Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience

PURPOSE: Magnetic resonance‐guided radiotherapy (MRgRT) is desired for the treatment of diseases in the abdominothoracic region, which has a broad imaging area and continuous motion. To ensure accurate treatment delivery, an effective image quality assurance (QA) program, with a phantom that covers...

Descripción completa

Detalles Bibliográficos
Autores principales: Marasini, Shanti, Zhang, Hailei, Dyke, Lara, Cole, Mike, Quinn, Benjamin, Curcuru, Austen, Gu, Bruce, Flores, Rocco, Kim, Taeho
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10562018/
https://www.ncbi.nlm.nih.gov/pubmed/37307238
http://dx.doi.org/10.1002/acm2.14066
_version_ 1785118035211714560
author Marasini, Shanti
Zhang, Hailei
Dyke, Lara
Cole, Mike
Quinn, Benjamin
Curcuru, Austen
Gu, Bruce
Flores, Rocco
Kim, Taeho
author_facet Marasini, Shanti
Zhang, Hailei
Dyke, Lara
Cole, Mike
Quinn, Benjamin
Curcuru, Austen
Gu, Bruce
Flores, Rocco
Kim, Taeho
author_sort Marasini, Shanti
collection PubMed
description PURPOSE: Magnetic resonance‐guided radiotherapy (MRgRT) is desired for the treatment of diseases in the abdominothoracic region, which has a broad imaging area and continuous motion. To ensure accurate treatment delivery, an effective image quality assurance (QA) program, with a phantom that covers the field of view (FOV) similar to a human torso, is required. However, routine image QA for a large FOV is not readily available at many MRgRT centers. In this work, we present the clinical experience of the large FOV MRgRT Insight phantom for periodic daily and monthly comprehensive magnetic resonance imaging (MRI)‐QA and its feasibility compared to the existing institutional routine MRI‐QA procedures in 0.35 T MRgRT. METHODS: Three phantoms; ViewRay cylindrical water phantom, Fluke 76–907 uniformity and linearity phantom, and Modus QA large FOV MRgRT Insight phantom, were imaged on the 0.35 T MR‐Linac. The measurements were made in MRI mode with the true fast imaging with steady‐state free precession (TRUFI) sequence. The ViewRay cylindrical water phantom was imaged in a single‐position setup whereas the Fluke phantom and Insight phantom were imaged in three different orientations: axial, sagittal, and coronal. Additionally, the phased array coil QA was performed using the horizontal base plate of the Insight phantom by placing the desired coil around the base section which was compared to an in‐house built Polyurethane foam phantom for reference. RESULT: The Insight phantom captured image artifacts across the entire planar field of view, up to 400 mm, in a single image acquisition, which is beyond the FOV of the conventional phantoms. The geometric distortion test showed a similar distortion of 0.45 ± 0.01  and 0.41 ± 0.01 mm near the isocenter, that is, within 300 mm lengths for Fluke and Insight phantoms, respectively, but showed higher geometric distortion of 0.8 ± 0.4 mm in the peripheral region between 300 and 400 mm of the imaging slice for the Insight phantom. The Insight phantom with multiple image quality features and its accompanying software utilized the modulation transform function (MTF) to evaluate the image spatial resolution. The average MTF values were 0.35 ± 0.01, 0.35 ± 0.01, and 0.34 ± 0.03 for axial, coronal, and sagittal images, respectively. The plane alignment and spatial accuracy of the ViewRay water phantom were measured manually. The phased array coil test for both the Insight phantom and the Polyurethane foam phantoms ensured the proper functionality of each coil element. CONCLUSION: The multifunctional large FOV Insight phantom helps in tracking MR imaging quality of the system to a larger extent compared to the routine daily and monthly QA phantoms currently used in our institute. Also, the Insight phantom is found to be more feasible for routine QA with easy setup.
format Online
Article
Text
id pubmed-10562018
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-105620182023-10-10 Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience Marasini, Shanti Zhang, Hailei Dyke, Lara Cole, Mike Quinn, Benjamin Curcuru, Austen Gu, Bruce Flores, Rocco Kim, Taeho J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: Magnetic resonance‐guided radiotherapy (MRgRT) is desired for the treatment of diseases in the abdominothoracic region, which has a broad imaging area and continuous motion. To ensure accurate treatment delivery, an effective image quality assurance (QA) program, with a phantom that covers the field of view (FOV) similar to a human torso, is required. However, routine image QA for a large FOV is not readily available at many MRgRT centers. In this work, we present the clinical experience of the large FOV MRgRT Insight phantom for periodic daily and monthly comprehensive magnetic resonance imaging (MRI)‐QA and its feasibility compared to the existing institutional routine MRI‐QA procedures in 0.35 T MRgRT. METHODS: Three phantoms; ViewRay cylindrical water phantom, Fluke 76–907 uniformity and linearity phantom, and Modus QA large FOV MRgRT Insight phantom, were imaged on the 0.35 T MR‐Linac. The measurements were made in MRI mode with the true fast imaging with steady‐state free precession (TRUFI) sequence. The ViewRay cylindrical water phantom was imaged in a single‐position setup whereas the Fluke phantom and Insight phantom were imaged in three different orientations: axial, sagittal, and coronal. Additionally, the phased array coil QA was performed using the horizontal base plate of the Insight phantom by placing the desired coil around the base section which was compared to an in‐house built Polyurethane foam phantom for reference. RESULT: The Insight phantom captured image artifacts across the entire planar field of view, up to 400 mm, in a single image acquisition, which is beyond the FOV of the conventional phantoms. The geometric distortion test showed a similar distortion of 0.45 ± 0.01  and 0.41 ± 0.01 mm near the isocenter, that is, within 300 mm lengths for Fluke and Insight phantoms, respectively, but showed higher geometric distortion of 0.8 ± 0.4 mm in the peripheral region between 300 and 400 mm of the imaging slice for the Insight phantom. The Insight phantom with multiple image quality features and its accompanying software utilized the modulation transform function (MTF) to evaluate the image spatial resolution. The average MTF values were 0.35 ± 0.01, 0.35 ± 0.01, and 0.34 ± 0.03 for axial, coronal, and sagittal images, respectively. The plane alignment and spatial accuracy of the ViewRay water phantom were measured manually. The phased array coil test for both the Insight phantom and the Polyurethane foam phantoms ensured the proper functionality of each coil element. CONCLUSION: The multifunctional large FOV Insight phantom helps in tracking MR imaging quality of the system to a larger extent compared to the routine daily and monthly QA phantoms currently used in our institute. Also, the Insight phantom is found to be more feasible for routine QA with easy setup. John Wiley and Sons Inc. 2023-06-12 /pmc/articles/PMC10562018/ /pubmed/37307238 http://dx.doi.org/10.1002/acm2.14066 Text en © 2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Marasini, Shanti
Zhang, Hailei
Dyke, Lara
Cole, Mike
Quinn, Benjamin
Curcuru, Austen
Gu, Bruce
Flores, Rocco
Kim, Taeho
Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience
title Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience
title_full Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience
title_fullStr Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience
title_full_unstemmed Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience
title_short Comprehensive MR imaging QA of 0.35 T MR‐Linac using a multi‐purpose large FOV phantom: A single‐institution experience
title_sort comprehensive mr imaging qa of 0.35 t mr‐linac using a multi‐purpose large fov phantom: a single‐institution experience
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10562018/
https://www.ncbi.nlm.nih.gov/pubmed/37307238
http://dx.doi.org/10.1002/acm2.14066
work_keys_str_mv AT marasinishanti comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT zhanghailei comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT dykelara comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT colemike comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT quinnbenjamin comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT curcuruausten comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT gubruce comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT floresrocco comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience
AT kimtaeho comprehensivemrimagingqaof035tmrlinacusingamultipurposelargefovphantomasingleinstitutionexperience