Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report
BACKGROUND: Xen Gel Stent implant is a new minimally invasive surgical treatment for glaucoma that has been proven effectiveness and safety profile. However, it may also lead to some complications. Xen Gel Stent occlusion is a relatively rare complication reported less frequently and has limited tre...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10585744/ https://www.ncbi.nlm.nih.gov/pubmed/37858210 http://dx.doi.org/10.1186/s12886-023-03109-7 |
_version_ | 1785123009205370880 |
---|---|
author | Zhang, Yao Xiang, Haotian Zhang, Yangyang Tang, Li |
author_facet | Zhang, Yao Xiang, Haotian Zhang, Yangyang Tang, Li |
author_sort | Zhang, Yao |
collection | PubMed |
description | BACKGROUND: Xen Gel Stent implant is a new minimally invasive surgical treatment for glaucoma that has been proven effectiveness and safety profile. However, it may also lead to some complications. Xen Gel Stent occlusion is a relatively rare complication reported less frequently and has limited treatment experience. In our case report, we proposed a novel surgical treatment using a 10 − 0 nylon suture to successfully recanalize the occluded Xen45 Gel Stent. CASE PRESENTATION: A 16-year-old female patient had bilateral juvenile glaucoma for the past 5 years. Her right eye had undergone three glaucoma surgeries but failed. At a presentation to our hospital, the right eye’s intraocular pressure (IOP) was 30 mmHg despite applying four different active principles. Xen45 Gel Stent implant was chosen for treatment, but six days after implantation, the IOP rose to 40 mmHg as a result of an anterior chamber tip occlusion of the Xen45 Gel Stent. Nd: YAG laser shockwave therapy was attempted but failed. The patient eventually had to return to the operating room for a revision procedure. The Xen45 Gel Stent was recanalized from the ab externo by making an L-shaped conjunctival incision at the fornix base and threading a 10 − 0 nylon suture through it. The IOP was successfully controlled in the 11-month follow-up without medication. CONCLUSION: If postoperative occlusion arises after Xen45 Gel Stent implantation, surgery using 10 − 0 nylon suture to recanalize Xen45 Gel Stent should be considered as a relatively safe, effective that does not require removal of Xen45 Gel Stent. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-03109-7. |
format | Online Article Text |
id | pubmed-10585744 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-105857442023-10-20 Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report Zhang, Yao Xiang, Haotian Zhang, Yangyang Tang, Li BMC Ophthalmol Case Report BACKGROUND: Xen Gel Stent implant is a new minimally invasive surgical treatment for glaucoma that has been proven effectiveness and safety profile. However, it may also lead to some complications. Xen Gel Stent occlusion is a relatively rare complication reported less frequently and has limited treatment experience. In our case report, we proposed a novel surgical treatment using a 10 − 0 nylon suture to successfully recanalize the occluded Xen45 Gel Stent. CASE PRESENTATION: A 16-year-old female patient had bilateral juvenile glaucoma for the past 5 years. Her right eye had undergone three glaucoma surgeries but failed. At a presentation to our hospital, the right eye’s intraocular pressure (IOP) was 30 mmHg despite applying four different active principles. Xen45 Gel Stent implant was chosen for treatment, but six days after implantation, the IOP rose to 40 mmHg as a result of an anterior chamber tip occlusion of the Xen45 Gel Stent. Nd: YAG laser shockwave therapy was attempted but failed. The patient eventually had to return to the operating room for a revision procedure. The Xen45 Gel Stent was recanalized from the ab externo by making an L-shaped conjunctival incision at the fornix base and threading a 10 − 0 nylon suture through it. The IOP was successfully controlled in the 11-month follow-up without medication. CONCLUSION: If postoperative occlusion arises after Xen45 Gel Stent implantation, surgery using 10 − 0 nylon suture to recanalize Xen45 Gel Stent should be considered as a relatively safe, effective that does not require removal of Xen45 Gel Stent. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-03109-7. BioMed Central 2023-10-19 /pmc/articles/PMC10585744/ /pubmed/37858210 http://dx.doi.org/10.1186/s12886-023-03109-7 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Case Report Zhang, Yao Xiang, Haotian Zhang, Yangyang Tang, Li Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
title | Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
title_full | Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
title_fullStr | Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
title_full_unstemmed | Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
title_short | Recanalization of Xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
title_sort | recanalization of xen45 gel stent implant occlusion using 10 − 0 nylon suture in refractory glaucoma: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10585744/ https://www.ncbi.nlm.nih.gov/pubmed/37858210 http://dx.doi.org/10.1186/s12886-023-03109-7 |
work_keys_str_mv | AT zhangyao recanalizationofxen45gelstentimplantocclusionusing100nylonsutureinrefractoryglaucomaacasereport AT xianghaotian recanalizationofxen45gelstentimplantocclusionusing100nylonsutureinrefractoryglaucomaacasereport AT zhangyangyang recanalizationofxen45gelstentimplantocclusionusing100nylonsutureinrefractoryglaucomaacasereport AT tangli recanalizationofxen45gelstentimplantocclusionusing100nylonsutureinrefractoryglaucomaacasereport |