined AR, artificial intelligence, and optical tracking to enhance...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Xin, Liu, Xiaoguang, Zhu, Bin, Hou, Xiangyu, Hai, Bao, Yu, Dongfang, Zheng, Wenhao, Li, Ranyang, Pan, Junjun, Yao, Youjie, Dai, Zailin, Zeng, Haijun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10525156/
https://www.ncbi.nlm.nih.gov/pubmed/37760196
http://dx.doi.org/10.3390/bioengineering10091094
Descripción
Sumario:Background: In minimally invasive spine surgery (MISS), where the surgeon cannot directly see the patient’s internal anatomical structure, the implementation of augmented reality (AR) technology may solve this problem. Methods: We combined AR, artificial intelligence, and optical tracking to enhance the augmented reality minimally invasive spine surgery (AR-MISS) system. The system has three functions: AR radiograph superimposition, AR real-time puncture needle tracking, and AR intraoperative navigation. The three functions of the system were evaluated through beagle animal experiments. Results: The AR radiographs were successfully superimposed on the real intraoperative videos. The anteroposterior (AP) and lateral errors of superimposed AR radiographs were 0.74 ± 0.21 mm and 1.13 ± 0.40 mm, respectively. The puncture needles could be tracked by the AR-MISS system in real time. The AP and lateral errors of the real-time AR needle tracking were 1.26 ± 0.20 mm and 1.22 ± 0.25 mm, respectively. With the help of AR radiographs and AR puncture needles, the puncture procedure could be guided visually by the system in real-time. The anteroposterior and lateral errors of AR-guided puncture were 2.47 ± 0.86 mm and 2.85 ± 1.17 mm, respectively. Conclusions: The results indicate that the AR-MISS system is accurate and applicable.