3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
Objective In this study, we investigated the optimal reconstruction algorithm in fluorodeoxyglucose (FDG) positron emission tomography (PET) with a short acquisition time. Materials and Methods In the phantom study, six spheres filled with FDG solution (sphere size: 6.23–37 mm; radioactivity ratio...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Thieme Medical and Scientific Publishers Pvt. Ltd.
2023
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10581748/ https://www.ncbi.nlm.nih.gov/pubmed/37854086 http://dx.doi.org/10.1055/s-0043-1774418 |
_version_ | 1785122196371275776 |
---|---|
author | Tsuda, Keisuke Suzuki, Takayuki Toya, Kazuhito Sato, Eisuke Fujii, Hirofumi |
author_facet | Tsuda, Keisuke Suzuki, Takayuki Toya, Kazuhito Sato, Eisuke Fujii, Hirofumi |
author_sort | Tsuda, Keisuke |
collection | PubMed |
description | Objective In this study, we investigated the optimal reconstruction algorithm in fluorodeoxyglucose (FDG) positron emission tomography (PET) with a short acquisition time. Materials and Methods In the phantom study, six spheres filled with FDG solution (sphere size: 6.23–37 mm; radioactivity ratio of spheres to background = 8:1) and placed in a National Electrical Manufacturers Association phantom were evaluated. Image acquisition time was 15 to 180 seconds, and the obtained image data were reconstructed using each of the Fourier rebinning (FORE) + ordered subsets expectation-maximization (OSEM) and 3D-OSEM algorithms. In the clinical study, mid-abdominal images of 19 patients were evaluated using regions of interest placed on areas of low, intermediate, and high radioactivity. All obtained images were investigated visually, and quantitatively using maximum standardized uptake value (SUV) and coefficient of variation (CV). Results In the phantom study, FORE + OSEM images with a short acquisition time had large CVs (poor image quality) but comparatively constant maximum SUVs. 3D-OSEM images showed comparatively constant CVs (good image quality) but significantly low maximum SUVs. The results of visual evaluation were well correlated with those of quantitative evaluation. Small spheres were obscured on 3D-OSEM images with short acquisition time, but image quality was not greatly deteriorated. The clinical and phantom studies yielded similar results. Conclusion FDG PET images with a short acquisition time reconstructed by FORE + OSEM showed poorer image quality than by 3D-OSEM. However, images obtained with a short acquisition time and reconstructed with FORE + OSEM showed clearer FDG uptake and more useful than 3D-OSEM in the light of the detection of lesions. |
format | Online Article Text |
id | pubmed-10581748 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Thieme Medical and Scientific Publishers Pvt. Ltd. |
record_format | MEDLINE/PubMed |
spelling | pubmed-105817482023-10-18 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time Tsuda, Keisuke Suzuki, Takayuki Toya, Kazuhito Sato, Eisuke Fujii, Hirofumi World J Nucl Med Objective In this study, we investigated the optimal reconstruction algorithm in fluorodeoxyglucose (FDG) positron emission tomography (PET) with a short acquisition time. Materials and Methods In the phantom study, six spheres filled with FDG solution (sphere size: 6.23–37 mm; radioactivity ratio of spheres to background = 8:1) and placed in a National Electrical Manufacturers Association phantom were evaluated. Image acquisition time was 15 to 180 seconds, and the obtained image data were reconstructed using each of the Fourier rebinning (FORE) + ordered subsets expectation-maximization (OSEM) and 3D-OSEM algorithms. In the clinical study, mid-abdominal images of 19 patients were evaluated using regions of interest placed on areas of low, intermediate, and high radioactivity. All obtained images were investigated visually, and quantitatively using maximum standardized uptake value (SUV) and coefficient of variation (CV). Results In the phantom study, FORE + OSEM images with a short acquisition time had large CVs (poor image quality) but comparatively constant maximum SUVs. 3D-OSEM images showed comparatively constant CVs (good image quality) but significantly low maximum SUVs. The results of visual evaluation were well correlated with those of quantitative evaluation. Small spheres were obscured on 3D-OSEM images with short acquisition time, but image quality was not greatly deteriorated. The clinical and phantom studies yielded similar results. Conclusion FDG PET images with a short acquisition time reconstructed by FORE + OSEM showed poorer image quality than by 3D-OSEM. However, images obtained with a short acquisition time and reconstructed with FORE + OSEM showed clearer FDG uptake and more useful than 3D-OSEM in the light of the detection of lesions. Thieme Medical and Scientific Publishers Pvt. Ltd. 2023-09-13 /pmc/articles/PMC10581748/ /pubmed/37854086 http://dx.doi.org/10.1055/s-0043-1774418 Text en The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. ( https://creativecommons.org/licenses/by/4.0/ ) https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Tsuda, Keisuke Suzuki, Takayuki Toya, Kazuhito Sato, Eisuke Fujii, Hirofumi 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time |
title | 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time |
title_full | 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time |
title_fullStr | 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time |
title_full_unstemmed | 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time |
title_short | 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time |
title_sort | 3d-osem versus fore + osem: optimal reconstruction algorithm for fdg pet with a short acquisition time |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10581748/ https://www.ncbi.nlm.nih.gov/pubmed/37854086 http://dx.doi.org/10.1055/s-0043-1774418 |
work_keys_str_mv | AT tsudakeisuke 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime AT suzukitakayuki 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime AT toyakazuhito 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime AT satoeisuke 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime AT fujiihirofumi 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime |