Conversion of the bronchial tree into a conforming electrode to ablate the lung nodule in a porcine model

BACKGROUND: Radiofrequency ablation (RFA) is one of the treatment options for lung nodules. However, the need for exact delivery of the rigid metal electrode into the center of the target mass often leads to complications or suboptimal results. To overcome these limitations, a concept of conforming...

Descripción completa

Detalles Bibliográficos
Autores principales: Shah, Izaz Ali, Seol, Hee Yun, Cho, Youngdae, Ji, Wonjun, Seo, Jaeyoung, Lee, Cheolmin, Chon, Min-Ku, Shin, Donghoon, Kim, Justin H., Choo, Ki-Seok, Park, Junhui, Kim, Juhyung, Yoo, Hyoungsuk, Kim, June-Hong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10541426/
https://www.ncbi.nlm.nih.gov/pubmed/37775526
http://dx.doi.org/10.1038/s43856-023-00362-1
Descripción
Sumario:BACKGROUND: Radiofrequency ablation (RFA) is one of the treatment options for lung nodules. However, the need for exact delivery of the rigid metal electrode into the center of the target mass often leads to complications or suboptimal results. To overcome these limitations, a concept of conforming electrodes using a flexible material has been tested in this study. METHODS: A bronchoscopy-guided RFA (CAROL) under a temperature-controlled mode was tested in in-vivo and ex-vivo porcine lungs. Gallium-based liquid metal was used for turning the bronchial tree into temporary RF electrodes. A customized bronchoscopy-guided balloon-tipped guiding catheter (CAROL catheter) was used to make the procedure feasible under fluoroscopy imaging guidance. The computer simulation was also performed to gain further insight into the ablation results. Safety was also assessed including the liquid metal remaining in the body. RESULTS: The bronchial electrode injected from the CAROL catheter was able to turn the target site bronchial air pipe into a temporally multi-tined RF electrode. The mean volume of Gallium for each effective CAROL was 0.46 ± 0.47 ml. The ablation results showed highly efficacious and consistent results, especially in the peripheral lung. Most bronchial electrodes were also retrieved by either bronchoscopic suction immediately after the procedure or by natural expectoration thereafter. The liquid metal used in these experiments did not have any significant safety issues. Computer simulation also supports these results. CONCLUSION: The CAROL ablation was very effective and safe in porcine lungs showing encouraging potential to overcome the conventional approaches.