Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †

SIMPLE SUMMARY: Hypersomatotropism results from excess growth hormone production by a pituitary tumour and represents an important underlying cause of diabetes mellitus in cats. Diagnosis of hypersomatotropism is mainly based on the measurement of insulin-like growth factor-1 (IGF-1), ideally in com...

Descripción completa

Detalles Bibliográficos
Autores principales: Güssow, Arne, Thalmeier, Sabine, Gostelow, Ruth, Langenstein, Judith, Foerster, Gesine, Bauer, Natali, Hazuchova, Katarina
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10534906/
https://www.ncbi.nlm.nih.gov/pubmed/37756097
http://dx.doi.org/10.3390/vetsci10090575
_version_ 1785112504383307776
author Güssow, Arne
Thalmeier, Sabine
Gostelow, Ruth
Langenstein, Judith
Foerster, Gesine
Bauer, Natali
Hazuchova, Katarina
author_facet Güssow, Arne
Thalmeier, Sabine
Gostelow, Ruth
Langenstein, Judith
Foerster, Gesine
Bauer, Natali
Hazuchova, Katarina
author_sort Güssow, Arne
collection PubMed
description SIMPLE SUMMARY: Hypersomatotropism results from excess growth hormone production by a pituitary tumour and represents an important underlying cause of diabetes mellitus in cats. Diagnosis of hypersomatotropism is mainly based on the measurement of insulin-like growth factor-1 (IGF-1), ideally in combination with imaging of the head. The most well-validated assay (radioimmunoassay), however, requires facilities approved to handle radioactivity and is costly. This study validated an alternative method for IGF-1 measurement in cats, a chemiluminescence assay, and compared it to the radioimmunoassay. It also established a reference interval for IGF-1. This will increase availability of IGF-1 measurement and facilitate diagnosis of hypersomatotropism. ABSTRACT: Previously, radioimmunoassay (RIA) has been the only assay to measure insulin-like growth factor-1 (IGF-1) to diagnose hypersomatotropism (HS). Due to radiation concerns, availability, and the cost of IGF-1 RIA, validation of assays for automated analysers such as a chemiluminescent immunoassay (CLIA) is needed. The aim of this study was to validate a CLIA for measurement of feline IGF-1 (IMMULITE 2000(®) XPi, Siemens Medical Solutions Diagnostics, Malvern, PA, USA) compared to IGF1 RIA, establish reference interval (RI), and determine a cut-off value for diagnosis of HS in diabetic cats. Validation of assay performance included precision, linearity, and recovery studies. Right-sided RI was determined using surplus serum of 50 healthy adult cats. Surplus serum samples of diabetic cats with known IGF-1 concentration with (n = 32/68) and without HS (n = 36/68) were used for method comparison with RIA. The cut-off for diagnosis of HS was established using receiver operating characteristic (ROC) analysis. The intra-assay coefficient of variation (CV) was ≤4.7%, and the inter-assay CV was ≤5.6% for samples with low, medium, and high IGF-1 concentration. Linearity was excellent (R(2) > 0.99). The correlation between CLIA and RIA was very high (r(s) = 0.97), with a mean negative bias for CLIA of 24.5%. The upper limit of RI was 670 ng/mL. ROC analysis showed an area under the curve of 0.94, with best cut-off for diagnosis of HS at 746 ng/mL (sensitivity, 84.4%; specificity, 97.2%). The performance of CLIA was good, and the RI and cut-off for HS diagnosis established in this study allow for CLIA to be used in routine work-up of diabetic cats.
format Online
Article
Text
id pubmed-10534906
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-105349062023-09-29 Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats † Güssow, Arne Thalmeier, Sabine Gostelow, Ruth Langenstein, Judith Foerster, Gesine Bauer, Natali Hazuchova, Katarina Vet Sci Article SIMPLE SUMMARY: Hypersomatotropism results from excess growth hormone production by a pituitary tumour and represents an important underlying cause of diabetes mellitus in cats. Diagnosis of hypersomatotropism is mainly based on the measurement of insulin-like growth factor-1 (IGF-1), ideally in combination with imaging of the head. The most well-validated assay (radioimmunoassay), however, requires facilities approved to handle radioactivity and is costly. This study validated an alternative method for IGF-1 measurement in cats, a chemiluminescence assay, and compared it to the radioimmunoassay. It also established a reference interval for IGF-1. This will increase availability of IGF-1 measurement and facilitate diagnosis of hypersomatotropism. ABSTRACT: Previously, radioimmunoassay (RIA) has been the only assay to measure insulin-like growth factor-1 (IGF-1) to diagnose hypersomatotropism (HS). Due to radiation concerns, availability, and the cost of IGF-1 RIA, validation of assays for automated analysers such as a chemiluminescent immunoassay (CLIA) is needed. The aim of this study was to validate a CLIA for measurement of feline IGF-1 (IMMULITE 2000(®) XPi, Siemens Medical Solutions Diagnostics, Malvern, PA, USA) compared to IGF1 RIA, establish reference interval (RI), and determine a cut-off value for diagnosis of HS in diabetic cats. Validation of assay performance included precision, linearity, and recovery studies. Right-sided RI was determined using surplus serum of 50 healthy adult cats. Surplus serum samples of diabetic cats with known IGF-1 concentration with (n = 32/68) and without HS (n = 36/68) were used for method comparison with RIA. The cut-off for diagnosis of HS was established using receiver operating characteristic (ROC) analysis. The intra-assay coefficient of variation (CV) was ≤4.7%, and the inter-assay CV was ≤5.6% for samples with low, medium, and high IGF-1 concentration. Linearity was excellent (R(2) > 0.99). The correlation between CLIA and RIA was very high (r(s) = 0.97), with a mean negative bias for CLIA of 24.5%. The upper limit of RI was 670 ng/mL. ROC analysis showed an area under the curve of 0.94, with best cut-off for diagnosis of HS at 746 ng/mL (sensitivity, 84.4%; specificity, 97.2%). The performance of CLIA was good, and the RI and cut-off for HS diagnosis established in this study allow for CLIA to be used in routine work-up of diabetic cats. MDPI 2023-09-15 /pmc/articles/PMC10534906/ /pubmed/37756097 http://dx.doi.org/10.3390/vetsci10090575 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Güssow, Arne
Thalmeier, Sabine
Gostelow, Ruth
Langenstein, Judith
Foerster, Gesine
Bauer, Natali
Hazuchova, Katarina
Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †
title Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †
title_full Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †
title_fullStr Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †
title_full_unstemmed Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †
title_short Method Validation and Establishment of Reference Intervals for an Insulin-like Growth Factor-1 Chemiluminescent Immunoassay in Cats †
title_sort method validation and establishment of reference intervals for an insulin-like growth factor-1 chemiluminescent immunoassay in cats †
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10534906/
https://www.ncbi.nlm.nih.gov/pubmed/37756097
http://dx.doi.org/10.3390/vetsci10090575
work_keys_str_mv AT gussowarne methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats
AT thalmeiersabine methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats
AT gostelowruth methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats
AT langensteinjudith methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats
AT foerstergesine methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats
AT bauernatali methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats
AT hazuchovakatarina methodvalidationandestablishmentofreferenceintervalsforaninsulinlikegrowthfactor1chemiluminescentimmunoassayincats