Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer

OBJECTIVE: To evaluate the performance of newly developed glycated hemoglobin (HbA1c) clinical analytic reagents and HPLC columns, applied on Tosoh HLC-723 G8 Analyzer. METHODS: Newly developed reagents and columns were used on a Tosoh HLC-723 G8 Analyzer (standard mode) system to measure both of qu...

Descripción completa

Detalles Bibliográficos
Autores principales: Yuan, Bo, Yang, Wei, Zhang, Na, Shi, Hongyan, Dong, Shuangpeng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10582560/
https://www.ncbi.nlm.nih.gov/pubmed/37860220
http://dx.doi.org/10.1016/j.plabm.2023.e00338
_version_ 1785122358991781888
author Yuan, Bo
Yang, Wei
Zhang, Na
Shi, Hongyan
Dong, Shuangpeng
author_facet Yuan, Bo
Yang, Wei
Zhang, Na
Shi, Hongyan
Dong, Shuangpeng
author_sort Yuan, Bo
collection PubMed
description OBJECTIVE: To evaluate the performance of newly developed glycated hemoglobin (HbA1c) clinical analytic reagents and HPLC columns, applied on Tosoh HLC-723 G8 Analyzer. METHODS: Newly developed reagents and columns were used on a Tosoh HLC-723 G8 Analyzer (standard mode) system to measure both of qulity contorls and the clinical blood samples to evaluate the performances of these newly developed prodcuts including precision, accuracy, linearity, carryover, bias evaluation, correlation with commercial reagents, and stability according to CLSI recommendations. RESULTS: The CV of intra-assay precision and inter-assay precision of quality control and clinical blood sample assays using Lirimax products were both less than 3.00%. And the REs of accuracy were less than 6.00%. Linearity: R(2) = 0.9993 in the concentration range 4.77%–14.67%. Carryover: 0.05%. The Bland-Altman mean difference: −0.003583% HbA1c (CI: 0.07398: −0.08115); Passing-Bablok regression: y = 1.0022(0.9984:1.006)x-0.01097(-0.03776: 0.01582), R(2) = 0.9996. Stability evaluation was also acceptable. CONCLUSION: The performance of newly developed products was well evaluated for HbA1c measurement on a TOSOH G8 Analyzer which shows excellent suitability for clinical assay.
format Online
Article
Text
id pubmed-10582560
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-105825602023-10-19 Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer Yuan, Bo Yang, Wei Zhang, Na Shi, Hongyan Dong, Shuangpeng Pract Lab Med Original Research Article OBJECTIVE: To evaluate the performance of newly developed glycated hemoglobin (HbA1c) clinical analytic reagents and HPLC columns, applied on Tosoh HLC-723 G8 Analyzer. METHODS: Newly developed reagents and columns were used on a Tosoh HLC-723 G8 Analyzer (standard mode) system to measure both of qulity contorls and the clinical blood samples to evaluate the performances of these newly developed prodcuts including precision, accuracy, linearity, carryover, bias evaluation, correlation with commercial reagents, and stability according to CLSI recommendations. RESULTS: The CV of intra-assay precision and inter-assay precision of quality control and clinical blood sample assays using Lirimax products were both less than 3.00%. And the REs of accuracy were less than 6.00%. Linearity: R(2) = 0.9993 in the concentration range 4.77%–14.67%. Carryover: 0.05%. The Bland-Altman mean difference: −0.003583% HbA1c (CI: 0.07398: −0.08115); Passing-Bablok regression: y = 1.0022(0.9984:1.006)x-0.01097(-0.03776: 0.01582), R(2) = 0.9996. Stability evaluation was also acceptable. CONCLUSION: The performance of newly developed products was well evaluated for HbA1c measurement on a TOSOH G8 Analyzer which shows excellent suitability for clinical assay. Elsevier 2023-10-05 /pmc/articles/PMC10582560/ /pubmed/37860220 http://dx.doi.org/10.1016/j.plabm.2023.e00338 Text en © 2023 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Research Article
Yuan, Bo
Yang, Wei
Zhang, Na
Shi, Hongyan
Dong, Shuangpeng
Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
title Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
title_full Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
title_fullStr Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
title_full_unstemmed Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
title_short Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
title_sort evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on tosoh hlc-723 g8 analyzer
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10582560/
https://www.ncbi.nlm.nih.gov/pubmed/37860220
http://dx.doi.org/10.1016/j.plabm.2023.e00338
work_keys_str_mv AT yuanbo evaluationofnewlydevelopedglycatedhemoglobinclinicalanalyticreagentsandchromatographycolumnontosohhlc723g8analyzer
AT yangwei evaluationofnewlydevelopedglycatedhemoglobinclinicalanalyticreagentsandchromatographycolumnontosohhlc723g8analyzer
AT zhangna evaluationofnewlydevelopedglycatedhemoglobinclinicalanalyticreagentsandchromatographycolumnontosohhlc723g8analyzer
AT shihongyan evaluationofnewlydevelopedglycatedhemoglobinclinicalanalyticreagentsandchromatographycolumnontosohhlc723g8analyzer
AT dongshuangpeng evaluationofnewlydevelopedglycatedhemoglobinclinicalanalyticreagentsandchromatographycolumnontosohhlc723g8analyzer