O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?

BACKGROUND: Accurate non-invasive oxygen saturation (SpO2) measurement is crucial in healthcare, however is challenged by numerous devices using variable algorithms1. We compared two commonly used oximeter devices worn simultaneously in patients during an overnight polysomnogram. AIM: To assess (1)...

Descripción completa

Detalles Bibliográficos
Autores principales: McDermott, E, Cuesta, R, van Braak, E, Spiteri, M, Davis, S, Beranek, R, Kaur-Bains, S, Slater, B, Naughton, M, Roebuck, T, Nguy, R
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10591675/
http://dx.doi.org/10.1093/sleepadvances/zpad035.009
_version_ 1785124273856184320
author McDermott, E
Cuesta, R
van Braak, E
Spiteri, M
Davis, S
Beranek, R
Kaur-Bains, S
Slater, B
Naughton, M
Roebuck, T
Nguy, R
author_facet McDermott, E
Cuesta, R
van Braak, E
Spiteri, M
Davis, S
Beranek, R
Kaur-Bains, S
Slater, B
Naughton, M
Roebuck, T
Nguy, R
author_sort McDermott, E
collection PubMed
description BACKGROUND: Accurate non-invasive oxygen saturation (SpO2) measurement is crucial in healthcare, however is challenged by numerous devices using variable algorithms1. We compared two commonly used oximeter devices worn simultaneously in patients during an overnight polysomnogram. AIM: To assess (1) differences between two SpO2 recording devices and (2) SpO2 accuracy against SaO2. METHODS: SpO2 data of two devices were recorded, downloaded and analysed with corresponding software: device 1 (Rad7 [Masimo, Irvine, CA, USA] set at 2-4sec averaging time) with software 1 (PSG 4 [Compumedics, Melbourne, Australia]) and device 2 (Wrist-Ox2® [Nonin Medical Inc, Plymouth, MN, USA] set at 3 sec averaging time) with software 2 (Noxturnal [Nox Medical Inc, Reykjavik, Iceland]). Responsiveness of each oximeter to changes in the patients’ desaturations associated with apnoeas and hypopnoeas overnight was assessed. In addition, both device SpO2 values were compared with SaO2 from simultaneous arterial blood gas sampling. PRELIMINARY RESULTS: 7 patients (age 56.4±7.1 years, BMI 31.6±2.2 kg/m2, 100% male, 87% light skinned) were studied. The mean SpO2 with RAD7 was higher than the Nonin (95.9±1.6 vs 94.2±1.4%, mean difference 1.6±1.0%). Differences in the responsiveness of the oximeters during overnight desaturations were also observed. However, both device SpO2 values were found to be within ±2% of SaO2. CONCLUSIONS: Discrepancies in SpO2 measurement do occur between devices and may have important clinical implications especially when it comes to the measurement of hypoxemic burden during sleep. 1.Blanchet MA et al. Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients. Respir Care. 2023 May;68(5):565-574.
format Online
Article
Text
id pubmed-10591675
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-105916752023-10-24 O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant? McDermott, E Cuesta, R van Braak, E Spiteri, M Davis, S Beranek, R Kaur-Bains, S Slater, B Naughton, M Roebuck, T Nguy, R Sleep Adv Oral Presentations BACKGROUND: Accurate non-invasive oxygen saturation (SpO2) measurement is crucial in healthcare, however is challenged by numerous devices using variable algorithms1. We compared two commonly used oximeter devices worn simultaneously in patients during an overnight polysomnogram. AIM: To assess (1) differences between two SpO2 recording devices and (2) SpO2 accuracy against SaO2. METHODS: SpO2 data of two devices were recorded, downloaded and analysed with corresponding software: device 1 (Rad7 [Masimo, Irvine, CA, USA] set at 2-4sec averaging time) with software 1 (PSG 4 [Compumedics, Melbourne, Australia]) and device 2 (Wrist-Ox2® [Nonin Medical Inc, Plymouth, MN, USA] set at 3 sec averaging time) with software 2 (Noxturnal [Nox Medical Inc, Reykjavik, Iceland]). Responsiveness of each oximeter to changes in the patients’ desaturations associated with apnoeas and hypopnoeas overnight was assessed. In addition, both device SpO2 values were compared with SaO2 from simultaneous arterial blood gas sampling. PRELIMINARY RESULTS: 7 patients (age 56.4±7.1 years, BMI 31.6±2.2 kg/m2, 100% male, 87% light skinned) were studied. The mean SpO2 with RAD7 was higher than the Nonin (95.9±1.6 vs 94.2±1.4%, mean difference 1.6±1.0%). Differences in the responsiveness of the oximeters during overnight desaturations were also observed. However, both device SpO2 values were found to be within ±2% of SaO2. CONCLUSIONS: Discrepancies in SpO2 measurement do occur between devices and may have important clinical implications especially when it comes to the measurement of hypoxemic burden during sleep. 1.Blanchet MA et al. Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients. Respir Care. 2023 May;68(5):565-574. Oxford University Press 2023-10-23 /pmc/articles/PMC10591675/ http://dx.doi.org/10.1093/sleepadvances/zpad035.009 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of Sleep Research Society. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Oral Presentations
McDermott, E
Cuesta, R
van Braak, E
Spiteri, M
Davis, S
Beranek, R
Kaur-Bains, S
Slater, B
Naughton, M
Roebuck, T
Nguy, R
O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?
title O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?
title_full O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?
title_fullStr O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?
title_full_unstemmed O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?
title_short O009 Are there Differences in SpO2 Recordings from Two Oximeters on the same hand during an Overnight Sleep Study Significant?
title_sort o009 are there differences in spo2 recordings from two oximeters on the same hand during an overnight sleep study significant?
topic Oral Presentations
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10591675/
http://dx.doi.org/10.1093/sleepadvances/zpad035.009
work_keys_str_mv AT mcdermotte o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT cuestar o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT vanbraake o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT spiterim o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT daviss o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT beranekr o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT kaurbainss o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT slaterb o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT naughtonm o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT roebuckt o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant
AT nguyr o009aretheredifferencesinspo2recordingsfromtwooximetersonthesamehandduringanovernightsleepstudysignificant