Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant
BACKGROUND: Patients with glaucoma on topical glaucoma medication are often affected by dry eye symptoms and thus likely to rub or squeeze their eyelids. Here, we telemetrically measure peak intraocular pressure (IOP) during eyelid manoeuvres and eyelid rubbing. METHODS: Eleven patients with primary...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BMJ Publishing Group
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10579178/ https://www.ncbi.nlm.nih.gov/pubmed/35701079 http://dx.doi.org/10.1136/bjophthalmol-2021-320508 |
_version_ | 1785121669324472320 |
---|---|
author | van den Bosch, Jacqueline J O N Pennisi, Vincenzo Mansouri, Kaweh Weinreb, Robert N Thieme, Hagen Hoffmann, Michael B Choritz, Lars |
author_facet | van den Bosch, Jacqueline J O N Pennisi, Vincenzo Mansouri, Kaweh Weinreb, Robert N Thieme, Hagen Hoffmann, Michael B Choritz, Lars |
author_sort | van den Bosch, Jacqueline J O N |
collection | PubMed |
description | BACKGROUND: Patients with glaucoma on topical glaucoma medication are often affected by dry eye symptoms and thus likely to rub or squeeze their eyelids. Here, we telemetrically measure peak intraocular pressure (IOP) during eyelid manoeuvres and eyelid rubbing. METHODS: Eleven patients with primary open-angle glaucoma (POAG) previously implanted with a telemetric IOP sensor (Eyemate-IO) were instructed to look straight ahead for 1 min as a baseline measurement. Next, 6 repeats of blinking on instruction with 10 s intervals in between were performed. In addition, 5 repeats of eyelid closure (n=9), eyelid squeezing and eyelid rubbing (n=7) were performed with 15 s intervals in between. IOP was recorded via an external antenna placed around the study eye. Average peak IOP increases from baseline were analysed and tested against zero (no change) with one-sample t-tests. RESULTS: For eyelid rubbing, the average peak ∆ IOP increase (mean±SEM) was 59.1±9.6 mm Hg (p<0.001) from baseline. It was 42.2±5.8 mm Hg (p<0.0001) for eyelid squeezing, 3.8±0.6 mm Hg (n=9, p<0.01) for eyelid closure and 11.6±2.4 mm Hg (p<0.001) for voluntary blinking. No IOP change except for a short irregularity in the ocular pulse was observed during involuntary blinking. CONCLUSION: Eyelid manoeuvres in patients with POAG elicited brief increases in IOP that were particularly large with squeezing and rubbing. Further investigation of the potential implications for glaucoma progression is warranted. |
format | Online Article Text |
id | pubmed-10579178 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | BMJ Publishing Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-105791782023-10-18 Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant van den Bosch, Jacqueline J O N Pennisi, Vincenzo Mansouri, Kaweh Weinreb, Robert N Thieme, Hagen Hoffmann, Michael B Choritz, Lars Br J Ophthalmol Clinical Science BACKGROUND: Patients with glaucoma on topical glaucoma medication are often affected by dry eye symptoms and thus likely to rub or squeeze their eyelids. Here, we telemetrically measure peak intraocular pressure (IOP) during eyelid manoeuvres and eyelid rubbing. METHODS: Eleven patients with primary open-angle glaucoma (POAG) previously implanted with a telemetric IOP sensor (Eyemate-IO) were instructed to look straight ahead for 1 min as a baseline measurement. Next, 6 repeats of blinking on instruction with 10 s intervals in between were performed. In addition, 5 repeats of eyelid closure (n=9), eyelid squeezing and eyelid rubbing (n=7) were performed with 15 s intervals in between. IOP was recorded via an external antenna placed around the study eye. Average peak IOP increases from baseline were analysed and tested against zero (no change) with one-sample t-tests. RESULTS: For eyelid rubbing, the average peak ∆ IOP increase (mean±SEM) was 59.1±9.6 mm Hg (p<0.001) from baseline. It was 42.2±5.8 mm Hg (p<0.0001) for eyelid squeezing, 3.8±0.6 mm Hg (n=9, p<0.01) for eyelid closure and 11.6±2.4 mm Hg (p<0.001) for voluntary blinking. No IOP change except for a short irregularity in the ocular pulse was observed during involuntary blinking. CONCLUSION: Eyelid manoeuvres in patients with POAG elicited brief increases in IOP that were particularly large with squeezing and rubbing. Further investigation of the potential implications for glaucoma progression is warranted. BMJ Publishing Group 2023-10 2022-06-14 /pmc/articles/PMC10579178/ /pubmed/35701079 http://dx.doi.org/10.1136/bjophthalmol-2021-320508 Text en © Author(s) (or their employer(s)) 2023. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. https://creativecommons.org/licenses/by-nc/4.0/This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) . |
spellingShingle | Clinical Science van den Bosch, Jacqueline J O N Pennisi, Vincenzo Mansouri, Kaweh Weinreb, Robert N Thieme, Hagen Hoffmann, Michael B Choritz, Lars Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant |
title | Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant |
title_full | Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant |
title_fullStr | Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant |
title_full_unstemmed | Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant |
title_short | Effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an IOP sensor implant |
title_sort | effect of eyelid muscle action and rubbing on telemetrically obtained intraocular pressure in patients with glaucoma with an iop sensor implant |
topic | Clinical Science |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10579178/ https://www.ncbi.nlm.nih.gov/pubmed/35701079 http://dx.doi.org/10.1136/bjophthalmol-2021-320508 |
work_keys_str_mv | AT vandenboschjacquelinejon effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant AT pennisivincenzo effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant AT mansourikaweh effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant AT weinrebrobertn effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant AT thiemehagen effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant AT hoffmannmichaelb effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant AT choritzlars effectofeyelidmuscleactionandrubbingontelemetricallyobtainedintraocularpressureinpatientswithglaucomawithaniopsensorimplant |