Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study
This study aimed to estimate scattered radiation and its spatial distribution around three cone-beam computed tomography (CBCT) devices, in order to determine potential positions for an operator to stand if they needed to be inside the CBCT room. The following devices were tested: Morita Accuitomo (...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10573760/ https://www.ncbi.nlm.nih.gov/pubmed/37834843 http://dx.doi.org/10.3390/jcm12196199 |
_version_ | 1785120536075960320 |
---|---|
author | Petsaros, Sotirios Chatzipetros, Emmanouil Donta, Catherine Karaiskos, Pantelis Boziari, Argiro Papadakis, Evangelos Angelopoulos, Christos |
author_facet | Petsaros, Sotirios Chatzipetros, Emmanouil Donta, Catherine Karaiskos, Pantelis Boziari, Argiro Papadakis, Evangelos Angelopoulos, Christos |
author_sort | Petsaros, Sotirios |
collection | PubMed |
description | This study aimed to estimate scattered radiation and its spatial distribution around three cone-beam computed tomography (CBCT) devices, in order to determine potential positions for an operator to stand if they needed to be inside the CBCT room. The following devices were tested: Morita Accuitomo (CBCT1), Newtom Giano HR (CBCT2), Newtom VGi (CBCT3). Scattered radiation measurements were performed using different kVp, mA, and Field of View (FOV) options. An anthropomorphic phantom (NATHANIA) was placed inside the X-ray gantry to simulate clinical conditions. Scattered measurements were taken with the Inovision model 451P Victoreen ionization chamber once placed at fixed distances from each irradiation isocenter, away from the primary beam. A statistically significant (p < 0.001) difference was found in the mean value of the scattered radiation estimations between the CBCT devices. Scattered radiation was reduced with a different rate for each CBCT device as distance was increased. For CBCT1 the reduction was 0.047 μGy, for CBCT2 it was 0.036 μGy, and for CBCT3 it was 0.079 μGy, for every one meter from the X-ray gantry. Therefore, at certain distances from the central X-ray, the scattered radiation was below the critical level of 1 mGy, which is defined by the radiation protection guidelines as the exposure radiation limit of the general population. Consequently, an operator could stay inside the room accompanying the patient being scanned, if necessary. |
format | Online Article Text |
id | pubmed-10573760 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-105737602023-10-14 Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study Petsaros, Sotirios Chatzipetros, Emmanouil Donta, Catherine Karaiskos, Pantelis Boziari, Argiro Papadakis, Evangelos Angelopoulos, Christos J Clin Med Article This study aimed to estimate scattered radiation and its spatial distribution around three cone-beam computed tomography (CBCT) devices, in order to determine potential positions for an operator to stand if they needed to be inside the CBCT room. The following devices were tested: Morita Accuitomo (CBCT1), Newtom Giano HR (CBCT2), Newtom VGi (CBCT3). Scattered radiation measurements were performed using different kVp, mA, and Field of View (FOV) options. An anthropomorphic phantom (NATHANIA) was placed inside the X-ray gantry to simulate clinical conditions. Scattered measurements were taken with the Inovision model 451P Victoreen ionization chamber once placed at fixed distances from each irradiation isocenter, away from the primary beam. A statistically significant (p < 0.001) difference was found in the mean value of the scattered radiation estimations between the CBCT devices. Scattered radiation was reduced with a different rate for each CBCT device as distance was increased. For CBCT1 the reduction was 0.047 μGy, for CBCT2 it was 0.036 μGy, and for CBCT3 it was 0.079 μGy, for every one meter from the X-ray gantry. Therefore, at certain distances from the central X-ray, the scattered radiation was below the critical level of 1 mGy, which is defined by the radiation protection guidelines as the exposure radiation limit of the general population. Consequently, an operator could stay inside the room accompanying the patient being scanned, if necessary. MDPI 2023-09-26 /pmc/articles/PMC10573760/ /pubmed/37834843 http://dx.doi.org/10.3390/jcm12196199 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Petsaros, Sotirios Chatzipetros, Emmanouil Donta, Catherine Karaiskos, Pantelis Boziari, Argiro Papadakis, Evangelos Angelopoulos, Christos Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study |
title | Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study |
title_full | Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study |
title_fullStr | Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study |
title_full_unstemmed | Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study |
title_short | Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study |
title_sort | scattered radiation distribution utilizing three different cone-beam computed tomography devices for maxillofacial diagnostics: a research study |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10573760/ https://www.ncbi.nlm.nih.gov/pubmed/37834843 http://dx.doi.org/10.3390/jcm12196199 |
work_keys_str_mv | AT petsarossotirios scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy AT chatzipetrosemmanouil scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy AT dontacatherine scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy AT karaiskospantelis scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy AT boziariargiro scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy AT papadakisevangelos scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy AT angelopouloschristos scatteredradiationdistributionutilizingthreedifferentconebeamcomputedtomographydevicesformaxillofacialdiagnosticsaresearchstudy |