3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time

Objective  In this study, we investigated the optimal reconstruction algorithm in fluorodeoxyglucose (FDG) positron emission tomography (PET) with a short acquisition time. Materials and Methods  In the phantom study, six spheres filled with FDG solution (sphere size: 6.23–37 mm; radioactivity ratio...

Descripción completa

Detalles Bibliográficos
Autores principales: Tsuda, Keisuke, Suzuki, Takayuki, Toya, Kazuhito, Sato, Eisuke, Fujii, Hirofumi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Thieme Medical and Scientific Publishers Pvt. Ltd. 2023
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10581748/
https://www.ncbi.nlm.nih.gov/pubmed/37854086
http://dx.doi.org/10.1055/s-0043-1774418
_version_ 1785122196371275776
author Tsuda, Keisuke
Suzuki, Takayuki
Toya, Kazuhito
Sato, Eisuke
Fujii, Hirofumi
author_facet Tsuda, Keisuke
Suzuki, Takayuki
Toya, Kazuhito
Sato, Eisuke
Fujii, Hirofumi
author_sort Tsuda, Keisuke
collection PubMed
description Objective  In this study, we investigated the optimal reconstruction algorithm in fluorodeoxyglucose (FDG) positron emission tomography (PET) with a short acquisition time. Materials and Methods  In the phantom study, six spheres filled with FDG solution (sphere size: 6.23–37 mm; radioactivity ratio of spheres to background = 8:1) and placed in a National Electrical Manufacturers Association phantom were evaluated. Image acquisition time was 15 to 180 seconds, and the obtained image data were reconstructed using each of the Fourier rebinning (FORE) + ordered subsets expectation-maximization (OSEM) and 3D-OSEM algorithms. In the clinical study, mid-abdominal images of 19 patients were evaluated using regions of interest placed on areas of low, intermediate, and high radioactivity. All obtained images were investigated visually, and quantitatively using maximum standardized uptake value (SUV) and coefficient of variation (CV). Results  In the phantom study, FORE + OSEM images with a short acquisition time had large CVs (poor image quality) but comparatively constant maximum SUVs. 3D-OSEM images showed comparatively constant CVs (good image quality) but significantly low maximum SUVs. The results of visual evaluation were well correlated with those of quantitative evaluation. Small spheres were obscured on 3D-OSEM images with short acquisition time, but image quality was not greatly deteriorated. The clinical and phantom studies yielded similar results. Conclusion  FDG PET images with a short acquisition time reconstructed by FORE + OSEM showed poorer image quality than by 3D-OSEM. However, images obtained with a short acquisition time and reconstructed with FORE + OSEM showed clearer FDG uptake and more useful than 3D-OSEM in the light of the detection of lesions.
format Online
Article
Text
id pubmed-10581748
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Thieme Medical and Scientific Publishers Pvt. Ltd.
record_format MEDLINE/PubMed
spelling pubmed-105817482023-10-18 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time Tsuda, Keisuke Suzuki, Takayuki Toya, Kazuhito Sato, Eisuke Fujii, Hirofumi World J Nucl Med Objective  In this study, we investigated the optimal reconstruction algorithm in fluorodeoxyglucose (FDG) positron emission tomography (PET) with a short acquisition time. Materials and Methods  In the phantom study, six spheres filled with FDG solution (sphere size: 6.23–37 mm; radioactivity ratio of spheres to background = 8:1) and placed in a National Electrical Manufacturers Association phantom were evaluated. Image acquisition time was 15 to 180 seconds, and the obtained image data were reconstructed using each of the Fourier rebinning (FORE) + ordered subsets expectation-maximization (OSEM) and 3D-OSEM algorithms. In the clinical study, mid-abdominal images of 19 patients were evaluated using regions of interest placed on areas of low, intermediate, and high radioactivity. All obtained images were investigated visually, and quantitatively using maximum standardized uptake value (SUV) and coefficient of variation (CV). Results  In the phantom study, FORE + OSEM images with a short acquisition time had large CVs (poor image quality) but comparatively constant maximum SUVs. 3D-OSEM images showed comparatively constant CVs (good image quality) but significantly low maximum SUVs. The results of visual evaluation were well correlated with those of quantitative evaluation. Small spheres were obscured on 3D-OSEM images with short acquisition time, but image quality was not greatly deteriorated. The clinical and phantom studies yielded similar results. Conclusion  FDG PET images with a short acquisition time reconstructed by FORE + OSEM showed poorer image quality than by 3D-OSEM. However, images obtained with a short acquisition time and reconstructed with FORE + OSEM showed clearer FDG uptake and more useful than 3D-OSEM in the light of the detection of lesions. Thieme Medical and Scientific Publishers Pvt. Ltd. 2023-09-13 /pmc/articles/PMC10581748/ /pubmed/37854086 http://dx.doi.org/10.1055/s-0043-1774418 Text en The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. ( https://creativecommons.org/licenses/by/4.0/ ) https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Tsuda, Keisuke
Suzuki, Takayuki
Toya, Kazuhito
Sato, Eisuke
Fujii, Hirofumi
3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
title 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
title_full 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
title_fullStr 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
title_full_unstemmed 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
title_short 3D-OSEM versus FORE + OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
title_sort 3d-osem versus fore + osem: optimal reconstruction algorithm for fdg pet with a short acquisition time
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10581748/
https://www.ncbi.nlm.nih.gov/pubmed/37854086
http://dx.doi.org/10.1055/s-0043-1774418
work_keys_str_mv AT tsudakeisuke 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime
AT suzukitakayuki 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime
AT toyakazuhito 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime
AT satoeisuke 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime
AT fujiihirofumi 3dosemversusforeosemoptimalreconstructionalgorithmforfdgpetwithashortacquisitiontime