Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom

BACKGROUND: Osteoporosis is a systemic bone disease which can increase the risk of osteoporotic fractures. Dual-energy X-ray absorptiometry (DXA) is considered as the clinical standard for diagnosing osteoporosis by detecting the bone mineral density (BMD) in patients, but it has flaws in distinguis...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, Feng, Zhu, Hongmei, Ma, Jinlian, Miao, Liqiong, Chen, Shuang, Yin, Zijie, Wang, Huan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10578019/
https://www.ncbi.nlm.nih.gov/pubmed/37845720
http://dx.doi.org/10.1186/s13018-023-04174-w
_version_ 1785121434876510208
author Liu, Feng
Zhu, Hongmei
Ma, Jinlian
Miao, Liqiong
Chen, Shuang
Yin, Zijie
Wang, Huan
author_facet Liu, Feng
Zhu, Hongmei
Ma, Jinlian
Miao, Liqiong
Chen, Shuang
Yin, Zijie
Wang, Huan
author_sort Liu, Feng
collection PubMed
description BACKGROUND: Osteoporosis is a systemic bone disease which can increase the risk of osteoporotic fractures. Dual-energy X-ray absorptiometry (DXA) is considered as the clinical standard for diagnosing osteoporosis by detecting the bone mineral density (BMD) in patients, but it has flaws in distinguishing between calcification and other degenerative diseases, thus leading to inaccurate BMD levels in subjects. Mindways quantitative computed tomography (Mindways QCT) is a classical QCT system. Similar to DXA, Mindways QCT can directly present the density of trabecular bone, vascular or tissue calcification; therefore, it is more accurate and sensitive than DXA and has been widely applied in clinic to evaluate osteoporosis. iCare QCT osteodensitometry was a new phantom-based QCT system, recently developed by iCare Inc. (China). It has been gradually applied in clinic by its superiority of taking 3-dimensional BMD of bone and converting BMD values to T value automatically. This study aimed at evaluating the osteoporosis detection rate of iCare QCT, compared with synchronous Mindways QCT (USA). METHODS: In this study, 131 patients who underwent hip phantom-based CT scan were included. Bone mineral density (BMD) of the unified region of interests (ROI) defined at the European spine phantom (ESP, German QRM) including L1 (low), L2 (medium), and L3 (high) vertebral bodies was detected for QCT quality control and horizontal calibration. Every ESP scan were taken for 10 times, and the mean BMD values measured by iCare QCT and Mindways QCT were compared. Hip CT scan was conducted with ESP as calibration individually. T-scores gained from iCare QCT and Mindways QCT were analyzed with Pearson correlation test. The detection rates of osteoporosis were compared between iCare QCT and Mindways QCT. The unified region of interests (ROI) was delineated in the QCT software. RESULTS: The results showed that there was no significant difference between iCare QCT and Mindways QCT in the evaluation of L1, L2, and L3 vertebrae bodies in ESP. A strong correlation between iCare QCT and Mindways QCT in the assessment of hip T-score was found. It was illustrated that iCare QCT had a higher detection rate of osteoporosis with the assessment of hip T-score than Mindways QCT did. In patients < 50 years subgroup, the detection rate of osteoporosis with iCare QCT and Mindways QCT was equal. In patients ≥ 50 years subgroup, the detection rate of osteoporosis with iCare QCT (35/92, 38.0%) was higher than that with Mindways QCT. In female subgroup, the detection rate of osteoporosis with iCare QCT was significantly higher than Mindways QCT. In male subgroup, the detection rate of osteoporosis with iCare QCT was also markedly higher than Mindways QCT. The detection rate of osteoporosis by iCare QCT was higher than Mindways QCT with hip bone assessment. Of course, the results of the present study remain to be further verified by multicenter studies in the future.
format Online
Article
Text
id pubmed-10578019
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-105780192023-10-17 Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom Liu, Feng Zhu, Hongmei Ma, Jinlian Miao, Liqiong Chen, Shuang Yin, Zijie Wang, Huan J Orthop Surg Res Research Article BACKGROUND: Osteoporosis is a systemic bone disease which can increase the risk of osteoporotic fractures. Dual-energy X-ray absorptiometry (DXA) is considered as the clinical standard for diagnosing osteoporosis by detecting the bone mineral density (BMD) in patients, but it has flaws in distinguishing between calcification and other degenerative diseases, thus leading to inaccurate BMD levels in subjects. Mindways quantitative computed tomography (Mindways QCT) is a classical QCT system. Similar to DXA, Mindways QCT can directly present the density of trabecular bone, vascular or tissue calcification; therefore, it is more accurate and sensitive than DXA and has been widely applied in clinic to evaluate osteoporosis. iCare QCT osteodensitometry was a new phantom-based QCT system, recently developed by iCare Inc. (China). It has been gradually applied in clinic by its superiority of taking 3-dimensional BMD of bone and converting BMD values to T value automatically. This study aimed at evaluating the osteoporosis detection rate of iCare QCT, compared with synchronous Mindways QCT (USA). METHODS: In this study, 131 patients who underwent hip phantom-based CT scan were included. Bone mineral density (BMD) of the unified region of interests (ROI) defined at the European spine phantom (ESP, German QRM) including L1 (low), L2 (medium), and L3 (high) vertebral bodies was detected for QCT quality control and horizontal calibration. Every ESP scan were taken for 10 times, and the mean BMD values measured by iCare QCT and Mindways QCT were compared. Hip CT scan was conducted with ESP as calibration individually. T-scores gained from iCare QCT and Mindways QCT were analyzed with Pearson correlation test. The detection rates of osteoporosis were compared between iCare QCT and Mindways QCT. The unified region of interests (ROI) was delineated in the QCT software. RESULTS: The results showed that there was no significant difference between iCare QCT and Mindways QCT in the evaluation of L1, L2, and L3 vertebrae bodies in ESP. A strong correlation between iCare QCT and Mindways QCT in the assessment of hip T-score was found. It was illustrated that iCare QCT had a higher detection rate of osteoporosis with the assessment of hip T-score than Mindways QCT did. In patients < 50 years subgroup, the detection rate of osteoporosis with iCare QCT and Mindways QCT was equal. In patients ≥ 50 years subgroup, the detection rate of osteoporosis with iCare QCT (35/92, 38.0%) was higher than that with Mindways QCT. In female subgroup, the detection rate of osteoporosis with iCare QCT was significantly higher than Mindways QCT. In male subgroup, the detection rate of osteoporosis with iCare QCT was also markedly higher than Mindways QCT. The detection rate of osteoporosis by iCare QCT was higher than Mindways QCT with hip bone assessment. Of course, the results of the present study remain to be further verified by multicenter studies in the future. BioMed Central 2023-10-16 /pmc/articles/PMC10578019/ /pubmed/37845720 http://dx.doi.org/10.1186/s13018-023-04174-w Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research Article
Liu, Feng
Zhu, Hongmei
Ma, Jinlian
Miao, Liqiong
Chen, Shuang
Yin, Zijie
Wang, Huan
Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom
title Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom
title_full Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom
title_fullStr Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom
title_full_unstemmed Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom
title_short Performance of iCare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in European spine phantom
title_sort performance of icare quantitative computed tomography in bone mineral density assessment of the hip and vertebral bodies in european spine phantom
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10578019/
https://www.ncbi.nlm.nih.gov/pubmed/37845720
http://dx.doi.org/10.1186/s13018-023-04174-w
work_keys_str_mv AT liufeng performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom
AT zhuhongmei performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom
AT majinlian performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom
AT miaoliqiong performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom
AT chenshuang performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom
AT yinzijie performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom
AT wanghuan performanceoficarequantitativecomputedtomographyinbonemineraldensityassessmentofthehipandvertebralbodiesineuropeanspinephantom