Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon

Currently, there is a lack of methods and tools that efficiently evaluate the auto-feathering junctions created by multileaf collimator (MLCs) for supine volumetric modulated arc therapy (VMAT) craniospinal irradiation (CSI) plans. We have investigated the feasibility of stitching together multi-iso...

Descripción completa

Detalles Bibliográficos
Autores principales: Misa, Joshua, McCarthy, Shane, St. Clair, William, Pokhrel, Damodar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10588278/
https://www.ncbi.nlm.nih.gov/pubmed/37868562
http://dx.doi.org/10.7759/cureus.45640
_version_ 1785123547565260800
author Misa, Joshua
McCarthy, Shane
St. Clair, William
Pokhrel, Damodar
author_facet Misa, Joshua
McCarthy, Shane
St. Clair, William
Pokhrel, Damodar
author_sort Misa, Joshua
collection PubMed
description Currently, there is a lack of methods and tools that efficiently evaluate the auto-feathering junctions created by multileaf collimator (MLCs) for supine volumetric modulated arc therapy (VMAT) craniospinal irradiation (CSI) plans. We have investigated the feasibility of stitching together multi-isocenter fluence maps to then analyze the feathered junctions for patient-specific quality assurance (QA). Furthermore, we investigated the capability of Halcyon for the treatment of CSI patients. Three patients, who previously underwent VMAT CSI treatment on TrueBeam (6-MV flattening filter-free (FFF)) for 36 Gy in 20 fractions were replanned for Halcyon. A multi-isocenter approach with only translational superior-inferior shifts was used for both platforms. Each isocenter consists of two full arcs with anterior avoidance sectors, ±5° collimator rotations between arcs, and 5-8 cm of overlapping MLC auto-feathering junctions. All plans were QA’d via electronic portal imaging device (EPID) portal dosimetry and analyzed with a gamma criteria of 3%/3 mm. A variety of plan quality metrics were analyzed to evaluate dose distributions to the target, doses to organs at risk (OARs), and integral dose to the patient. A MATLAB script was developed to stitch the calculated and measured fluence maps in order to perform patient-specific QA for the composite fluence. The Halcyon plans provided highly conformal and homogenous dose distributions to the entire CSI target, superior to the clinical TrueBeam plans, while sparing critical organs with significantly lower values of V10Gy and V18Gy by up to 2% and 2.5%, respectively. Qualitative depictions of vertical dose profiles from the stitched DICOM of the entire CSI target for both planned and delivered fluence maps demonstrated equivalency, with slightly lower average pass rates with Halcyon (97%) compared to TrueBeam (99.9%). This approach to stitch multiple measured versus calculated EPID fluence maps has shown to be a feasible and accurate method and will be helpful for comprehensive VMAT CSI QA on both platforms. Further implementation of this script will be used in examining dosimetric impacts of daily patient positioning errors at MLC auto-feathering junctions.
format Online
Article
Text
id pubmed-10588278
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-105882782023-10-21 Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon Misa, Joshua McCarthy, Shane St. Clair, William Pokhrel, Damodar Cureus Medical Physics Currently, there is a lack of methods and tools that efficiently evaluate the auto-feathering junctions created by multileaf collimator (MLCs) for supine volumetric modulated arc therapy (VMAT) craniospinal irradiation (CSI) plans. We have investigated the feasibility of stitching together multi-isocenter fluence maps to then analyze the feathered junctions for patient-specific quality assurance (QA). Furthermore, we investigated the capability of Halcyon for the treatment of CSI patients. Three patients, who previously underwent VMAT CSI treatment on TrueBeam (6-MV flattening filter-free (FFF)) for 36 Gy in 20 fractions were replanned for Halcyon. A multi-isocenter approach with only translational superior-inferior shifts was used for both platforms. Each isocenter consists of two full arcs with anterior avoidance sectors, ±5° collimator rotations between arcs, and 5-8 cm of overlapping MLC auto-feathering junctions. All plans were QA’d via electronic portal imaging device (EPID) portal dosimetry and analyzed with a gamma criteria of 3%/3 mm. A variety of plan quality metrics were analyzed to evaluate dose distributions to the target, doses to organs at risk (OARs), and integral dose to the patient. A MATLAB script was developed to stitch the calculated and measured fluence maps in order to perform patient-specific QA for the composite fluence. The Halcyon plans provided highly conformal and homogenous dose distributions to the entire CSI target, superior to the clinical TrueBeam plans, while sparing critical organs with significantly lower values of V10Gy and V18Gy by up to 2% and 2.5%, respectively. Qualitative depictions of vertical dose profiles from the stitched DICOM of the entire CSI target for both planned and delivered fluence maps demonstrated equivalency, with slightly lower average pass rates with Halcyon (97%) compared to TrueBeam (99.9%). This approach to stitch multiple measured versus calculated EPID fluence maps has shown to be a feasible and accurate method and will be helpful for comprehensive VMAT CSI QA on both platforms. Further implementation of this script will be used in examining dosimetric impacts of daily patient positioning errors at MLC auto-feathering junctions. Cureus 2023-09-20 /pmc/articles/PMC10588278/ /pubmed/37868562 http://dx.doi.org/10.7759/cureus.45640 Text en Copyright © 2023, Misa et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Medical Physics
Misa, Joshua
McCarthy, Shane
St. Clair, William
Pokhrel, Damodar
Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon
title Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon
title_full Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon
title_fullStr Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon
title_full_unstemmed Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon
title_short Development and Quality Assurance of Multileaf Collimator (MLC) Auto-Feathering Junctions for Multi-Isocenter Supine Volumetric Modulated Arc Therapy (VMAT) Craniospinal Axis Irradiation on Halcyon
title_sort development and quality assurance of multileaf collimator (mlc) auto-feathering junctions for multi-isocenter supine volumetric modulated arc therapy (vmat) craniospinal axis irradiation on halcyon
topic Medical Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10588278/
https://www.ncbi.nlm.nih.gov/pubmed/37868562
http://dx.doi.org/10.7759/cureus.45640
work_keys_str_mv AT misajoshua developmentandqualityassuranceofmultileafcollimatormlcautofeatheringjunctionsformultiisocentersupinevolumetricmodulatedarctherapyvmatcraniospinalaxisirradiationonhalcyon
AT mccarthyshane developmentandqualityassuranceofmultileafcollimatormlcautofeatheringjunctionsformultiisocentersupinevolumetricmodulatedarctherapyvmatcraniospinalaxisirradiationonhalcyon
AT stclairwilliam developmentandqualityassuranceofmultileafcollimatormlcautofeatheringjunctionsformultiisocentersupinevolumetricmodulatedarctherapyvmatcraniospinalaxisirradiationonhalcyon
AT pokhreldamodar developmentandqualityassuranceofmultileafcollimatormlcautofeatheringjunctionsformultiisocentersupinevolumetricmodulatedarctherapyvmatcraniospinalaxisirradiationonhalcyon