Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out

OBJECTIVE: This study aimed to introduce a method to extract the 3-dimensional spatial position of the femoral head implant from 2-dimensional fluoroscopic projections, allowing surgeons to assess fixation much more accurately and prevent cut-out complications in proximal femoral nailing. METHODS: T...

Descripción completa

Detalles Bibliográficos
Autores principales: Subaşı, Ömer, Aslan, Lercan, Oral, Atacan, Demirhan, Mehmet, Seyahi, Aksel, Lazoğlu, İsmail
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Turkish Association of Orthopaedics and Traumatology, and Turkish Society of Orthopaedics and Traumatology 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10542487/
https://www.ncbi.nlm.nih.gov/pubmed/37670446
http://dx.doi.org/10.5152/j.aott.2023.22175
_version_ 1785114104940199936
author Subaşı, Ömer
Aslan, Lercan
Oral, Atacan
Demirhan, Mehmet
Seyahi, Aksel
Lazoğlu, İsmail
author_facet Subaşı, Ömer
Aslan, Lercan
Oral, Atacan
Demirhan, Mehmet
Seyahi, Aksel
Lazoğlu, İsmail
author_sort Subaşı, Ömer
collection PubMed
description OBJECTIVE: This study aimed to introduce a method to extract the 3-dimensional spatial position of the femoral head implant from 2-dimensional fluoroscopic projections, allowing surgeons to assess fixation much more accurately and prevent cut-out complications in proximal femoral nailing. METHODS: To define a safety region for the tip in the femoral head, a novel 3-dimensional distance-based risk parameter called TSD3D was introduced. An intersection algorithm was developed that solely takes the fluoroscopic anteroposterior and lateral distances to reveal the 3-dimensional location of the screw or Kirschner wire tip, enabling the utilization of the 3-dimensional parameter. Orthogonal perspectives of 6 femur proximal bone substitutes with randomly inserted Kirschner wires were imaged under fluoroscopy. The developed algorithm was used to calculate the implant tip location in 3-dimensional from 2-dimensional images for each case. Algorithm accuracy was validated with the computed tomography-obtained 3-dimensional models of the same femur substitutes. RESULTS: The newly introduced risk parameter successfully visualizes 3-dimensional safety regions. Utilizing the 2-dimensional fluoroscopic distances as inputs to the algorithm, the 3-dimensional position of the implanted Kirschner wire tip is calculated with a maximum of 9.8% error for a single Cartesian-coordinate measurement comparison. CONCLUSION: By incorporating the newly introduced 3-dimensional risk parameter, surgeons can more precisely evaluate the position of the implant and avoid cut-out complications, instead of relying solely on misleading 2-dimensional fluoroscopic projections of the femoral head.
format Online
Article
Text
id pubmed-10542487
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Turkish Association of Orthopaedics and Traumatology, and Turkish Society of Orthopaedics and Traumatology
record_format MEDLINE/PubMed
spelling pubmed-105424872023-10-03 Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out Subaşı, Ömer Aslan, Lercan Oral, Atacan Demirhan, Mehmet Seyahi, Aksel Lazoğlu, İsmail Acta Orthop Traumatol Turc Research Article OBJECTIVE: This study aimed to introduce a method to extract the 3-dimensional spatial position of the femoral head implant from 2-dimensional fluoroscopic projections, allowing surgeons to assess fixation much more accurately and prevent cut-out complications in proximal femoral nailing. METHODS: To define a safety region for the tip in the femoral head, a novel 3-dimensional distance-based risk parameter called TSD3D was introduced. An intersection algorithm was developed that solely takes the fluoroscopic anteroposterior and lateral distances to reveal the 3-dimensional location of the screw or Kirschner wire tip, enabling the utilization of the 3-dimensional parameter. Orthogonal perspectives of 6 femur proximal bone substitutes with randomly inserted Kirschner wires were imaged under fluoroscopy. The developed algorithm was used to calculate the implant tip location in 3-dimensional from 2-dimensional images for each case. Algorithm accuracy was validated with the computed tomography-obtained 3-dimensional models of the same femur substitutes. RESULTS: The newly introduced risk parameter successfully visualizes 3-dimensional safety regions. Utilizing the 2-dimensional fluoroscopic distances as inputs to the algorithm, the 3-dimensional position of the implanted Kirschner wire tip is calculated with a maximum of 9.8% error for a single Cartesian-coordinate measurement comparison. CONCLUSION: By incorporating the newly introduced 3-dimensional risk parameter, surgeons can more precisely evaluate the position of the implant and avoid cut-out complications, instead of relying solely on misleading 2-dimensional fluoroscopic projections of the femoral head. Turkish Association of Orthopaedics and Traumatology, and Turkish Society of Orthopaedics and Traumatology 2023-07-01 /pmc/articles/PMC10542487/ /pubmed/37670446 http://dx.doi.org/10.5152/j.aott.2023.22175 Text en 2023 authors https://creativecommons.org/licenses/by-nc/4.0/ Content of this journal is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License. (https://creativecommons.org/licenses/by-nc/4.0/)
spellingShingle Research Article
Subaşı, Ömer
Aslan, Lercan
Oral, Atacan
Demirhan, Mehmet
Seyahi, Aksel
Lazoğlu, İsmail
Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
title Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
title_full Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
title_fullStr Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
title_full_unstemmed Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
title_short Intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
title_sort intraoperative fluoroscopic safety assessment of femoral head implants with 3-dimensional risk parameters to minimize cut-out
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10542487/
https://www.ncbi.nlm.nih.gov/pubmed/37670446
http://dx.doi.org/10.5152/j.aott.2023.22175
work_keys_str_mv AT subasıomer intraoperativefluoroscopicsafetyassessmentoffemoralheadimplantswith3dimensionalriskparameterstominimizecutout
AT aslanlercan intraoperativefluoroscopicsafetyassessmentoffemoralheadimplantswith3dimensionalriskparameterstominimizecutout
AT oralatacan intraoperativefluoroscopicsafetyassessmentoffemoralheadimplantswith3dimensionalriskparameterstominimizecutout
AT demirhanmehmet intraoperativefluoroscopicsafetyassessmentoffemoralheadimplantswith3dimensionalriskparameterstominimizecutout
AT seyahiaksel intraoperativefluoroscopicsafetyassessmentoffemoralheadimplantswith3dimensionalriskparameterstominimizecutout
AT lazogluismail intraoperativefluoroscopicsafetyassessmentoffemoralheadimplantswith3dimensionalriskparameterstominimizecutout