Analytical Validation of a Spiral Microfluidic Chip with Hydrofoil-Shaped Pillars for the Enrichment of Circulating Tumor Cells
The isolation of circulating tumor cells (CTCs) from peripheral blood with high efficiency remains a challenge hindering the utilization of CTC enrichment methods in clinical practice. Here, we propose a microfluidic channel design for the size-based hydrodynamic enrichment of CTCs from blood in an...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10605072/ https://www.ncbi.nlm.nih.gov/pubmed/37887131 http://dx.doi.org/10.3390/bios13100938 |
Sumario: | The isolation of circulating tumor cells (CTCs) from peripheral blood with high efficiency remains a challenge hindering the utilization of CTC enrichment methods in clinical practice. Here, we propose a microfluidic channel design for the size-based hydrodynamic enrichment of CTCs from blood in an epitope-independent and high-throughput manner. The microfluidic channel comprises a spiral-shaped part followed by a widening part, incorporating successive streamlined pillars, that improves the enrichment efficiency. The design was tested against two benchmark designs, a spiral microfluidic channel and a spiral microfluidic channel followed by a widening channel without the hydrofoils, by processing 5 mL of healthy blood samples spiked with 100 MCF-7 cells. The results proved that the design with hydrofoil-shaped pillars perform significantly better in terms of recovery (recovery rate of 67.9% compared to 23.6% in spiral and 56.7% in spiral with widening section), at a cost of slightly lower white blood cell (WBC) depletion (depletion rate of 94.2% compared to 98.6% in spiral and 94.2% in spiral with widening section), at 1500 µL/min flow rate. For analytical validation, the design was further tested with A549, SKOV-3, and BT-474 cell lines, yielding recovery rates of 62.3 ± 8.4%, 71.0 ± 6.5%, and 82.9 ± 9.9%, respectively. The results are consistent with the size and deformability variation in the respective cell lines, where the increasing size and decreasing deformability affect the recovery rate in a positive manner. The analysis before and after the microfluidic chip process showed that the process does not affect cell viability. |
---|