Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone
There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upo...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10575311/ https://www.ncbi.nlm.nih.gov/pubmed/37837170 http://dx.doi.org/10.3390/s23198340 |
_version_ | 1785120894772838400 |
---|---|
author | Tomazini, Žan Geršak, Gregor Beguš, Samo |
author_facet | Tomazini, Žan Geršak, Gregor Beguš, Samo |
author_sort | Tomazini, Žan |
collection | PubMed |
description | There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards. |
format | Online Article Text |
id | pubmed-10575311 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-105753112023-10-14 Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone Tomazini, Žan Geršak, Gregor Beguš, Samo Sensors (Basel) Article There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards. MDPI 2023-10-09 /pmc/articles/PMC10575311/ /pubmed/37837170 http://dx.doi.org/10.3390/s23198340 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Tomazini, Žan Geršak, Gregor Beguš, Samo Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone |
title | Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone |
title_full | Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone |
title_fullStr | Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone |
title_full_unstemmed | Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone |
title_short | Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone |
title_sort | sphygmomanometer dynamic pressure measurement using a condenser microphone |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10575311/ https://www.ncbi.nlm.nih.gov/pubmed/37837170 http://dx.doi.org/10.3390/s23198340 |
work_keys_str_mv | AT tomazinizan sphygmomanometerdynamicpressuremeasurementusingacondensermicrophone AT gersakgregor sphygmomanometerdynamicpressuremeasurementusingacondensermicrophone AT begussamo sphygmomanometerdynamicpressuremeasurementusingacondensermicrophone |