Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study
Objectives Collimators have a significant role in image quality and detectability in single-photon emission computed tomography (SPECT) imaging. Using an appropriate alloy that effectively absorbs scattered photons, without induced secondary x-rays, and with proper rigidity and weight may provide a...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Thieme Medical and Scientific Publishers Pvt. Ltd.
2023
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10581757/ https://www.ncbi.nlm.nih.gov/pubmed/37854088 http://dx.doi.org/10.1055/s-0043-1771287 |
_version_ | 1785122198509322240 |
---|---|
author | Darami, Maryam Mahmoudian, Babak Ljungberg, Michael Pirayesh Islamian, Jalil |
author_facet | Darami, Maryam Mahmoudian, Babak Ljungberg, Michael Pirayesh Islamian, Jalil |
author_sort | Darami, Maryam |
collection | PubMed |
description | Objectives Collimators have a significant role in image quality and detectability in single-photon emission computed tomography (SPECT) imaging. Using an appropriate alloy that effectively absorbs scattered photons, without induced secondary x-rays, and with proper rigidity and weight may provide an effective approach to the image improvement that conventionally collimators made of lead (Pb). Materials and Methods A Siemens E.CAM SPECT imaging system equipped with low-energy high-resolution (LEHR) collimator was simulated by the Simulating Medical Imaging Nuclear Detectors Monte Carlo program. Experimental and simulated data were compared based on a 2-mm (99m) Tc point source in an acrylic cylindrical Deluxe phantom (Data Spectrum, Inc). Seven types of tungsten (W) alloys (Wolfmet), with W content from 90 to 97% by weight, were then used as collimator materials of the simulated system. Camera parameters, such as energy- and spatial resolution, image contrast, and collimator-related parameters, such as fraction of septal penetration, scatter-to-primary ratios, and percentage of induced secondary x-rays, due to interactions in the collimator, were evaluated. Results Acceptable conformity was found for the simulated and experiment systems in terms of energy spectra, 10.113 and 10.140%, full width at half-maximum (FWHM) of the point spread function (PSF) curves, 8.78 and 9.06 mm, sensitivity, 78.46 and 78.34 cps/MBq, and contrast in images of 19.1 mm cold spheres in the Deluxe phantom, 79.17 and 78.97%, respectively. Results on the parameters of the simulated system with LEHR collimator made from the alloys showed that the alloy consisting of 90% W, 6% nickel, and 4% copper provided an FWHM of 8.76 mm, resulting in a 0.2% improvement in spatial resolution. Furthermore, all the Wolfmet collimators showed a 48% reduction in the amount of X-rays production compared to the Pb. Conclusion A Wolfmet LEHR collimator, made by a combination of W (90%), Ni (6%), and Cu (6%) provides a better image quality and detectability compared to the Pb. |
format | Online Article Text |
id | pubmed-10581757 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Thieme Medical and Scientific Publishers Pvt. Ltd. |
record_format | MEDLINE/PubMed |
spelling | pubmed-105817572023-10-18 Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study Darami, Maryam Mahmoudian, Babak Ljungberg, Michael Pirayesh Islamian, Jalil World J Nucl Med Objectives Collimators have a significant role in image quality and detectability in single-photon emission computed tomography (SPECT) imaging. Using an appropriate alloy that effectively absorbs scattered photons, without induced secondary x-rays, and with proper rigidity and weight may provide an effective approach to the image improvement that conventionally collimators made of lead (Pb). Materials and Methods A Siemens E.CAM SPECT imaging system equipped with low-energy high-resolution (LEHR) collimator was simulated by the Simulating Medical Imaging Nuclear Detectors Monte Carlo program. Experimental and simulated data were compared based on a 2-mm (99m) Tc point source in an acrylic cylindrical Deluxe phantom (Data Spectrum, Inc). Seven types of tungsten (W) alloys (Wolfmet), with W content from 90 to 97% by weight, were then used as collimator materials of the simulated system. Camera parameters, such as energy- and spatial resolution, image contrast, and collimator-related parameters, such as fraction of septal penetration, scatter-to-primary ratios, and percentage of induced secondary x-rays, due to interactions in the collimator, were evaluated. Results Acceptable conformity was found for the simulated and experiment systems in terms of energy spectra, 10.113 and 10.140%, full width at half-maximum (FWHM) of the point spread function (PSF) curves, 8.78 and 9.06 mm, sensitivity, 78.46 and 78.34 cps/MBq, and contrast in images of 19.1 mm cold spheres in the Deluxe phantom, 79.17 and 78.97%, respectively. Results on the parameters of the simulated system with LEHR collimator made from the alloys showed that the alloy consisting of 90% W, 6% nickel, and 4% copper provided an FWHM of 8.76 mm, resulting in a 0.2% improvement in spatial resolution. Furthermore, all the Wolfmet collimators showed a 48% reduction in the amount of X-rays production compared to the Pb. Conclusion A Wolfmet LEHR collimator, made by a combination of W (90%), Ni (6%), and Cu (6%) provides a better image quality and detectability compared to the Pb. Thieme Medical and Scientific Publishers Pvt. Ltd. 2023-09-06 /pmc/articles/PMC10581757/ /pubmed/37854088 http://dx.doi.org/10.1055/s-0043-1771287 Text en The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. ( https://creativecommons.org/licenses/by/4.0/ ) https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Darami, Maryam Mahmoudian, Babak Ljungberg, Michael Pirayesh Islamian, Jalil Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study |
title | Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study |
title_full | Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study |
title_fullStr | Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study |
title_full_unstemmed | Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study |
title_short | Impact of Wolfmet Tungsten Alloys as Parallel-Hole Collimator Material on Single-Photon Emission Computed Tomography Image Quality and Functional Parameters: A Simulating Medical Imaging Nuclear Detectors Monte Carlo Study |
title_sort | impact of wolfmet tungsten alloys as parallel-hole collimator material on single-photon emission computed tomography image quality and functional parameters: a simulating medical imaging nuclear detectors monte carlo study |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10581757/ https://www.ncbi.nlm.nih.gov/pubmed/37854088 http://dx.doi.org/10.1055/s-0043-1771287 |
work_keys_str_mv | AT daramimaryam impactofwolfmettungstenalloysasparallelholecollimatormaterialonsinglephotonemissioncomputedtomographyimagequalityandfunctionalparametersasimulatingmedicalimagingnucleardetectorsmontecarlostudy AT mahmoudianbabak impactofwolfmettungstenalloysasparallelholecollimatormaterialonsinglephotonemissioncomputedtomographyimagequalityandfunctionalparametersasimulatingmedicalimagingnucleardetectorsmontecarlostudy AT ljungbergmichael impactofwolfmettungstenalloysasparallelholecollimatormaterialonsinglephotonemissioncomputedtomographyimagequalityandfunctionalparametersasimulatingmedicalimagingnucleardetectorsmontecarlostudy AT pirayeshislamianjalil impactofwolfmettungstenalloysasparallelholecollimatormaterialonsinglephotonemissioncomputedtomographyimagequalityandfunctionalparametersasimulatingmedicalimagingnucleardetectorsmontecarlostudy |