Cargando…
SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level
Objectives: We examined the washing evaluation method of the flusher disinfector installed in the medical site by visual evaluation using an ISO test soil and adenosine triphosphate (ATP) measurement. Methods: The test soil shown in ISO15883-5 was applied to the bedpan using 2 methods (N = 10) and w...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cambridge University Press
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10571205/ http://dx.doi.org/10.1017/ash.2023.103 |
_version_ | 1785119935887835136 |
---|---|
author | Shima, Mari Yoshida, Rika Okubo, Takashi |
author_facet | Shima, Mari Yoshida, Rika Okubo, Takashi |
author_sort | Shima, Mari |
collection | PubMed |
description | Objectives: We examined the washing evaluation method of the flusher disinfector installed in the medical site by visual evaluation using an ISO test soil and adenosine triphosphate (ATP) measurement. Methods: The test soil shown in ISO15883-5 was applied to the bedpan using 2 methods (N = 10) and was then visually evaluated using a 4-step scale. The ATP value on the surface of the bedpan corresponding to each scale was measured, and the correlation with the visual scale was confirmed. In addition, a visual evaluation was performed using a different flusher disinfector and bedpan (N = 3). Results: In the visual evaluation, when the test soil was applied to the entire surface of the bedpan, it remained in the parts where the water flow was hard to hit. When the test soil was applied to the surface of the bedpan except the lid, no soil remained. The 4-step visual evaluation scale and the logarithm of ATP value showed a positive correlation. In visual evaluations with different flusher disinfector and bedpan combinations, the residual soil patterns tended to be different. Some bedpans showed washing failure related to improper design. Conclusions: Poor cleaning of flusher disinfectors may occur when the amount of water used is insufficient or when the bedpan is significantly contaminated. It is important to carry out flusher disinfector washing evaluation at medical sites to evaluate the function of flusher disinfectors. An appropriate program must be utilized for staff education and appropriate management. Because flusher disinfectors and bedpans differ, flusher disinfector cleaning evaluations should be carried out at all facilities. |
format | Online Article Text |
id | pubmed-10571205 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Cambridge University Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-105712052023-10-14 SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level Shima, Mari Yoshida, Rika Okubo, Takashi Antimicrob Steward Healthc Epidemiol Sterilization and Disinfection Objectives: We examined the washing evaluation method of the flusher disinfector installed in the medical site by visual evaluation using an ISO test soil and adenosine triphosphate (ATP) measurement. Methods: The test soil shown in ISO15883-5 was applied to the bedpan using 2 methods (N = 10) and was then visually evaluated using a 4-step scale. The ATP value on the surface of the bedpan corresponding to each scale was measured, and the correlation with the visual scale was confirmed. In addition, a visual evaluation was performed using a different flusher disinfector and bedpan (N = 3). Results: In the visual evaluation, when the test soil was applied to the entire surface of the bedpan, it remained in the parts where the water flow was hard to hit. When the test soil was applied to the surface of the bedpan except the lid, no soil remained. The 4-step visual evaluation scale and the logarithm of ATP value showed a positive correlation. In visual evaluations with different flusher disinfector and bedpan combinations, the residual soil patterns tended to be different. Some bedpans showed washing failure related to improper design. Conclusions: Poor cleaning of flusher disinfectors may occur when the amount of water used is insufficient or when the bedpan is significantly contaminated. It is important to carry out flusher disinfector washing evaluation at medical sites to evaluate the function of flusher disinfectors. An appropriate program must be utilized for staff education and appropriate management. Because flusher disinfectors and bedpans differ, flusher disinfector cleaning evaluations should be carried out at all facilities. Cambridge University Press 2023-03-16 /pmc/articles/PMC10571205/ http://dx.doi.org/10.1017/ash.2023.103 Text en © The Society for Healthcare Epidemiology of America 2023 https://creativecommons.org/licenses/by/4.0/This is an Open Access article, distributed under the terms of the Creative Commons Attribution licence (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted re-use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Sterilization and Disinfection Shima, Mari Yoshida, Rika Okubo, Takashi SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level |
title | SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level |
title_full | SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level |
title_fullStr | SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level |
title_full_unstemmed | SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level |
title_short | SG-APSIC1024: The importance of the washing evaluation of flusher disinfector in the medical site: Visual evaluation with the ISO standardized test soil and adenosine triphosphate level |
title_sort | sg-apsic1024: the importance of the washing evaluation of flusher disinfector in the medical site: visual evaluation with the iso standardized test soil and adenosine triphosphate level |
topic | Sterilization and Disinfection |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10571205/ http://dx.doi.org/10.1017/ash.2023.103 |
work_keys_str_mv | AT shimamari sgapsic1024theimportanceofthewashingevaluationofflusherdisinfectorinthemedicalsitevisualevaluationwiththeisostandardizedtestsoilandadenosinetriphosphatelevel AT yoshidarika sgapsic1024theimportanceofthewashingevaluationofflusherdisinfectorinthemedicalsitevisualevaluationwiththeisostandardizedtestsoilandadenosinetriphosphatelevel AT okubotakashi sgapsic1024theimportanceofthewashingevaluationofflusherdisinfectorinthemedicalsitevisualevaluationwiththeisostandardizedtestsoilandadenosinetriphosphatelevel |