Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients

SIMPLE SUMMARY: Ionizing radiotherapy (RT) can cause malfunctions to cardiac implantable electronic devices (CIEDs), posing a risk to patient safety. The aim of the study was to quantify and categorize the risk of CIED malfunction during RT and to analyze national guidelines/recommendations for adeq...

Descripción completa

Detalles Bibliográficos
Autores principales: Lisowski, Dominik, Lutyj, Paul, Abazari, Arya, Weick, Stefan, Traub, Jan, Polat, Bülent, Flentje, Michael, Kraft, Johannes
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10571836/
https://www.ncbi.nlm.nih.gov/pubmed/37835524
http://dx.doi.org/10.3390/cancers15194830
_version_ 1785120094725079040
author Lisowski, Dominik
Lutyj, Paul
Abazari, Arya
Weick, Stefan
Traub, Jan
Polat, Bülent
Flentje, Michael
Kraft, Johannes
author_facet Lisowski, Dominik
Lutyj, Paul
Abazari, Arya
Weick, Stefan
Traub, Jan
Polat, Bülent
Flentje, Michael
Kraft, Johannes
author_sort Lisowski, Dominik
collection PubMed
description SIMPLE SUMMARY: Ionizing radiotherapy (RT) can cause malfunctions to cardiac implantable electronic devices (CIEDs), posing a risk to patient safety. The aim of the study was to quantify and categorize the risk of CIED malfunction during RT and to analyze national guidelines/recommendations for adequate patient safety. We found that the CIED malfunction rate is low (2.8%) in cancer patients receiving RT. The probability of CIED malfunction does not correlate with the applied photon energy, the dose at the CIED nor the treated area. Battery depletion occurs very rarely (1.2%). The national guidelines strongly differ in regard to safety the recommendations for patients with a CIED during and after RT. ABSTRACT: Purpose: This study analyses a large number of cancer patients with CIEDs for device malfunction and premature battery depletion by device interrogation after each radiotherapy fraction and compares different guidelines in regard to patient safety. Methods: From 2007 to 2022, a cohort of 255 patients was analyzed for CIED malfunctions via immediate device interrogation after every RT fraction. Results: Out of 324 series of radiotherapy treatments, with a total number of 5742 CIED interrogations, nine device malfunctions (2.8%) occurred. Switching into back-up/safety mode and software errors occurred four times each. Once, automatic read-out could not be performed. The median prescribed cumulative dose at planning target volume (PTV) associated with CIED malfunction was 45.0 Gy (IQR 36.0–64.0 Gy), with a median dose per fraction of 2.31 Gy (IQR 2.0–3.0 Gy). The median maximum dose at the CIED at time of malfunction was 0.3 Gy (IQR 0.0–1.3 Gy). No correlation between CIED malfunction and maximum photon energy (p = 0.07), maximum dose at the CIED (p = 0.59) nor treatment localization (p = 0.41) could be detected. After excluding the nine malfunctions, premature battery depletion was only observed three times (1.2%). Depending on the national guidelines, 1–9 CIED malfunctions in this study would have been detected on the day of occurrence and in none of the cases would patient safety have been compromised. Conclusion: Radiation-induced malfunctions of CIEDs and premature battery depletion are rare. If recommendations of national safety guidelines are followed, only a portion of the malfunctions would be detected directly after occurrence. Nevertheless, patient safety would not be compromised.
format Online
Article
Text
id pubmed-10571836
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-105718362023-10-14 Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients Lisowski, Dominik Lutyj, Paul Abazari, Arya Weick, Stefan Traub, Jan Polat, Bülent Flentje, Michael Kraft, Johannes Cancers (Basel) Article SIMPLE SUMMARY: Ionizing radiotherapy (RT) can cause malfunctions to cardiac implantable electronic devices (CIEDs), posing a risk to patient safety. The aim of the study was to quantify and categorize the risk of CIED malfunction during RT and to analyze national guidelines/recommendations for adequate patient safety. We found that the CIED malfunction rate is low (2.8%) in cancer patients receiving RT. The probability of CIED malfunction does not correlate with the applied photon energy, the dose at the CIED nor the treated area. Battery depletion occurs very rarely (1.2%). The national guidelines strongly differ in regard to safety the recommendations for patients with a CIED during and after RT. ABSTRACT: Purpose: This study analyses a large number of cancer patients with CIEDs for device malfunction and premature battery depletion by device interrogation after each radiotherapy fraction and compares different guidelines in regard to patient safety. Methods: From 2007 to 2022, a cohort of 255 patients was analyzed for CIED malfunctions via immediate device interrogation after every RT fraction. Results: Out of 324 series of radiotherapy treatments, with a total number of 5742 CIED interrogations, nine device malfunctions (2.8%) occurred. Switching into back-up/safety mode and software errors occurred four times each. Once, automatic read-out could not be performed. The median prescribed cumulative dose at planning target volume (PTV) associated with CIED malfunction was 45.0 Gy (IQR 36.0–64.0 Gy), with a median dose per fraction of 2.31 Gy (IQR 2.0–3.0 Gy). The median maximum dose at the CIED at time of malfunction was 0.3 Gy (IQR 0.0–1.3 Gy). No correlation between CIED malfunction and maximum photon energy (p = 0.07), maximum dose at the CIED (p = 0.59) nor treatment localization (p = 0.41) could be detected. After excluding the nine malfunctions, premature battery depletion was only observed three times (1.2%). Depending on the national guidelines, 1–9 CIED malfunctions in this study would have been detected on the day of occurrence and in none of the cases would patient safety have been compromised. Conclusion: Radiation-induced malfunctions of CIEDs and premature battery depletion are rare. If recommendations of national safety guidelines are followed, only a portion of the malfunctions would be detected directly after occurrence. Nevertheless, patient safety would not be compromised. MDPI 2023-10-02 /pmc/articles/PMC10571836/ /pubmed/37835524 http://dx.doi.org/10.3390/cancers15194830 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Lisowski, Dominik
Lutyj, Paul
Abazari, Arya
Weick, Stefan
Traub, Jan
Polat, Bülent
Flentje, Michael
Kraft, Johannes
Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
title Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
title_full Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
title_fullStr Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
title_full_unstemmed Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
title_short Impact of Radiotherapy on Malfunctions and Battery Life of Cardiac Implantable Electronic Devices in Cancer Patients
title_sort impact of radiotherapy on malfunctions and battery life of cardiac implantable electronic devices in cancer patients
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10571836/
https://www.ncbi.nlm.nih.gov/pubmed/37835524
http://dx.doi.org/10.3390/cancers15194830
work_keys_str_mv AT lisowskidominik impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT lutyjpaul impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT abazariarya impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT weickstefan impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT traubjan impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT polatbulent impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT flentjemichael impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients
AT kraftjohannes impactofradiotherapyonmalfunctionsandbatterylifeofcardiacimplantableelectronicdevicesincancerpatients