Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion

BACKGROUND AND OBJECTIVES: A totally implantable venous access port (TIVAP) is a commonly used intravenous infusion device for patients receiving chemotherapy or long-term infusion therapy. To improve the success rate of one-time insertion of the Huber needle, we developed a novel Y-Z magnetic TIVAP...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhang, Miaomiao, Lyu, Xin, Wang, Juanjuan, Shi, Aihua, Zheng, Yunyun, Lyu, Yi, Yan, Xiaopeng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10560742/
https://www.ncbi.nlm.nih.gov/pubmed/37818210
http://dx.doi.org/10.3389/fsurg.2023.1241780
_version_ 1785117787888287744
author Zhang, Miaomiao
Lyu, Xin
Wang, Juanjuan
Shi, Aihua
Zheng, Yunyun
Lyu, Yi
Yan, Xiaopeng
author_facet Zhang, Miaomiao
Lyu, Xin
Wang, Juanjuan
Shi, Aihua
Zheng, Yunyun
Lyu, Yi
Yan, Xiaopeng
author_sort Zhang, Miaomiao
collection PubMed
description BACKGROUND AND OBJECTIVES: A totally implantable venous access port (TIVAP) is a commonly used intravenous infusion device for patients receiving chemotherapy or long-term infusion therapy. To improve the success rate of one-time insertion of the Huber needle, we developed a novel Y-Z magnetic TIVAP (Y-Z MTIVAP), which we produced using three-dimensional printing technology. MATERIALS AND METHODS: The Y-Z MTIVAP includes a magnetic port body and a magnetic positioning device. For testing, we established four venous port implantation models using the two types of TIVAPs and two implantation depth ranges (≤5 mm and >5 mm). Twenty nurses performed Huber needle puncture with the four models, and we recorded the number of attempts required for successful needle insertion, the operation time, and the operator's satisfaction. RESULTS: The success rate for one-time needle insertion with the Y-Z MTIVAP was significantly higher than that with the traditional TIVAP at either depth range (100% vs. 75% at ≤5 mm, p = 0.047; 95% vs. 35% at >5 mm, p < 0.001). With increasing implantation depth, the success rate for one-time insertion was significantly reduced with the traditional TIVAP (75% at ≤5 mm vs. 35% vs. >5 mm, p = 0.025), but the success rate with the Y-Z MTIVAP was not significantly affected (100% vs. 95%, p = 1.000). The operation time with the Y-Z MTIVAP was significantly shorter than that with the traditional TIVAP at either depth range (both p < 0.001), and 90% of operators reported that the Y-Z MTIVAP was superior to the traditional TIVAP. CONCLUSIONS: The theoretical design of Y-Z MTIVAP is feasible, and the preliminary in vitro simulation experiment shows that it can significantly improve puncture success rate and shortened operation time.
format Online
Article
Text
id pubmed-10560742
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-105607422023-10-10 Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion Zhang, Miaomiao Lyu, Xin Wang, Juanjuan Shi, Aihua Zheng, Yunyun Lyu, Yi Yan, Xiaopeng Front Surg Surgery BACKGROUND AND OBJECTIVES: A totally implantable venous access port (TIVAP) is a commonly used intravenous infusion device for patients receiving chemotherapy or long-term infusion therapy. To improve the success rate of one-time insertion of the Huber needle, we developed a novel Y-Z magnetic TIVAP (Y-Z MTIVAP), which we produced using three-dimensional printing technology. MATERIALS AND METHODS: The Y-Z MTIVAP includes a magnetic port body and a magnetic positioning device. For testing, we established four venous port implantation models using the two types of TIVAPs and two implantation depth ranges (≤5 mm and >5 mm). Twenty nurses performed Huber needle puncture with the four models, and we recorded the number of attempts required for successful needle insertion, the operation time, and the operator's satisfaction. RESULTS: The success rate for one-time needle insertion with the Y-Z MTIVAP was significantly higher than that with the traditional TIVAP at either depth range (100% vs. 75% at ≤5 mm, p = 0.047; 95% vs. 35% at >5 mm, p < 0.001). With increasing implantation depth, the success rate for one-time insertion was significantly reduced with the traditional TIVAP (75% at ≤5 mm vs. 35% vs. >5 mm, p = 0.025), but the success rate with the Y-Z MTIVAP was not significantly affected (100% vs. 95%, p = 1.000). The operation time with the Y-Z MTIVAP was significantly shorter than that with the traditional TIVAP at either depth range (both p < 0.001), and 90% of operators reported that the Y-Z MTIVAP was superior to the traditional TIVAP. CONCLUSIONS: The theoretical design of Y-Z MTIVAP is feasible, and the preliminary in vitro simulation experiment shows that it can significantly improve puncture success rate and shortened operation time. Frontiers Media S.A. 2023-09-25 /pmc/articles/PMC10560742/ /pubmed/37818210 http://dx.doi.org/10.3389/fsurg.2023.1241780 Text en © 2023 Zhang, Lyu, Wang, Shi, Zheng, Lyu and Yan. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY) (https://creativecommons.org/licenses/by/4.0/) . The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Surgery
Zhang, Miaomiao
Lyu, Xin
Wang, Juanjuan
Shi, Aihua
Zheng, Yunyun
Lyu, Yi
Yan, Xiaopeng
Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
title Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
title_full Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
title_fullStr Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
title_full_unstemmed Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
title_short Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
title_sort value of a novel y-z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion
topic Surgery
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10560742/
https://www.ncbi.nlm.nih.gov/pubmed/37818210
http://dx.doi.org/10.3389/fsurg.2023.1241780
work_keys_str_mv AT zhangmiaomiao valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion
AT lyuxin valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion
AT wangjuanjuan valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion
AT shiaihua valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion
AT zhengyunyun valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion
AT lyuyi valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion
AT yanxiaopeng valueofanovelyzmagnetictotallyimplantablevenousaccessportinimprovingthesuccessrateofonetimeneedleinsertion