Cargando…

Fiber-optic hydrophone for detection of high-intensity ultrasound waves

Fiber-optic hydrophones (FOHs) are widely used to detect high-intensity focused ultrasound (HIFU) fields. The most common type consists of an uncoated single-mode fiber with a perpendicularly cleaved end face. The main disadvantage of these hydrophones is their low signal-to-noise ratio (SNR). To in...

Descripción completa

Detalles Bibliográficos
Autores principales: Aytac Kipergil, Esra, Martin, Eleanor, Mathews, Sunish J., Papakonstantinou, Ioannis, Alles, Erwin J., Desjardins, Adrien E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Optica Publishing Group 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10575604/
https://www.ncbi.nlm.nih.gov/pubmed/37186722
http://dx.doi.org/10.1364/OL.488862
Descripción
Sumario:Fiber-optic hydrophones (FOHs) are widely used to detect high-intensity focused ultrasound (HIFU) fields. The most common type consists of an uncoated single-mode fiber with a perpendicularly cleaved end face. The main disadvantage of these hydrophones is their low signal-to-noise ratio (SNR). To increase the SNR, signal averaging is performed, but the associated increased acquisition times hinder ultrasound field scans. In this study, with a view to increasing SNR while withstanding HIFU pressures, the bare FOH paradigm is extended to include a partially reflective coating on the fiber end face. Here, a numerical model based on the general transfer-matrix method was implemented. Based on the simulation results, a single-layer, 172 nm TiO(2)-coated FOH was fabricated. The frequency range of the hydrophone was verified from 1 to 30 MHz. The SNR of the acoustic measurement with the coated sensor was 21 dB higher than that of the uncoated one. The coated sensor successfully withstood a peak positive pressure of 35 MPa for 6000 pulses.