Cargando…
Biomechanical causes for failure of the Physiomesh/Securestrap system
This study investigates the mechanical behavior of the Physiomesh/Securestrap system, a hernia repair system used for IPOM procedures associated with high failure rates. The study involved conducting mechanical experiments and numerical simulations to investigate the mechanical behavior of the Physi...
Autores principales: | Zamkowski, Mateusz, Tomaszewska, Agnieszka, Lubowiecka, Izabela, Śmietański, Maciej |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10579252/ https://www.ncbi.nlm.nih.gov/pubmed/37845369 http://dx.doi.org/10.1038/s41598-023-44940-8 |
Ejemplares similares
-
Ventralight ST and SorbaFix Versus Physiomesh and Securestrap in a Porcine Model
por: Deeken, Corey R., et al.
Publicado: (2013) -
MEsh FIxation in Laparoendsocopic Repair of Large M3 inguinal hernias: multicenter, double-blinded, randomized controlled trial—study protocol for a MEFI Trial
por: Zamkowski, Mateusz, et al.
Publicado: (2023) -
Preclinical evaluation of the effect of the combined use of the Ethicon Securestrap(®) Open Absorbable Strap Fixation Device and Ethicon Physiomesh™ Open Flexible Composite Mesh Device on surgeon stress during ventral hernia repair
por: Sutton, Nadia, et al.
Publicado: (2017) -
Preclinical evaluation of the effect of the combined use of the Ethicon Securestrap(®) Open Absorbable Strap Fixation Device and Ethicon Physiomesh™ Open Flexible Composite Mesh Device on surgeon stress during ventral hernia repair [Corrigendum]
Publicado: (2018) -
Post-partum abdominal wall insufficiency syndrome (PPAWIS): lessons learned from a single surgeon’s experience based on 200 cases
por: Śmietański, Maciej, et al.
Publicado: (2022)