Development and Validation of a Novel In Vitro Joint Testing System for Reproduction of In Vivo Dynamic Muscle Force

In vitro biomechanical experiments utilizing cadaveric specimens are one of the most effective methods for rehearsing surgical procedures, testing implants, and guiding postoperative rehabilitation. Applying dynamic physiological muscle force to the specimens is a challenge to reconstructing the env...

Descripción completa

Detalles Bibliográficos
Autores principales: Yang, Yangyang, Wang, Yufan, Zheng, Nan, Cheng, Rongshan, Zou, Diyang, Zhao, Jie, Tsai, Tsung-Yuan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10525521/
https://www.ncbi.nlm.nih.gov/pubmed/37760108
http://dx.doi.org/10.3390/bioengineering10091006
Descripción
Sumario:In vitro biomechanical experiments utilizing cadaveric specimens are one of the most effective methods for rehearsing surgical procedures, testing implants, and guiding postoperative rehabilitation. Applying dynamic physiological muscle force to the specimens is a challenge to reconstructing the environment of bionic mechanics in vivo, which is often ignored in the in vitro experiment. The current work aims to establish a hardware platform and numerical computation methods to reproduce dynamic muscle forces that can be applied to mechanical testing on in vitro specimens. Dynamic muscle loading is simulated through numerical computation, and the inputs of the platform will be derived. Then, the accuracy and robustness of the platform will be evaluated through actual muscle loading tests in vitro. The tests were run on three muscles (gastrocnemius lateralis, the rectus femoris, and the semitendinosus) around the knee joint and the results showed that the platform can accurately reproduce the magnitude of muscle strength (errors range from −6.2% to 1.81%) and changing pattern (goodness-of-fit range coefficient ranges from 0.00 to 0.06) of target muscle forces. The robustness of the platform is mainly manifested in that the platform can still accurately reproduce muscle force after changing the hardware combination. Additionally, the standard deviation of repeated test results is very small (standard ranges of hardware combination 1: 0.34 N~2.79 N vs. hardware combination 2: 0.68 N~2.93 N). Thus, the platform can stably and accurately reproduce muscle forces in vitro, and it has great potential to be applied in the future musculoskeletal loading system.