Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study

Objective  In this study, we compare the diagnostic accuracy of dual-energy (DE) computed tomography pulmonary angiography (CTPA) derived contrast enhancement (DECTPA, CTPA images with iodine maps) with standard dual-energy pulmonary angiography (SCTPA) for diagnosis of subsegmental pulmonary emboli...

Descripción completa

Detalles Bibliográficos
Autores principales: Yadav, Vivek, Singhal, Manphool, Maralakunte, Muniraju, Sharma, Navneet, Sharma, Arun, Lal, Anupam
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Thieme Medical and Scientific Publishers Pvt. Ltd. 2023
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10556308/
https://www.ncbi.nlm.nih.gov/pubmed/37811170
http://dx.doi.org/10.1055/s-0043-1764489
_version_ 1785116847269478400
author Yadav, Vivek
Singhal, Manphool
Maralakunte, Muniraju
Sharma, Navneet
Sharma, Arun
Lal, Anupam
author_facet Yadav, Vivek
Singhal, Manphool
Maralakunte, Muniraju
Sharma, Navneet
Sharma, Arun
Lal, Anupam
author_sort Yadav, Vivek
collection PubMed
description Objective  In this study, we compare the diagnostic accuracy of dual-energy (DE) computed tomography pulmonary angiography (CTPA) derived contrast enhancement (DECTPA, CTPA images with iodine maps) with standard dual-energy pulmonary angiography (SCTPA) for diagnosis of subsegmental pulmonary embolism in the cases with clinical suspicion of acute pulmonary embolism (APE). Materials and Methods  We included 50 cases with clinical suspicion of APE that were referred for CTPA. All the patients underwent CTPA in the dual-energy protocol. Two radiologists evaluated the images. The first radiologist interpreted the SCTPA images (vascular images) and the second radiologist interpreted the DECTPA (CTPA images with iodine maps) for findings of APE. We calculated the sensitivity, specificity, and negative predictive value of DECTPA vis-à-vis SCTPA images. Results  The DECTPA with the advantage of iodine map utilization yielded higher detection of thrombi in peripheral subsegmental arteries (72 vs. 99; p  = – 0.001) as compared to the SCTPA images by identification of 18 new perfusion defects (interquartile range [IQR]: 0–1) that were consistent with APE. Filling defects were identified in 27 (IQR: 0–4) more subsegmental arteries supplying these 18 areas, which were not detected on SCTPA alone. These 18 perfusion defects were identified in 13 cases. In these 13 cases, 4 new cases were diagnosed that were negative on CTPA ( p  = –0.125). In the evaluation of the APE, sensitivity and specificity were calculated and it was found that DECTPA showed 100% sensitivity and 86% specificity with 100% negative predictive value in the detection of thrombi as compared to the routine CTPA. Conclusion  DECTPA has higher sensitivity and negative predictive value in the detection of the subsegmental perfusion defect identification as compared to SCTPA.
format Online
Article
Text
id pubmed-10556308
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Thieme Medical and Scientific Publishers Pvt. Ltd.
record_format MEDLINE/PubMed
spelling pubmed-105563082023-10-07 Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study Yadav, Vivek Singhal, Manphool Maralakunte, Muniraju Sharma, Navneet Sharma, Arun Lal, Anupam Indian J Radiol Imaging Objective  In this study, we compare the diagnostic accuracy of dual-energy (DE) computed tomography pulmonary angiography (CTPA) derived contrast enhancement (DECTPA, CTPA images with iodine maps) with standard dual-energy pulmonary angiography (SCTPA) for diagnosis of subsegmental pulmonary embolism in the cases with clinical suspicion of acute pulmonary embolism (APE). Materials and Methods  We included 50 cases with clinical suspicion of APE that were referred for CTPA. All the patients underwent CTPA in the dual-energy protocol. Two radiologists evaluated the images. The first radiologist interpreted the SCTPA images (vascular images) and the second radiologist interpreted the DECTPA (CTPA images with iodine maps) for findings of APE. We calculated the sensitivity, specificity, and negative predictive value of DECTPA vis-à-vis SCTPA images. Results  The DECTPA with the advantage of iodine map utilization yielded higher detection of thrombi in peripheral subsegmental arteries (72 vs. 99; p  = – 0.001) as compared to the SCTPA images by identification of 18 new perfusion defects (interquartile range [IQR]: 0–1) that were consistent with APE. Filling defects were identified in 27 (IQR: 0–4) more subsegmental arteries supplying these 18 areas, which were not detected on SCTPA alone. These 18 perfusion defects were identified in 13 cases. In these 13 cases, 4 new cases were diagnosed that were negative on CTPA ( p  = –0.125). In the evaluation of the APE, sensitivity and specificity were calculated and it was found that DECTPA showed 100% sensitivity and 86% specificity with 100% negative predictive value in the detection of thrombi as compared to the routine CTPA. Conclusion  DECTPA has higher sensitivity and negative predictive value in the detection of the subsegmental perfusion defect identification as compared to SCTPA. Thieme Medical and Scientific Publishers Pvt. Ltd. 2023-05-18 /pmc/articles/PMC10556308/ /pubmed/37811170 http://dx.doi.org/10.1055/s-0043-1764489 Text en Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. ( https://creativecommons.org/licenses/by-nc-nd/4.0/ ) https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License, which permits unrestricted reproduction and distribution, for non-commercial purposes only; and use and reproduction, but not distribution, of adapted material for non-commercial purposes only, provided the original work is properly cited.
spellingShingle Yadav, Vivek
Singhal, Manphool
Maralakunte, Muniraju
Sharma, Navneet
Sharma, Arun
Lal, Anupam
Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study
title Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study
title_full Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study
title_fullStr Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study
title_full_unstemmed Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study
title_short Comparison of Dual-Energy Computed Tomography Pulmonary Angiography-Derived Contrast Enhancement with Standard Dual-Energy Pulmonary Angiography in Diagnosing Subsegmental Pulmonary Embolism: A Prospective Study
title_sort comparison of dual-energy computed tomography pulmonary angiography-derived contrast enhancement with standard dual-energy pulmonary angiography in diagnosing subsegmental pulmonary embolism: a prospective study
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10556308/
https://www.ncbi.nlm.nih.gov/pubmed/37811170
http://dx.doi.org/10.1055/s-0043-1764489
work_keys_str_mv AT yadavvivek comparisonofdualenergycomputedtomographypulmonaryangiographyderivedcontrastenhancementwithstandarddualenergypulmonaryangiographyindiagnosingsubsegmentalpulmonaryembolismaprospectivestudy
AT singhalmanphool comparisonofdualenergycomputedtomographypulmonaryangiographyderivedcontrastenhancementwithstandarddualenergypulmonaryangiographyindiagnosingsubsegmentalpulmonaryembolismaprospectivestudy
AT maralakuntemuniraju comparisonofdualenergycomputedtomographypulmonaryangiographyderivedcontrastenhancementwithstandarddualenergypulmonaryangiographyindiagnosingsubsegmentalpulmonaryembolismaprospectivestudy
AT sharmanavneet comparisonofdualenergycomputedtomographypulmonaryangiographyderivedcontrastenhancementwithstandarddualenergypulmonaryangiographyindiagnosingsubsegmentalpulmonaryembolismaprospectivestudy
AT sharmaarun comparisonofdualenergycomputedtomographypulmonaryangiographyderivedcontrastenhancementwithstandarddualenergypulmonaryangiographyindiagnosingsubsegmentalpulmonaryembolismaprospectivestudy
AT lalanupam comparisonofdualenergycomputedtomographypulmonaryangiographyderivedcontrastenhancementwithstandarddualenergypulmonaryangiographyindiagnosingsubsegmentalpulmonaryembolismaprospectivestudy