Cargando…

Propagation-based phase-contrast imaging of the breast: image quality and the effect of X-ray energy and radiation dose

OBJECTIVES: Propagation-based phase-contrast computed tomography (PB-CT) is a new imaging technique that exploits refractive and absorption properties of X-rays to enhance soft tissue contrast and improve image quality. This study compares image quality of PB-CT and absorption-based CT (AB-CT) for b...

Descripción completa

Detalles Bibliográficos
Autores principales: Gunaseelan, Indusaa, Amin Zadeh, Alaleh, Arhatari, Benedicta, Maksimenko, Anton, Hall, Christopher, Hausermann, Daniel, Kumar, Beena, Fox, Jane, Quiney, Harry, Lockie, Darren, Lewis, Sarah, Brennan, Patrick, Gureyev, Timur, Tavakoli Taba, Seyedamir
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10546460/
https://www.ncbi.nlm.nih.gov/pubmed/37665247
http://dx.doi.org/10.1259/bjr.20221189
Descripción
Sumario:OBJECTIVES: Propagation-based phase-contrast computed tomography (PB-CT) is a new imaging technique that exploits refractive and absorption properties of X-rays to enhance soft tissue contrast and improve image quality. This study compares image quality of PB-CT and absorption-based CT (AB-CT) for breast imaging while exploring X-ray energy and radiation dose. METHODS: Thirty-nine mastectomy samples were scanned at energy levels of 28-34keV using a flat panel detector at radiation dose levels of 4mGy and 2mGy. Image quality was assessed using signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), spatial resolution (res) and visibility (vis). Statistical analysis was performed to compare PB-CT images against their corresponding AB-CT images scanned at 32keV and 4mGy. RESULTS: The PB-CT images at 4mGy, across nearly all energy levels, demonstrated superior image quality than AB-CT images at the same dose. At some energy levels, the 2mGy PB-CT images also showed better image quality in terms of CNR/Res and vis compared to the 4mGy AB-CT images. At both investigated doses, SNR and SNR/res were found to have a statistically significant difference across all energy levels. The difference in vis was statistically significant at some energy levels. CONCLUSION: This study demonstrates superior image quality of PB-CT over AB-CT, with X-ray energy playing a crucial role in determining image quality parameters. ADVANCES IN KNOWLEDGE: Our findings reveal that standard dose PB-CT outperforms standard dose AB-CT across all image quality metrics. Additionally, we demonstrate that low dose PB-CT can produce superior images compared to standard dose AB-CT in terms of CNR/Res and vis.