Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model
BACKGROUND: Little is known about the stability of adjustable-loop devices (ALDs) for anterior cruciate ligament (ACL) reconstruction (ACLR). PURPOSE: To evaluate the stabilization behavior of 3 different types of ALDs for all-inside ACLR in a full-construct surgical technique-based manner. STUDY DE...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10541758/ https://www.ncbi.nlm.nih.gov/pubmed/37786476 http://dx.doi.org/10.1177/23259671231201461 |
_version_ | 1785113966875246592 |
---|---|
author | Bachmaier, Samuel Monaco, Edoardo Smith, Patrick A. Frank, Rachel M. Matzkin, Elizabeth G. Wijdicks, Coen A. |
author_facet | Bachmaier, Samuel Monaco, Edoardo Smith, Patrick A. Frank, Rachel M. Matzkin, Elizabeth G. Wijdicks, Coen A. |
author_sort | Bachmaier, Samuel |
collection | PubMed |
description | BACKGROUND: Little is known about the stability of adjustable-loop devices (ALDs) for anterior cruciate ligament (ACL) reconstruction (ACLR). PURPOSE: To evaluate the stabilization behavior of 3 different types of ALDs for all-inside ACLR in a full-construct surgical technique-based manner. STUDY DESIGN: Controlled laboratory study. METHODS: The femoral and tibial devices of Ultrabutton (Smith & Nephew), Infinity (Conmed), and TightRope II (Arthrex) were applied to quadrupled bovine tendon grafts (n = 8 each) with tibial-sided traction applied (350 N) for graft tensioning in a simulated fully extended knee. Knotless femoral graft fixation was based on either a suture-locking device (SLD; Ultrabutton), button-locking device (BLD; Infinity), or dual-locking device (DLD; TightRope II). All constructs were progressively loaded (50 N/500 cycles) from 50 to 300 N for 3000 cycles (0.75 Hz), including complete unloading situations and pull to failure (50 mm/min). Construct elongation, stiffness, and ultimate load were analyzed. RESULTS: BLD showed significantly greater initial elongation (-2.69 ± 0.15 mm) than DLD (-3.19 ± 0.21 mm; P < .001) but behaved similarly to SLD (-2.93 ± 0.23 mm). While DLD and SLD had the smallest initial elongation at the same significance level, they behaved opposite to each other with gradually increasing peak loading. At the end of testing, DLD had the lowest (-0.64 ± 0.32 mm) and SLD the highest (3.41 ± 1.01 mm) total elongation (P < .003 for both). SLD displayed significantly higher dynamic elongation (6.34 ± 0.23 mm) than BLD (3.21 ± 0.61 mm) and DLD (2.56 ± 0.31 mm) (P < .001 for both). The failure load of BLD (865.0 ± 183.8 N) was significantly lower (P < .026) compared with SLD and DLD (>1000 N). The predominant failure mode was suture rupture and tibial bone breakage with button subsidence (SLD, n = 4). No significant difference in stiffness between constructs was found. CONCLUSION: While DLD successfully restricted critical construct elongation, BLD partially and SLD completely exceeded the clinical failure threshold (>3 mm) of plastic elongation with loop lengthening during increasing cyclic peak loading with complete unloading. Higher failure loads of SLD and DLD implants (>1000 N) were achieved at similar construct stiffness to BLD. CLINICAL RELEVANCE: A detailed biomechanical understanding of the stabilization potential is pertinent to the continued evolution of ALDs to improve clinical outcomes. |
format | Online Article Text |
id | pubmed-10541758 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-105417582023-10-02 Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model Bachmaier, Samuel Monaco, Edoardo Smith, Patrick A. Frank, Rachel M. Matzkin, Elizabeth G. Wijdicks, Coen A. Orthop J Sports Med Original Research BACKGROUND: Little is known about the stability of adjustable-loop devices (ALDs) for anterior cruciate ligament (ACL) reconstruction (ACLR). PURPOSE: To evaluate the stabilization behavior of 3 different types of ALDs for all-inside ACLR in a full-construct surgical technique-based manner. STUDY DESIGN: Controlled laboratory study. METHODS: The femoral and tibial devices of Ultrabutton (Smith & Nephew), Infinity (Conmed), and TightRope II (Arthrex) were applied to quadrupled bovine tendon grafts (n = 8 each) with tibial-sided traction applied (350 N) for graft tensioning in a simulated fully extended knee. Knotless femoral graft fixation was based on either a suture-locking device (SLD; Ultrabutton), button-locking device (BLD; Infinity), or dual-locking device (DLD; TightRope II). All constructs were progressively loaded (50 N/500 cycles) from 50 to 300 N for 3000 cycles (0.75 Hz), including complete unloading situations and pull to failure (50 mm/min). Construct elongation, stiffness, and ultimate load were analyzed. RESULTS: BLD showed significantly greater initial elongation (-2.69 ± 0.15 mm) than DLD (-3.19 ± 0.21 mm; P < .001) but behaved similarly to SLD (-2.93 ± 0.23 mm). While DLD and SLD had the smallest initial elongation at the same significance level, they behaved opposite to each other with gradually increasing peak loading. At the end of testing, DLD had the lowest (-0.64 ± 0.32 mm) and SLD the highest (3.41 ± 1.01 mm) total elongation (P < .003 for both). SLD displayed significantly higher dynamic elongation (6.34 ± 0.23 mm) than BLD (3.21 ± 0.61 mm) and DLD (2.56 ± 0.31 mm) (P < .001 for both). The failure load of BLD (865.0 ± 183.8 N) was significantly lower (P < .026) compared with SLD and DLD (>1000 N). The predominant failure mode was suture rupture and tibial bone breakage with button subsidence (SLD, n = 4). No significant difference in stiffness between constructs was found. CONCLUSION: While DLD successfully restricted critical construct elongation, BLD partially and SLD completely exceeded the clinical failure threshold (>3 mm) of plastic elongation with loop lengthening during increasing cyclic peak loading with complete unloading. Higher failure loads of SLD and DLD implants (>1000 N) were achieved at similar construct stiffness to BLD. CLINICAL RELEVANCE: A detailed biomechanical understanding of the stabilization potential is pertinent to the continued evolution of ALDs to improve clinical outcomes. SAGE Publications 2023-09-29 /pmc/articles/PMC10541758/ /pubmed/37786476 http://dx.doi.org/10.1177/23259671231201461 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by-nc-nd/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 License (https://creativecommons.org/licenses/by-nc-nd/4.0/) which permits non-commercial use, reproduction and distribution of the work as published without adaptation or alteration, without further permission provided the original work is attributed as specified on the SAGE and Open Access page (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Research Bachmaier, Samuel Monaco, Edoardo Smith, Patrick A. Frank, Rachel M. Matzkin, Elizabeth G. Wijdicks, Coen A. Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model |
title | Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model |
title_full | Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model |
title_fullStr | Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model |
title_full_unstemmed | Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model |
title_short | Biomechanical Comparison of 3 Adjustable-Loop Suspensory Devices for All-Inside ACL Reconstruction: A Time-Zero Full-Construct Model |
title_sort | biomechanical comparison of 3 adjustable-loop suspensory devices for all-inside acl reconstruction: a time-zero full-construct model |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10541758/ https://www.ncbi.nlm.nih.gov/pubmed/37786476 http://dx.doi.org/10.1177/23259671231201461 |
work_keys_str_mv | AT bachmaiersamuel biomechanicalcomparisonof3adjustableloopsuspensorydevicesforallinsideaclreconstructionatimezerofullconstructmodel AT monacoedoardo biomechanicalcomparisonof3adjustableloopsuspensorydevicesforallinsideaclreconstructionatimezerofullconstructmodel AT smithpatricka biomechanicalcomparisonof3adjustableloopsuspensorydevicesforallinsideaclreconstructionatimezerofullconstructmodel AT frankrachelm biomechanicalcomparisonof3adjustableloopsuspensorydevicesforallinsideaclreconstructionatimezerofullconstructmodel AT matzkinelizabethg biomechanicalcomparisonof3adjustableloopsuspensorydevicesforallinsideaclreconstructionatimezerofullconstructmodel AT wijdickscoena biomechanicalcomparisonof3adjustableloopsuspensorydevicesforallinsideaclreconstructionatimezerofullconstructmodel |