Cargando…

The effect of pupil size on the measurement of corneal birefringence properties: preliminary study

We used a partial Mueller matrix polarimeter to measure the corneal anisotropic properties at three pupil sizes (dilated, natural, and constricted). The geometrical parameters of first order isochromes were described by quadrilaterals. These parameters are statistically significantly different betwe...

Descripción completa

Detalles Bibliográficos
Autores principales: Sobczak, Marcelina, Asejczyk, Magdalena, Wilczyński, Maciej
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10576786/
https://www.ncbi.nlm.nih.gov/pubmed/37838761
http://dx.doi.org/10.1038/s41598-023-44706-2
Descripción
Sumario:We used a partial Mueller matrix polarimeter to measure the corneal anisotropic properties at three pupil sizes (dilated, natural, and constricted). The geometrical parameters of first order isochromes were described by quadrilaterals. These parameters are statistically significantly different between the three pupil sizes. The pupillary size changes do not influence the azimuth angle distribution α. The retardation R and birefringence distributions show asymmetry in the nasal–temporal cross-section. There are differences between pupil sizes for both nasal and temporal parts of the cornea for these distributions. Iridial light scattering and diffraction might be the reason for these differences.