Visible-light optical coherence tomography platform for the characterization of the skin barrier
We demonstrate a free-space, trolley-mountable Fourier domain visible-light optical coherence tomography (OCT) system for studying the stratum corneum in non-palmar human skin. An axial resolution of 1 µm in tissue and at least −75 dB sensitivity have been achieved. High-quality B-scans, containing...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Optica Publishing Group
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10549730/ https://www.ncbi.nlm.nih.gov/pubmed/37799680 http://dx.doi.org/10.1364/BOE.494356 |
Sumario: | We demonstrate a free-space, trolley-mountable Fourier domain visible-light optical coherence tomography (OCT) system for studying the stratum corneum in non-palmar human skin. An axial resolution of 1 µm in tissue and at least −75 dB sensitivity have been achieved. High-quality B-scans, containing 1600 A-scans, are acquired at a rate of 39 Hz. Images from the dorsal hand, ventral wrist and ventral forearm areas are obtained, with a clearly resolved stratum corneum layer (typically 5–15 µm thick) presenting as a hypoechogenic dark layer below the bright entrance signal, similar to that found in palmar skin with traditional OCT systems. We find that the appearance of the stratum corneum layer strongly depends on its water content, becoming brighter after occlusive hydration. |
---|