A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner
BACKGROUND: Positron Emission Mammography (PEM) is a nuclear medicine imaging tool, playing a significant role in the diagnosis of patients with breast cancer. These days, many research has been done in order to improve the performance of this system. OBJECTIVE: This study aims to propose a new meth...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Shiraz University of Medical Sciences
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10589686/ https://www.ncbi.nlm.nih.gov/pubmed/37868946 http://dx.doi.org/10.31661/jbpe.v0i0.2009-1190 |
_version_ | 1785123838800953344 |
---|---|
author | Roshani, Delband Setayeshi, Saeed |
author_facet | Roshani, Delband Setayeshi, Saeed |
author_sort | Roshani, Delband |
collection | PubMed |
description | BACKGROUND: Positron Emission Mammography (PEM) is a nuclear medicine imaging tool, playing a significant role in the diagnosis of patients with breast cancer. These days, many research has been done in order to improve the performance of this system. OBJECTIVE: This study aims to propose a new method for optimizing the size of axial Field of View (FOV) in PEMs and improving the performance of the systems. MATERIAL AND METHODS: In this analytical study, a conventional Inveon PET is simulated using GATE in order to validate the simulation. For this simulation, the mean relative difference is 2.91%, showing the precision and correction of simulation and consequently it is benchmarked. In the next step, for design of the new optimized detector, several validated simulations are performed in order to find the best geometry. RESULTS: The best result is obtained with the axial FOV of 101.7 mm. It has 1.6×1.6×15 mm(3) lutetium yttrium orthosilicate (LYSO) crystals. The detector consists of 6 block rings with 30 detector blocks in each ring. In this paper, the performance of the scanner is improved and the geometry is optimized. Sensitivity and scatter fraction of the designed scanner are 4.65% and 21.2%, respectively, also noise equivalent count rate (NECR) is 105.442 kcps. CONCLUSION: The results showed 1 up to 3% improvement in the sensitivity of this new detector compared with different PEMs. |
format | Online Article Text |
id | pubmed-10589686 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Shiraz University of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-105896862023-10-22 A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner Roshani, Delband Setayeshi, Saeed J Biomed Phys Eng Original Article BACKGROUND: Positron Emission Mammography (PEM) is a nuclear medicine imaging tool, playing a significant role in the diagnosis of patients with breast cancer. These days, many research has been done in order to improve the performance of this system. OBJECTIVE: This study aims to propose a new method for optimizing the size of axial Field of View (FOV) in PEMs and improving the performance of the systems. MATERIAL AND METHODS: In this analytical study, a conventional Inveon PET is simulated using GATE in order to validate the simulation. For this simulation, the mean relative difference is 2.91%, showing the precision and correction of simulation and consequently it is benchmarked. In the next step, for design of the new optimized detector, several validated simulations are performed in order to find the best geometry. RESULTS: The best result is obtained with the axial FOV of 101.7 mm. It has 1.6×1.6×15 mm(3) lutetium yttrium orthosilicate (LYSO) crystals. The detector consists of 6 block rings with 30 detector blocks in each ring. In this paper, the performance of the scanner is improved and the geometry is optimized. Sensitivity and scatter fraction of the designed scanner are 4.65% and 21.2%, respectively, also noise equivalent count rate (NECR) is 105.442 kcps. CONCLUSION: The results showed 1 up to 3% improvement in the sensitivity of this new detector compared with different PEMs. Shiraz University of Medical Sciences 2023-10-01 /pmc/articles/PMC10589686/ /pubmed/37868946 http://dx.doi.org/10.31661/jbpe.v0i0.2009-1190 Text en Copyright: © Journal of Biomedical Physics and Engineering https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License, ( http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Roshani, Delband Setayeshi, Saeed A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner |
title | A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner |
title_full | A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner |
title_fullStr | A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner |
title_full_unstemmed | A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner |
title_short | A New Method for Optimizing the Size of Axial FOV in TOF-PEM to Improve Performance of the Scanner |
title_sort | new method for optimizing the size of axial fov in tof-pem to improve performance of the scanner |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10589686/ https://www.ncbi.nlm.nih.gov/pubmed/37868946 http://dx.doi.org/10.31661/jbpe.v0i0.2009-1190 |
work_keys_str_mv | AT roshanidelband anewmethodforoptimizingthesizeofaxialfovintofpemtoimproveperformanceofthescanner AT setayeshisaeed anewmethodforoptimizingthesizeofaxialfovintofpemtoimproveperformanceofthescanner AT roshanidelband newmethodforoptimizingthesizeofaxialfovintofpemtoimproveperformanceofthescanner AT setayeshisaeed newmethodforoptimizingthesizeofaxialfovintofpemtoimproveperformanceofthescanner |