Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination
The establishment and validation of a straightforward, accurate, and eco-friendly capillary zone electrophoretic-diode array detection (CZE-DAD) procedure has been presented for concurrent measurement of two common antibiotics, namely, linezolid (LIN) and cefixime trihydrate (CEF), in their binary m...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10548557/ https://www.ncbi.nlm.nih.gov/pubmed/37794494 http://dx.doi.org/10.1186/s13065-023-01049-3 |
_version_ | 1785115292791209984 |
---|---|
author | Habeeb, Maha R. Morshedy, Samir M. Daabees, Hoda G. Elonsy, Sohila M. |
author_facet | Habeeb, Maha R. Morshedy, Samir M. Daabees, Hoda G. Elonsy, Sohila M. |
author_sort | Habeeb, Maha R. |
collection | PubMed |
description | The establishment and validation of a straightforward, accurate, and eco-friendly capillary zone electrophoretic-diode array detection (CZE-DAD) procedure has been presented for concurrent measurement of two common antibiotics, namely, linezolid (LIN) and cefixime trihydrate (CEF), in their binary mixture or combined dosage form. The selected fused silica capillary has total and effective lengths equal to 58.5 cm and 50 cm, respectively, with a 50 µm internal diameter. Injections were performed utilizing 100 mM borate buffer at pH 10.2 as the background electrolyte (BGE) with a 15.0 s injection time. The finally utilized voltage was 30 kV. DAD was programmed to measure LIN at 250 nm and CEF at 285 nm. In less than 6 min, the two cited drugs were resolved at 2.51 and 5.47 min for LIN and CEF respectively. The introduced procedure had a linear response in the concentration range of 5–50 μg/mL for both analytes with correlation coefficients > 0.9999. Detection and quantification limits were 1.213 and 4.042 μg/mL, respectively, for LIN and 0.301 and 1.004 μg/mL, respectively, for CEF. Validation was conducted according to the International Council for Harmonization (ICH), concerning linearity, detection and quantitation limits, range, accuracy, precision, selectivity, and robustness. Precision was found acceptable due to the low relative standard deviation (RSD%) values that did not exceed 1.86% either for repeatability or for intermediate precision. Additionally, the adequately recovered concentrations and the low values of percentage relative error (Er%) provide evidence of the accuracy of the proposed method. On the other hand, the robustness of the introduced method was affirmed by the acceptable RSD% values that did not exceed 0.6% after deliberate changes in the following procedure parameters: buffer concentration, buffer pH, and wavelength. Finally, the ability of the presented method to quantify the two tested drugs in laboratory-prepared tablets was confirmed by the adequate recoveries (≥ 99%) utilizing the standard-addition procedure, along with the absence of any significant difference between the proposed method and the reference method as proven by the student’s t-test and the variance-ratio F-test values that did not exceed the theoretical ones. The analytical Eco-Scale and the analytical GREEness metric (AGREE) were the tools utilized for greenness assessment. This CZE procedure is the first electro-driven separation method that was utilized for the analysis of both antibiotics in their combined laboratory-prepared tablets with no interference from the co-formulated adjuvants. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s13065-023-01049-3. |
format | Online Article Text |
id | pubmed-10548557 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-105485572023-10-05 Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination Habeeb, Maha R. Morshedy, Samir M. Daabees, Hoda G. Elonsy, Sohila M. BMC Chem Research The establishment and validation of a straightforward, accurate, and eco-friendly capillary zone electrophoretic-diode array detection (CZE-DAD) procedure has been presented for concurrent measurement of two common antibiotics, namely, linezolid (LIN) and cefixime trihydrate (CEF), in their binary mixture or combined dosage form. The selected fused silica capillary has total and effective lengths equal to 58.5 cm and 50 cm, respectively, with a 50 µm internal diameter. Injections were performed utilizing 100 mM borate buffer at pH 10.2 as the background electrolyte (BGE) with a 15.0 s injection time. The finally utilized voltage was 30 kV. DAD was programmed to measure LIN at 250 nm and CEF at 285 nm. In less than 6 min, the two cited drugs were resolved at 2.51 and 5.47 min for LIN and CEF respectively. The introduced procedure had a linear response in the concentration range of 5–50 μg/mL for both analytes with correlation coefficients > 0.9999. Detection and quantification limits were 1.213 and 4.042 μg/mL, respectively, for LIN and 0.301 and 1.004 μg/mL, respectively, for CEF. Validation was conducted according to the International Council for Harmonization (ICH), concerning linearity, detection and quantitation limits, range, accuracy, precision, selectivity, and robustness. Precision was found acceptable due to the low relative standard deviation (RSD%) values that did not exceed 1.86% either for repeatability or for intermediate precision. Additionally, the adequately recovered concentrations and the low values of percentage relative error (Er%) provide evidence of the accuracy of the proposed method. On the other hand, the robustness of the introduced method was affirmed by the acceptable RSD% values that did not exceed 0.6% after deliberate changes in the following procedure parameters: buffer concentration, buffer pH, and wavelength. Finally, the ability of the presented method to quantify the two tested drugs in laboratory-prepared tablets was confirmed by the adequate recoveries (≥ 99%) utilizing the standard-addition procedure, along with the absence of any significant difference between the proposed method and the reference method as proven by the student’s t-test and the variance-ratio F-test values that did not exceed the theoretical ones. The analytical Eco-Scale and the analytical GREEness metric (AGREE) were the tools utilized for greenness assessment. This CZE procedure is the first electro-driven separation method that was utilized for the analysis of both antibiotics in their combined laboratory-prepared tablets with no interference from the co-formulated adjuvants. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s13065-023-01049-3. Springer International Publishing 2023-10-04 /pmc/articles/PMC10548557/ /pubmed/37794494 http://dx.doi.org/10.1186/s13065-023-01049-3 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Habeeb, Maha R. Morshedy, Samir M. Daabees, Hoda G. Elonsy, Sohila M. Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
title | Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
title_full | Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
title_fullStr | Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
title_full_unstemmed | Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
title_short | Development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
title_sort | development, validation and greenness assessment of a new electro-driven separation method for simultaneous analysis of cefixime trihydrate and linezolid in their fixed dose combination |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10548557/ https://www.ncbi.nlm.nih.gov/pubmed/37794494 http://dx.doi.org/10.1186/s13065-023-01049-3 |
work_keys_str_mv | AT habeebmahar developmentvalidationandgreennessassessmentofanewelectrodrivenseparationmethodforsimultaneousanalysisofcefiximetrihydrateandlinezolidintheirfixeddosecombination AT morshedysamirm developmentvalidationandgreennessassessmentofanewelectrodrivenseparationmethodforsimultaneousanalysisofcefiximetrihydrateandlinezolidintheirfixeddosecombination AT daabeeshodag developmentvalidationandgreennessassessmentofanewelectrodrivenseparationmethodforsimultaneousanalysisofcefiximetrihydrateandlinezolidintheirfixeddosecombination AT elonsysohilam developmentvalidationandgreennessassessmentofanewelectrodrivenseparationmethodforsimultaneousanalysisofcefiximetrihydrateandlinezolidintheirfixeddosecombination |