FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk

Disclosure: J. Li: None. L. Lee: None. F. Wang: None. C. Yang: None. C. Huang: None. Background: Adequate iodine status in lactating women is defined by a maternal median urinary iodine concentration above 100 μg/L. However, the current standard does not account for the secretion of iodine into brea...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Jia-Zhen, Lee, Lin-Hsuan, Wang, Fan-Fen, Yang, Chen-Chang, Jui Huang, Chun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10555403/
http://dx.doi.org/10.1210/jendso/bvad114.1841
_version_ 1785116649198714880
author Li, Jia-Zhen
Lee, Lin-Hsuan
Wang, Fan-Fen
Yang, Chen-Chang
Jui Huang, Chun
author_facet Li, Jia-Zhen
Lee, Lin-Hsuan
Wang, Fan-Fen
Yang, Chen-Chang
Jui Huang, Chun
author_sort Li, Jia-Zhen
collection PubMed
description Disclosure: J. Li: None. L. Lee: None. F. Wang: None. C. Yang: None. C. Huang: None. Background: Adequate iodine status in lactating women is defined by a maternal median urinary iodine concentration above 100 μg/L. However, the current standard does not account for the secretion of iodine into breastmilk and could not truly reflect the amount of iodine delivered to the infants. Measuring breastmilk median iodine concentration (BMIC) is of crucial importance but the method to measure BMIC has not been developed and validated in Taiwan. Method: We adopted the ammonia dilution method without prior sample digestion to measure BMIC by inductively coupled plasma mass spectrometry (ICP-MS). Samples and iodine calibrators were prepared into an aqueous solution containing Triton X-100, 0.5% ammonia solution and tellurium ((128)Te) as the internal standard. Precision, accuracy, serial dilution, and recovery tests were performed for method validation. Results: The intra-assay and inter-assay coefficient of variation for the four human breast milk samples with different iodine concentrations ranged from 3.2% to 4.7% and 2.3% to 5.5%, respectively. The standard NIST 1549 milk powder was prepared into three different concentrations of 50 μg/L, 100 μg/L and 200 μg/L to assess accuracy and the biases were <5%. A recovery of 95-105% was achieved for four human breast milk samples spiked with sodium iodide solution. The serial dilution test confirmed linearity up to 500 μg/L. The limit for detection and quantification was 0.73 μg/L and 2.19 μg/L, respectively. Conclusion: The results of the current study confirmed that this ICP-MS method is accurate and reliable in measuring BMIC. Presentation: Friday, June 16, 2023
format Online
Article
Text
id pubmed-10555403
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-105554032023-10-06 FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk Li, Jia-Zhen Lee, Lin-Hsuan Wang, Fan-Fen Yang, Chen-Chang Jui Huang, Chun J Endocr Soc Thyroid Disclosure: J. Li: None. L. Lee: None. F. Wang: None. C. Yang: None. C. Huang: None. Background: Adequate iodine status in lactating women is defined by a maternal median urinary iodine concentration above 100 μg/L. However, the current standard does not account for the secretion of iodine into breastmilk and could not truly reflect the amount of iodine delivered to the infants. Measuring breastmilk median iodine concentration (BMIC) is of crucial importance but the method to measure BMIC has not been developed and validated in Taiwan. Method: We adopted the ammonia dilution method without prior sample digestion to measure BMIC by inductively coupled plasma mass spectrometry (ICP-MS). Samples and iodine calibrators were prepared into an aqueous solution containing Triton X-100, 0.5% ammonia solution and tellurium ((128)Te) as the internal standard. Precision, accuracy, serial dilution, and recovery tests were performed for method validation. Results: The intra-assay and inter-assay coefficient of variation for the four human breast milk samples with different iodine concentrations ranged from 3.2% to 4.7% and 2.3% to 5.5%, respectively. The standard NIST 1549 milk powder was prepared into three different concentrations of 50 μg/L, 100 μg/L and 200 μg/L to assess accuracy and the biases were <5%. A recovery of 95-105% was achieved for four human breast milk samples spiked with sodium iodide solution. The serial dilution test confirmed linearity up to 500 μg/L. The limit for detection and quantification was 0.73 μg/L and 2.19 μg/L, respectively. Conclusion: The results of the current study confirmed that this ICP-MS method is accurate and reliable in measuring BMIC. Presentation: Friday, June 16, 2023 Oxford University Press 2023-10-05 /pmc/articles/PMC10555403/ http://dx.doi.org/10.1210/jendso/bvad114.1841 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the Endocrine Society. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Thyroid
Li, Jia-Zhen
Lee, Lin-Hsuan
Wang, Fan-Fen
Yang, Chen-Chang
Jui Huang, Chun
FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk
title FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk
title_full FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk
title_fullStr FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk
title_full_unstemmed FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk
title_short FRI495 Analytical Validation Of An Inductively Coupled Plasma Mass Spectrometry Method For Measuring Iodine Content In Breast Milk
title_sort fri495 analytical validation of an inductively coupled plasma mass spectrometry method for measuring iodine content in breast milk
topic Thyroid
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10555403/
http://dx.doi.org/10.1210/jendso/bvad114.1841
work_keys_str_mv AT lijiazhen fri495analyticalvalidationofaninductivelycoupledplasmamassspectrometrymethodformeasuringiodinecontentinbreastmilk
AT leelinhsuan fri495analyticalvalidationofaninductivelycoupledplasmamassspectrometrymethodformeasuringiodinecontentinbreastmilk
AT wangfanfen fri495analyticalvalidationofaninductivelycoupledplasmamassspectrometrymethodformeasuringiodinecontentinbreastmilk
AT yangchenchang fri495analyticalvalidationofaninductivelycoupledplasmamassspectrometrymethodformeasuringiodinecontentinbreastmilk
AT juihuangchun fri495analyticalvalidationofaninductivelycoupledplasmamassspectrometrymethodformeasuringiodinecontentinbreastmilk